Datasheet<br />
<strong>Collimators</strong> <strong>for</strong> <strong>Radiography</strong><br />
<strong>and</strong> <strong>Fluoroscopy</strong><br />
Answers <strong>for</strong> life.
<strong>Collimators</strong> <strong>for</strong> <strong>Radiography</strong> <strong>and</strong> <strong>Fluoroscopy</strong><br />
Product Description<br />
The purpose of the multileaf collimator is to collimate<br />
exactly the radiation beam emerging from the X-ray tube<br />
assembly onto the size of the object to be displayed.<br />
Collimation is manual with the aid of the field light<br />
localizer (except A 01 II <strong>and</strong> AF 01 II) or motor-driven by<br />
automatic <strong>for</strong>mat collimation system.<br />
The current source-image distance (SID) <strong>and</strong> the size<br />
of the collimated radiation field are displayed on a LCD<br />
display except A 01 II, AF 01 II.<br />
A laser line localizer is available <strong>for</strong> centering the bucky<br />
except <strong>for</strong> type A 01 II, AF 01 II.<br />
For optimizing the image <strong>and</strong> radiation quality three<br />
different Cu-filter can be selected.<br />
Features<br />
◾ Lamellae close to the source to shield off extrafocal<br />
radiation<br />
◾ Additional prefilters integrated in the collimator <strong>for</strong><br />
manual (AL 01 II eL) or motorized operation (AL 02 II eL,<br />
A 01 II, AF 01 II, AFL 01 II eL)<br />
◾ Bucky light generated by laser results in a sharp<br />
light line <strong>for</strong> all source-image distances (not A 01 II <strong>and</strong><br />
AF 01 II)<br />
2<br />
◾ Display of field sizes optionally in cm (default) or inch<br />
where available<br />
◾ Rotation by ± 45° with a stop in 0° position by swiveling<br />
flange (not A 01 II <strong>and</strong> AF 01 II)<br />
◾ Tape measure integrated in a user friendly way in the<br />
collimator (not A 01 II <strong>and</strong> AF 01 II)<br />
◾ Two foam protected accessory rails facilitate attachment<br />
of additional devices <strong>and</strong> further filters (not A 01<br />
II <strong>and</strong> AF 01 II)<br />
Options <strong>and</strong> accessories<br />
Set of three-field templates <strong>for</strong><br />
SID 90 – 110 cm / 36 – 42‘’<br />
SID 110 – 130 cm / 42 – 50‘’<br />
SID 130 – 175 cm / 50 – 70‘’<br />
SID 175 – 220 cm / 70 – 90‘’<br />
The collimator is prepared <strong>for</strong> installation of a DAP<br />
chamber as well as a CANopen DAP chamber.<br />
Customizing<br />
In order to fulfill your particular requirements, we also<br />
provide you with customized versions of this product.
Technical Data<br />
AL 01 II eL AL 02 II eL AFL 01 II eL AF 01 II A 01 II<br />
Part number 10092612 10092614 10092618 10092603 10092606<br />
Collimation<br />
Minimum field size<br />
Maximum field size<br />
Radiation field scale tolerance<br />
Iris<br />
(Data <strong>for</strong> SID = 100 cm)<br />
Shape<br />
Minimum diameter<br />
Maximum diameter<br />
Additional prefilter<br />
Type<br />
3 cm x 3 cm with SID 100 cm<br />
43 cm x 43 cm with SID 90 cm<br />
< 0.75% of the SID<br />
None None Octagonal<br />
7 cm<br />
50 cm<br />
0, 0.1, 0.2, 0.3 mm Cu<br />
Octagonal<br />
7 cm<br />
50 cm<br />
Operation Manual Motor-driven Motor-driven Motor-driven Motor-driven<br />
Light localizer<br />
(Data <strong>for</strong> SID = 100 cm)<br />
Projection LED<br />
Bucky light line<br />
Brightness<br />
Contrast<br />
Radiation field – light field tol.<br />
Time switch<br />
Radiological properties<br />
Inherent filtration<br />
Radiation shielding<br />
24 Vdc / 25 W<br />
Laser<br />
> 200 lx<br />
> 4.5<br />
< 1.2% of the SID<br />
0...255 sec via CAN-Bus,<br />
10, 30, 60 or 90 sec fix per jumper<br />
1.0 mm Al at 70 kV<br />
150 kV<br />
No light<br />
localizers<br />
0.5 mm Al at 70 kV<br />
150 kV<br />
Leakage radiation < 0.5 mGy/h (at 150 kV / 3mA, at 1 m distance from tube assembly)<br />
Electrical interface<br />
Collimator power supply<br />
Light localizer power supply<br />
CAN interface power supply<br />
24 Vdc (± 10%), max. 2 A<br />
28 – 40 Vac (± 10%) max. 3 A or 24 Vdc (± 10%), max. 3 A<br />
13 – 27 Vdc (± 10%) 150 mA<br />
None<br />
24 Vdc (± 10%), max. 2 A<br />
Data interface CAN-BUS CAN-BUS CANopen DS402<br />
Mechanical interfaces<br />
Flange to the tube assembly<br />
Source flange distance<br />
Accessory rails<br />
Diameter: 136 mm, thickness: 2 mm<br />
80 mm<br />
2 rails<br />
Width: 178 mm; Length: 178 mm; Height of the guide:<br />
2 mm<br />
Quality st<strong>and</strong>ards CE, UL, CSA, DHHS<br />
Weight 10.5 ± 1.0 kg<br />
Ambient conditions<br />
10 ≤ T ≤ 40° C<br />
Operating temperature<br />
– 40 ≤ T ≤ 70° C<br />
Storage temperature<br />
20 – 75%<br />
Relative humidity<br />
70 – 106 kPa<br />
Barometric pressure<br />
70 – 106 kPa<br />
Service maintenance-free<br />
No accessory rail available<br />
3
<strong>Collimators</strong> <strong>for</strong> <strong>Radiography</strong> <strong>and</strong> <strong>Fluoroscopy</strong><br />
Models AL01 II eL, AL02 II eL<br />
<strong>and</strong> AFL01 II eL<br />
Model A01 II, AF01 II<br />
All dimensions in mm<br />
4
<strong>Collimators</strong> <strong>for</strong> <strong>Radiography</strong> <strong>and</strong> <strong>Fluoroscopy</strong><br />
Your Contact<br />
If you would like any further in<strong>for</strong>mation or to arrange<br />
a consultation, please feel free to contact us.<br />
<strong>Siemens</strong> AG<br />
Healthcare Sector<br />
Clinical Products<br />
Components & Vacuum Technology<br />
H CP CV OEM<br />
Henkestr. 127<br />
DE-91052 Erlangen<br />
Germany<br />
Phone: +49 9131 84-6911<br />
Fax: +49 9131 84-2501<br />
E-mail: oem.healthcare@siemens.com<br />
www.siemens.com/oemproducts<br />
5
The <strong>Siemens</strong> Healthcare Business<br />
Unit Components & Vacuum<br />
Technology is ISO 9001 <strong>and</strong> ISO<br />
13485 certified, manufactures in<br />
accordance with the Quality System<br />
Regulations (QSR) as defined by the<br />
U.S. Food <strong>and</strong> Drug Administration<br />
(FDA) <strong>and</strong> endeavors to comply with<br />
legal requirements concerning the<br />
environmental compatibility of its<br />
products.<br />
<strong>Siemens</strong> reserves the right to<br />
modify the design <strong>and</strong> specifications<br />
contained herein without prior<br />
notice. All rights reserved,<br />
particularly in connection with<br />
patent applications or registrations<br />
of utility model or design. This<br />
document is not considered to be<br />
a contractual specification. Kindly<br />
contact <strong>Siemens</strong> AG prior to using<br />
this in<strong>for</strong>mation <strong>for</strong> equipment<br />
design.<br />
Local Contact In<strong>for</strong>mation<br />
<strong>Siemens</strong> AG<br />
Healthcare Sector<br />
H CP CV OEM<br />
Henkestr. 127<br />
DE-91052 Erlangen<br />
Germany<br />
Phone: +49 9131 84-6911<br />
Fax: +49 9131 84-2501<br />
oem.healthcare@siemens.com<br />
Global <strong>Siemens</strong> Headquarters<br />
<strong>Siemens</strong> AG<br />
Wittelsbacherplatz 2<br />
80333 Muenchen<br />
Germany<br />
These components <strong>and</strong><br />
configurations are not finished<br />
medical devices. Compliance with<br />
all laws <strong>and</strong> regulations that are<br />
applicable to finished medical<br />
devices is the responsibility of the<br />
assembler/manufacturer of the<br />
finished medical device.<br />
The in<strong>for</strong>mation in this document<br />
contains general descriptions of<br />
the technical options available, which<br />
do not always have to be present<br />
in individual cases. The required<br />
features should there<strong>for</strong>e be<br />
specified in each individual case at<br />
the time of closing the contract.<br />
The components are labeled as<br />
“Manufactured by <strong>Siemens</strong>”.<br />
However, the buyer shall not market<br />
the components using the “<strong>Siemens</strong>”<br />
br<strong>and</strong> name <strong>and</strong>/or trademark.<br />
Global <strong>Siemens</strong> Healthcare<br />
Headquarters<br />
<strong>Siemens</strong> AG<br />
Healthcare Sector<br />
Henkestrasse 127<br />
91052 Erlangen<br />
Germany<br />
Phone: +49 9131 84-0<br />
www.siemens.com/healthcare<br />
Legal Manufacturer<br />
<strong>Siemens</strong> AG<br />
Wittelsbacherplatz 2<br />
DE-80333 Muenchen<br />
Germany<br />
Printed in Germany | CC 283 041200 | 10306993_DS_<strong>Collimators</strong>_<strong>for</strong>_R&F 2012-04 | © 04.2012, <strong>Siemens</strong> AG<br />
www.siemens.com/healthcare<br />
Global Business Unit<br />
<strong>Siemens</strong> AG<br />
Medical Solutions<br />
Components <strong>and</strong> Vacuum Technology<br />
Henkestr. 127<br />
DE-91052 Erlangen<br />
Germany<br />
Phone: +49 9131 84-0<br />
www.siemens.com/oemproducts<br />
The buyer may integrate these<br />
components into the medical system<br />
using its own br<strong>and</strong>s <strong>and</strong> labels.<br />
The product names <strong>and</strong>/or br<strong>and</strong>s<br />
referred to are the property of their<br />
respective trademark holders.<br />
This product bears a CE marking in<br />
accordance with the provisions of<br />
the 93/42/EEC directive concerning<br />
medical products.