CAS 750 Monitor - Diagnos

idel.medical.com

CAS 750 Monitor - Diagnos

CAS 750 Monitor

Multi-Parameter Monitor

• Easy to handle

• Compact

• 2 kg (for fully configured version)

For all Patients

• Adult

• Pediatric

• Neonatal

ECG

• For 3 or 5 Lead Applications.

Monitors with ECG have

Respiration and Temperature

included.

Oximetry Choices

• Masimo SET ®

• Nellcor ® OxiMax ®

EtCO 2

• Oridion MicroStream ®

• For intubated and nonintubated

patients.

NIBP

• Fast, accurate and reliable.

• Motion Artifact Extraction.

Any configuration you want…

CAS 750E-1 ECG

CAS 750E-2 ECG and SpO 2

CAS 750E-3 ECG, SpO 2 and NIBP

CAS 750E-3C ECG, SpO 2 and EtCO 2

CAS 750E-4 ECG, SpO 2 , EtCO 2 and NIBP

CAS 750C-2 EtCO 2 and SpO 2

CAS 750C-3 EtCO 2 , SpO 2 and NIBP

CAS 750S-2 SpO 2 and NIBP

Power

• Integrated AC power supply.

• User replaceable NiMH

battery pack.

• DC Power option.

Display

• Up to 3 Waveforms.

• Grouped Numerics.

• Trend and Alarm History.

Accessories / Options

• Printer with infrared interface.

• Nurse Call Interface.

• Roll Stand with basket.

• Swiveled Hard Mount.

• Carrying Case.

®


CAS 750 Multi-Parameter Monitor

ECG

Input:

3 or 5 leads

Leads:

I, II, III, aVL, aVF, aVR, V

Frequency Response: 0.5 to 40 Hz

Gain: Automatic or selectable (X0.5, X1, X2, X4, X8 or X16)

Pacer Detection: Yes

Pacer Rejection: Yes

Heart Rate Range: Adult 15-300 BPM, Neonate 15-350 BPM

Accuracy:

± 3 BPM or ± 3 %, whichever is greater

Resolution:

1 BPM

Respiration (ECG)

Method:

Range:

Accuracy:

Resolution:

Impedance Pneumography

0-150 BrPM

0-122 BrPM, ± 1BrPM

123-150 BrPM, ± 2 BrPM

1 BrPM

Temperature

Input:

YSI ® 400 compatible

Range: 28.1-43.0 °C

Accuracy: ± 0.1 °C

Resolution: 0.1 °C

NIBP

Technique:

Oscillometric (MAXNIBP ® Technology)

Patient Range: Adult - Neonate

Measurement Range: Adult

Neonate

Systolic: 30-255 mmHg 30-135 mmHg

Diastolic: 15-220 mmHg 15-110 mmHg

MAP: 20-235 mmHg 20-125 mmHg

Pulse Rate Range: 30-240 BPM 40-240 BPM

Accuracy:

Blood Pressure: ± 5 mmHg, Standard deviation no greater

than 8 mmHg.

Pulse Rate: ± 2% or ± 2 BPM, whichever is greater.

Oximetry

Masimo SET ®

%SpO 2 Range: 0-100%

%SpO 2 Accuracy: 70-100%, ± 2 digits (1 S.D.) Adult

70-100%, ± 3 digits (1 S.D.) Neonate

Pulse Rate Range: 25-240 BPM

Pulse Rate Accuracy: ± 3 BPM

Nellcor ® OxiMax ®

%SpO 2 Range: 1-100%

%SpO 2 Accuracy: 70-100%, ± 2 digits (1 S.D.) Adult

70-100%, ± 3 digits (1 S.D.) Neonate

Pulse Rate Range: 20-240 BPM

Pulse Rate Accuracy: ± 3 digits

EtCO 2

Method: Sidestream Capnography (MicroStream ® )

Units:

mmHg, % or kPa

Sample Rate: 50 ml/min ± 7.5 ml/min

Range:

0-99 mmHg (0-13.2 kPa and 0-13.0 vol% at sea

level)

Warm-Up Time: 30 seconds typical, reaches steady-state

accuracy 20 minutes after power up.

Accuracy:

0-38 mmHg, ± 2 mmHg

39-99 mmHg, ± 5% of reading

Resolution:

1 mmHg, 0.1 % or 0.1 kPa

Respiration Rate: 0-150 breaths/min

Patient Alarms

Adjustable Alarms: High and low alarms for Heart Rate, %SpO 2 ,

Pulse Rate, EtCO 2 , FiCO 2, Respiration Rate,

Systolic Pressure, Diastolic Pressure,

No Respiration.

Fixed Alarms: Asystole

Alarm Indicators: Audible

Red Patient Alarm LED’s

Yellow Equipment Alarm LED’s

Message Window

Operating Modes

Patient Modes:

NIBP Modes:

Other Modes:

Neonate or Adult

Manual, STAT or Automatic (at preset intervals).

Stand-by mode and Sleep Mode.

Display

Numerics:

HR, PR, %SpO 2 , NIBP (Systolic, Diastolic

and MAP or PR), EtCO 2 , (mmHg, % or

kPa), FiCO 2 (for C versions only),

RR, Temp (in °C or °F).

Waveforms: Configurable (up to 3), ECG, SpO 2 ,

Impedance Respiration, CO 2 .

Sweep Speeds: 12.5, 25 or 50 mm/s for ECG and SpO 2 .

3.0, 6.25 or 12.5 mm/s for Resp. and CO 2 .

History:

480 1-minute entries of all parameters

Alarm History (25 most recent alarms)

Physical Dimensions & Weight (Fully configured unit with battery)

H x W x D:

170 mm x 215 mm x 102 mm

Weight:

2.0 kg

Operating Environment

Operating Temp.: 0-50°C

Storage Temp.: -20 to 60°C

Humidity:

15-95% RH, non-condensing

Altitude:

-382 to 3048 m (1060 hPa to 697 hPa)

Power

Source:

AC:

DC (optional):

Battery:

Charge Time:

Operating Time:

Chassis Leakage

Current:

External AC power or internal battery.

100-240 VAC, 50/60 Hz, 0.5 A, 0.5 A fuse.

12 VDC

Smart NiMH battery pack (user removable).

3-5 hours

3 hrs (750E-4 with NIBP at 15 minute intervals).

100 µA (maximum)

Standards

Units comply with the following requirements:

ANSI/AAMI SP-10: 2002

EN 60601-1, EN 60601-1-2

EN 60601-2-27, EN 60601-2-30, EN 60601-20-49

EN 864, EN 865

UL classified - UL 60601-1, CAN/CSA C22.2 No. 601.1

CE marking according to Directive 93/42/EEC

The CAS logo is a registered trademark of CAS Medical Systems, Inc.

MAXNIBP ® is a registered trademark of CAS Medical Systems, Inc.

Masimo SET ® is a registered trademark of Masimo Inc.

Nellcor ® OxiMax ® is a registered trade mark of Nellcor Puritan Bennett Inc.

Oridion MicroStream ® is a registered trademark of Oridion Medical, Ltd.

YSI ® is a registered trademark of YSI, Inc.

Certificate Number FM72259

Distributed by:

Product marking: 0086

Corporate Office:

CAS MEDICAL SYSTEMS, INC.

44 East Industrial Road, Branford, CT 06405 USA

phone: +1-203 488 6056, fax: +1-203 488 9438

sales@casmed.com

www.casmed.com

Authorized European Union Representative:

mossaConsulting GmbH

Bollbrügg 22, 23570 Lübeck, Germany

Tel: +49-4502 880 557, fax: +49-4502 880 559

mossa.rod@t-online.de

CAS reserves the right to make changes to this sheet and the product at any time without notice. 21-05-0127 REV 01

More magazines by this user