The <strong>Dolomite</strong> Centre Ltd<br />

Specifications<br />

<strong>Sensor</strong> Specification a<br />

Part No.<br />

3000238<br />

Part No.<br />

3000239<br />

Part No.<br />

3000240<br />

Part No.<br />

3000241<br />

Part No.<br />

3000242<br />

Max. <strong>Flow</strong> <strong>Rate</strong> (µl/min) 4000 1000 40.0 7.0 1.5<br />

Calib. Min <strong>Flow</strong> (µl/min) 200 40 1.0 0.25 0.05<br />

Typical Resolution (µl/min)<br />

300 to<br />

5000<br />

100 to<br />

1500<br />

7 1.5 0.5<br />

Capillary internal diameter<br />

(mm)<br />

Max operating pressure<br />

(bar)<br />

Typical flow detection<br />

response time (ms)<br />

2.0 1.0 0.48 0.15 0.025<br />

3 5 150 350 350<br />

30 30 20 20 20<br />

Operating temperature (°C) 0 to 60 0 to 60 10 to 45 10 to 45 10 to 45<br />

Fluid Connectors<br />

UNF 10-32 2B<br />

PEEK Micro Fitting for 360µm<br />

Capillaries<br />

Max Internal Volume (µl) - - 10.3 1 0.15<br />

Typ. Total Mass (g) 36 36 33 33 33<br />

Connected Fluid Capillary,<br />

ID (mm)<br />

Connected Fluid Capillary,<br />

OD (mm)<br />

≥0.8 b ≥0.5 b N/A N/A N/A<br />

1.6 1.6 0.36 0.36 0.36<br />

Wetted Materials<br />

Glass (Duran ® )capillary with<br />

PEEK fittings & Teflon ® seal<br />

Fused Silica (quartz)Capillary<br />

with PEEK fittings<br />

Housing Material<br />

Power Supply (DC)<br />

Operating Current<br />

Power Dissipation<br />

Interface<br />

Chemically inert PBT<br />

ABS. max. rating, Minimum = 7V, Typical = 9V, Maximum = 18V<br />

VDD = 9V, no load, Typical Value = 20mA<br />

VDD = 9V, 3kΩ at RS232 output, Typical Value = 27mA<br />

VDD = 9V, no load, Typical Value = 180mW<br />

RS-232<br />

a<br />

Reference conditions: 1bar abs, const. 23°C, offset 0, medium H 2O, vertical mounting position.<br />

b<br />

Values smaller than this may cause turbulences that influence the accuracy of measurements.<br />

Warning: To avoid blockage the use of an appropriate inline filter is highly recommended, especially for the<br />

smaller capillary flow sensors (Part No.3000240, 3000241 or 3000242).<br />

<strong>Microfluidic</strong> <strong>Flow</strong> <strong>Rate</strong> <strong>Sensor</strong>_product datasheet

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!