The <strong>Dolomite</strong> Centre Ltd<br />
Specifications<br />
<strong>Sensor</strong> Specification a<br />
Part No.<br />
3000238<br />
Part No.<br />
3000239<br />
Part No.<br />
3000240<br />
Part No.<br />
3000241<br />
Part No.<br />
3000242<br />
Max. <strong>Flow</strong> <strong>Rate</strong> (µl/min) 4000 1000 40.0 7.0 1.5<br />
Calib. Min <strong>Flow</strong> (µl/min) 200 40 1.0 0.25 0.05<br />
Typical Resolution (µl/min)<br />
300 to<br />
5000<br />
100 to<br />
1500<br />
7 1.5 0.5<br />
Capillary internal diameter<br />
(mm)<br />
Max operating pressure<br />
(bar)<br />
Typical flow detection<br />
response time (ms)<br />
2.0 1.0 0.48 0.15 0.025<br />
3 5 150 350 350<br />
30 30 20 20 20<br />
Operating temperature (°C) 0 to 60 0 to 60 10 to 45 10 to 45 10 to 45<br />
Fluid Connectors<br />
UNF 10-32 2B<br />
PEEK Micro Fitting for 360µm<br />
Capillaries<br />
Max Internal Volume (µl) - - 10.3 1 0.15<br />
Typ. Total Mass (g) 36 36 33 33 33<br />
Connected Fluid Capillary,<br />
ID (mm)<br />
Connected Fluid Capillary,<br />
OD (mm)<br />
≥0.8 b ≥0.5 b N/A N/A N/A<br />
1.6 1.6 0.36 0.36 0.36<br />
Wetted Materials<br />
Glass (Duran ® )capillary with<br />
PEEK fittings & Teflon ® seal<br />
Fused Silica (quartz)Capillary<br />
with PEEK fittings<br />
Housing Material<br />
Power Supply (DC)<br />
Operating Current<br />
Power Dissipation<br />
Interface<br />
Chemically inert PBT<br />
ABS. max. rating, Minimum = 7V, Typical = 9V, Maximum = 18V<br />
VDD = 9V, no load, Typical Value = 20mA<br />
VDD = 9V, 3kΩ at RS232 output, Typical Value = 27mA<br />
VDD = 9V, no load, Typical Value = 180mW<br />
RS-232<br />
a<br />
Reference conditions: 1bar abs, const. 23°C, offset 0, medium H 2O, vertical mounting position.<br />
b<br />
Values smaller than this may cause turbulences that influence the accuracy of measurements.<br />
Warning: To avoid blockage the use of an appropriate inline filter is highly recommended, especially for the<br />
smaller capillary flow sensors (Part No.3000240, 3000241 or 3000242).<br />
<strong>Microfluidic</strong> <strong>Flow</strong> <strong>Rate</strong> <strong>Sensor</strong>_product datasheet