Analysis of Cr (VI) in Trivalent Chromate Coating Film with UH5300

Analysis of Cr (VI) in Trivalent Chromate Coating Film with UH5300

Analysis of Cr (VI) in Trivalent Chromate Coating Film with UH5300

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>Analysis</strong> <strong>of</strong> <strong>Cr</strong> (<strong>VI</strong>) <strong>in</strong> <strong>Trivalent</strong> <strong>Chromate</strong> Coat<strong>in</strong>g <strong>Film</strong> <strong>with</strong> <strong>UH5300</strong><br />

<strong>UH5300</strong><br />

INTRODUCTION<br />

The standard <strong>in</strong>stallation <strong>of</strong> <strong>UH5300</strong> spectrophotometer <strong>in</strong>cludes an automatic 6-cell turret. When analyz<strong>in</strong>g many<br />

samples, the <strong>in</strong>creased work efficiency can be expected by us<strong>in</strong>g the “<strong>in</strong>telligent Start” function. <strong>UH5300</strong><br />

spectrophotometer also focuses on the absorbance differences between cells. All cells are measured for the auto-zero<br />

values, which are recorded as the correction values. Therefore, the <strong>in</strong>strument is capable <strong>of</strong> handl<strong>in</strong>g the analysis where<br />

a slight absorbance difference is be<strong>in</strong>g measured. This time, <strong>Cr</strong>(<strong>VI</strong>) <strong>in</strong> water samples was analyzed by us<strong>in</strong>g this function.<br />

As a result <strong>of</strong> the analysis, a calibration curve <strong>with</strong> a high correlation coefficient (R 2 = 0.9999) was obta<strong>in</strong>ed.<br />

Analyte<br />

<strong>Analysis</strong> method<br />

METHOD<br />

: <strong>Cr</strong>(<strong>VI</strong>)<br />

: Diphenylcarbazide absorptiometry<br />

(Simple analysis by us<strong>in</strong>g a water quality test kit<br />

<strong>of</strong> Kyoritsu Chemical Check Lab., Corp.)<br />

Reagent : Water quality test kit No.31 <strong>Cr</strong> 6+ ,<br />

Kyoritsu Chemical Check Lab., Corp.<br />

Range <strong>of</strong> quantitative analysis : 0.02 - 1.0 mg/L<br />

Refer to JIS K0102 65.2 for the <strong>of</strong>ficial method<br />

Instrument<br />

: <strong>UH5300</strong><br />

ANALYTICAL CONDITIONS<br />

Slit<br />

: 1 nm<br />

Sample<br />

Stir<br />

PREPARATION<br />

25 mL<br />

Add reaction reagent R-1 (*1)<br />

Let stand for 5 m<strong>in</strong><br />

Solution for measurement<br />

Temperature: Room temperature<br />

*1 Kyoritsu Chemical Check Lab., Corp.<br />

Reagent Set for Water Analyzer No.31 <strong>Cr</strong> 6+<br />

Scan speed<br />

: 400 nm/m<strong>in</strong><br />

Measurement wavelength<br />

: 542 nm<br />

** Contact Kyoritsu Chemical Check Lab.<br />

Corp. for the <strong>in</strong>formation <strong>in</strong>clud<strong>in</strong>g the<br />

effects <strong>of</strong> the <strong>in</strong>terfer<strong>in</strong>g substances<br />

and the details <strong>of</strong> the reagents.<br />

High throughput measurement by “<strong>in</strong>telligent Start” function<br />

Open the sample<br />

chamber<br />

Set the sample<br />

Set the cell <strong>in</strong> the designated<br />

position.<br />

6 cells can be used.<br />

Close the sample<br />

chamber<br />

When the lid <strong>of</strong> the sample<br />

chamber is closed, the sample<br />

measurement will start<br />

automatically.<br />

Completion <strong>of</strong><br />

Start <strong>of</strong> Measurement<br />

measurement<br />

• When the sample cell is set and the sample chamber is closed, the measurement will start<br />

automatically. (Patent pend<strong>in</strong>g: Patent application No. 2012-181302)<br />

• The start<strong>in</strong>g operation after switch<strong>in</strong>g the sample cells is omitted.<br />

KEY WORDS<br />

Electronics/Semiconductor Related, Other Electronics/Semiconductor Related,<br />

Environmental Chemistry, Environment, <strong>Cr</strong>(<strong>VI</strong>), Absorption Spectrum,<br />

Calibration Curve, Color<strong>in</strong>g Reagent, Hexavalent Chromium, UV, <strong>UH5300</strong>,<br />

U-5100, U-2900,U-2910<br />

Spectrophotometer<br />

(UV)<br />

Sheet No.<br />

UV120005-01<br />

Hitachi High-Technologies Corporation


<strong>Analysis</strong> <strong>of</strong> <strong>Cr</strong>(<strong>VI</strong>) <strong>in</strong> <strong>Trivalent</strong> <strong>Chromate</strong> Coat<strong>in</strong>g <strong>Film</strong> <strong>with</strong> <strong>UH5300</strong><br />

<strong>UH5300</strong><br />

0.8<br />

0.8<br />

0.6<br />

542 nm<br />

1 mg/L<br />

0.6<br />

0.5 mg/L<br />

Abs<br />

0.4<br />

Abs<br />

0.4<br />

0.2 mg/L<br />

0.1 mg/L<br />

0.2<br />

0.2<br />

0.02 mg/L<br />

0.0<br />

0.0 0.2 0.4 0.6 0.8 1.0<br />

Concentration (mg/L)<br />

Calibration Curve <strong>of</strong> <strong>Cr</strong> (<strong>VI</strong>)<br />

R 2 : 0.9999<br />

0.0<br />

400 500 600 700<br />

Wavelength (nm)<br />

Absorption Spectrum <strong>of</strong> <strong>Cr</strong> (<strong>VI</strong>)<br />

Blank<br />

Addition Recovery Test for Extract<br />

Extract Extract + 0.5 mg/L Recovery rate<br />

0.16 ± 0.002 0.69 ± 0.003 105.6 ± 0.6 %<br />

KEY WORDS<br />

Spectrophotometer<br />

Electronics/Semiconductor Related, Other Electronics/Semiconductor Related,<br />

(UV)<br />

Environmental Chemistry, Environment, <strong>Cr</strong>(<strong>VI</strong>), Absorption Spectrum,<br />

Calibration Curve, Color<strong>in</strong>g Reagent, Hexavalent Chromium, UV, <strong>UH5300</strong>, Sheet No. UV120005-02<br />

U-5100, U-2900, U-2910<br />