Cartridge Equilibration<br />

• Remove cartridges from shipping container and store as follows prior to use:<br />

Cartridge(s)<br />

Equilibration<br />

Time<br />

Storage<br />

Temperature<br />

Warm up<br />

Time<br />

BG, CC 72 hours 15-30°C (59-86°F) None<br />

H3, H4, GL,<br />

LA<br />

1 hour 15-30°C (59-86°F) None<br />

CR None 2-8°C (35.6-46.4°F) 15 min<br />

• If storage range fluctuates more than 8°C (14.4°F), cartridges must go through<br />

an additional equilibration prior to use.<br />

• Cartridges can be used in temperatures down to 12°C, 54°F if used within 4 hours<br />

of transfer from the 15-30°C storage.