IRMA TRUpoint
IRMA TRUpoint
IRMA TRUpoint
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Cartridge Equilibration<br />
• Remove cartridges from shipping container and store as follows prior to use:<br />
Cartridge(s)<br />
Equilibration<br />
Time<br />
Storage<br />
Temperature<br />
Warm up<br />
Time<br />
BG, CC 72 hours 15-30°C (59-86°F) None<br />
H3, H4, GL,<br />
LA<br />
1 hour 15-30°C (59-86°F) None<br />
CR None 2-8°C (35.6-46.4°F) 15 min<br />
• If storage range fluctuates more than 8°C (14.4°F), cartridges must go through<br />
an additional equilibration prior to use.<br />
• Cartridges can be used in temperatures down to 12°C, 54°F if used within 4 hours<br />
of transfer from the 15-30°C storage.