Views
3 years ago

IRMA TRUpoint

IRMA TRUpoint

Correlation Setting

Correlation Setting (Optional) • QA Users can establish correlation parameters for each IRMA TRUpoint product type to adjust IRMA TRUpoint patient test results to values that would be expected if samples were run on another reference analyzer. There are 2 correlation options: • Off: Correlation parameters will not be used by the analyzer to adjust IRMA TRUpoint patient test results. • On: Correlation parameters will be used by the analyzer to adjust IRMA TRUpoint patient test results.

Activating Correlations 1. Highlight Correlation from TEST SETTINGS and press edit. An Attention message appears, stating that adjusting correlation factors will affect future patient test results. Press ok. 2. Highlight the desired Product Type and press edit. 3. Highlight ”on” or ”off” button and press next. If ”on” is selected, go to “Establishing or Changing Correlation Factors”.