Operating manualTriOxmatic ® 700 IQTriOxmatic ® 700 IQ SWTriOxmatic 700 IQTriOxmatic 700 IQ SWIQ SENSOR NET D.O. sensorba25317e06 03/2005


Accuracy when going topressThe use of advanced technology and the high quality standard of ourinstruments are the result of continuous development. This may resultin differences between this operating manual and your instrument.Also, we cannot guarantee that there are absolutely no errors in thismanual. Therefore, we are sure you will understand that we cannotaccept any legal claims resulting from the data, figures or descriptions.NoteThe latest version of the present operating manual can be found on theInternet under www.WTW.com.Copyright© Weilheim 2005, WTW GmbHReprinting - even as excerpts - is only allowed with the explicit writtenauthorization of WTW GmbH, Weilheim.Printed in Germany.


TriOxmatic ® 700 IQ (SW)Contents1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-11.1 How to use this component operating manual . . . . . . . . 1-11.2 Structure of the TriOxmatic ® 700 IQ (SW) . . . . . . . . . . . 1-21.3 Recommended fields of application . . . . . . . . . . . . . . . . 1-22 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12.1 Authorized use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-22.2 General safety instructions . . . . . . . . . . . . . . . . . . . . . . . 2-23 Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13.1 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13.2 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13.3 Commissioning / Getting the instrument ready formeasuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-33.4 TriOxmatic ® 700 IQ (SW) setting table . . . . . . . . . . . . . 3-44 Measuring / Operation . . . . . . . . . . . . . . . . . . . . . . . . . 4-14.1 Measuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14.2 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14.2.1 General information on calibration . . . . . . . . . . 4-14.2.2 Calibration in water vapor-saturated air . . . . . . 4-24.3 Function check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-45 Maintenance, cleaning, disposal, and replacement . 5-15.1 General maintenance instructions . . . . . . . . . . . . . . . . . 5-15.2 Cleaning the sensor shaft and membrane . . . . . . . . . . . 5-15.3 Changing the electrolyte and membrane cap . . . . . . . . 5-45.4 Cleaning the electrodes . . . . . . . . . . . . . . . . . . . . . . . . . 5-95.4.1 Cleaning the gold working electrode . . . . . . . . 5-105.4.2 Cleaning the silver counter electrode . . . . . . . 5-115.5 Checking the sensor for freedom from zero-current . . 5-185.6 Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-195.7 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-195.8 Maintenance equipment and replacement parts . . . . . 5-206 What to do if... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-17 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-17.1 General data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-17.2 Electrical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-27.3 Measurement conditions . . . . . . . . . . . . . . . . . . . . . . . . 7-37.4 Characteristic data on delivery . . . . . . . . . . . . . . . . . . . . 7-48 Indexes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-10 - 1


ContentsTriOxmatic ® 700 IQ (SW)8.1 Explanation of the messages . . . . . . . . . . . . . . . . . . . . .8-18.1.1 Error messages . . . . . . . . . . . . . . . . . . . . . . . . .8-18.1.2 Info messages . . . . . . . . . . . . . . . . . . . . . . . . . .8-28.2 Status info . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-20 - 2


TriOxmatic ® 700 IQ (SW)Overview1 Overview1.1 How to use this component operating manualStructure of theIQ SENSOR NEToperating manualFig. 1-1Structure of the IQ SENSOR NET operating manualThe IQ SENSOR NET operating manual has a modular structure like theIQ SENSOR NET system itself. It consists of a system operating manualand the operating manuals of all the components used.Please file these component operating manuals into the ring binder ofthe system operating manual.1 - 1


OverviewTriOxmatic ® 700 IQ (SW)1.2 Structure of the TriOxmatic ® 700 IQ (SW)123 45Fig. 1-267 8Structure of the D. O. sensor (example: TriOxmatic ® 700 IQ)1 Protective hood2 WP 600 membrane cap3 Electrode unit4 ShaftElectrode unit:5 Gold working electrode (cathode)6 Insulator7 Silver counter electrode (anode)8 Reference electrode1.3 Recommended fields of applicationTriOxmatic 700 IQTriOxmatic 700 IQ SWStationary measurements in water/wastewater applications.Stationary measurements in seawater, aquaculture.1 - 2


TriOxmatic ® 700 IQ (SW)Safety2 SafetyThis component operating manual contains special instructions thatmust be followed during the operation of the TriOxmatic ® 700 IQ (SW)D.O. sensor. Thus, it is essential to read this component operatingmanual before carrying out any work using this sensor. In addition tothis manual, the SAFETY chapter of the IQ SENSOR NET systemoperating manual must be followed.Always keep this component operating manual together with thesystem operating manual and any other component operating manualsin the vicinity of the IQ SENSOR NET system.Special userqualificationsGeneral safetyinstructionsThe D.O. sensor was developed for applications in onlinemeasurement - essentially in wastewater treatment plants. Thus, weassume that the operators are familiar with the necessary precautionsto take when dealing with chemicals as a result of their professionaltraining and experience.The individual chapters of this operating manual use the followingsafety labels to indicate different levels of danger:Warningindicates instructions that must be followed precisely in order toprevent serious dangers to persons.Cautionindicates instructions that must be followed precisely in order toavoid slight injuries or damage to the instrument or theenvironment.Other labelsNoteindicates notes that draw your attention to special features.Noteindicates cross-references to other documents, e.g. operatingmanuals.2 - 1


SafetyTriOxmatic ® 700 IQ (SW)2.1 Authorized useThe authorized use of the TriOxmatic ® 700 IQ (SW) comprises its useas a D.O. sensor in the IQ SENSOR NET.The technical specifications according to chapter 7 TECHNICAL DATAmust be observed. Only operation according to the instructions in thisoperating manual is authorized.Any other use is considered to be unauthorized. Unauthorized useinvalidates any claims with regard to the guarantee.CautionOnly connect and operate the sensor together with IQ SENSOR NETaccessories.2.2 General safety instructionsThe sensor left the factory in a safe and secure technical condition.Function andoperational safetyThe failure-free function and operational safety of the sensor is onlyguaranteed if the generally applicable safety measures and the specialsafety instructions in this operating manual are followed during its use.The failure-free function and operational safety of the sensor is onlyguaranteed under the environmental conditions that are specified inchapter 7 TECHNICAL DATA.The specified temperature (chapter 7 TECHNICAL DATA) must bemaintained during the operation and transport of the sensor. Protectthe sensor, particularly against frost or overheating.CautionThe sensor may only be opened by specialists authorized byWTW. The only exceptions to this are the activities described inchapter 5 MAINTENANCE, CLEANING, DISPOSAL, AND REPLACEMENT.2 - 2


TriOxmatic ® 700 IQ (SW)SafetySafe operationIf safe operation is no longer possible, the sensor must be taken out ofoperation and secured against inadvertent operation.Safe operation is no longer possible if the sensor:• has been damaged in transport• has been stored under adverse conditions for a lengthy period oftime• is visibly damaged• no longer operates as described in this manual.If you are in any doubt, contact the supplier of your sensor.Obligations of theoperatorThe operator of the sensor must ensure that the following rules andregulations are followed when dealing with hazardous substances:• EEC directives for protective labor legislation• National protective labor legislation• Safety regulations• Safety data sheets of the chemical manufacturer.2 - 3


SafetyTriOxmatic ® 700 IQ (SW)2 - 4


CommissioningTriOxmatic ® 700 IQ (SW)Connecting the sensorto the sensorconnection cable1 Take the protective caps off the plug connections of the sensorandSACIQ sensor connection cable and keep them safe.2 Plug the jack of the SACIQ sensor connection cable onto theplug head connector of the sensor. At the same time, rotate thesocket so that the pin in the plug head connector (1) clicks intoone of the two holes in the jack.3 Then, screw the coupling ring (2) of the sensor connectioncable onto the sensor up to the stop.SACIQ21Fig. 3-1Connecting the sensor3 - 2


TriOxmatic ® 700 IQ (SW)Commissioning3.3 Commissioning / Getting the instrument ready formeasuring1 Pull the protective cap off the protective hood of the sensor.2 Leave the sensor to lie in the air for at least 60 minutes while itis switched on (polarization).NoteFor the sensor to polarize, the following conditions must be fulfilled:• The sensor is connected with the IQ SENSOR NET system via theSACIQ sensor connection cable.• The IQ SENSOR NET system is in operation• The sensor has been identified by the IQ SENSOR NET system.3 If required, assign a user-defined name to the sensor (seerelevant IQ SENSOR NET system operating manual).4 Set the sensor (see section 3.4).5 Calibrate the sensor (see section 4.2 CALIBRATION).NoteBoth sensor variants, TriOxmatic 700 IQ and TriOxmatic 700 IQ SW,run under the designation TriOxmatic700IQ in the IQ SENSOR NETsoftware.PrecisionmeasurementsRecommendation: In order to carry out precision measurements, leavethe sensor to polarize for a longer period of time, e.g. overnight, andrecalibrate it the following day.3 - 3


CommissioningTriOxmatic ® 700 IQ (SW)3.4 TriOxmatic ® 700 IQ (SW) setting tableSetting Selection/values ExplanationMeasuring modeMeasuring rangeConcentrationMeasuring rangeSaturation• Concentration• Saturation• 0 ... 60.0 mg/l• 0 ... 60.0 ppm• 0 ... 600 %Unit of the measured value on the measuredvalue display.These measuring ranges are available forselection.The measuring range is set permanently.Temperature mode • °C• °FUnit of the measured temperature value(Celsius, Fahrenheit).Salinity• On• OffDetermines whether the entered salinityshould be taken into account.Salinity input 2.0 ... 70.0 The entry of the salinity enables a saltcontent correction that compensates for theeffect of salt contents > 0.1 % on the oxygenmeasurement.The salt content correction is recommendedfor measurements in salt-contaminatedwastewater (salinity ≥ 2.0 corresponding to aconductivity of ≥ 3.4 mS/cm at a referencetemperature T REF = 20 °C).Temp. adjustment -1.5 K ... +1.5 K The temperature compensation enables thetolerances of the temperature sensor to bebalanced (shifting of the zero point by± 1.5 K).Notes:• Due to the thermal capacity of the sensor,it is necessary to place it in a containerwith at least 2 liters of water.• Leave the sensor in this container for atleast 15 minutes while stirringoccasionally until the balancing can beperformed.For temperature differences between thewater and sensor > 10 °C, leave thesensor for at least 1 hour in this containerwhile stirring occasionally until thebalancing can be performed.3 - 4


TriOxmatic ® 700 IQ (SW)CommissioningCarrying out settingsUsing s, switch from the measured value display to the main menu ofthe settings. Then navigate to the setting menu (setting table) of thesensor. The exact procedure is given in the relevant IQ SENSOR NETsystem operating manual.NoteFor information on oxygen measurement in solutions that contain salt,see WTW application report no. 1193118.3 - 5


CommissioningTriOxmatic ® 700 IQ (SW)3 - 6


TriOxmatic ® 700 IQ (SW)Measuring / Operation4 Measuring / Operation4.1 MeasuringFor measuring submerse the operable sensor in the test sample. Themeasured value is available immediately on submersing.4.2 Calibration4.2.1 General information on calibrationWhy calibrate?When to calibrate?Calibration procedure,During the operation of a D.O. sensor, the slope of the D.O. sensorchanges with time. The calibration procedure determines the currentslope of the sensor.Calibrate before measuring and at regular intervals (depending on theapplication).Ideally, calibration takes place in water vapor-saturated air. To do so,position the sensor approx. 2 cm above a water surface, for example ina narrow bucket or similar container with water.In the case of air temperatures under 5 °C we recommend to calibratenot in air but in air-saturated water that has a higher temperature. Youobtain air-saturated water by pouring water several times in and out oftwo vessels so that it sparkles.NoteFor calibration the membrane always has to be clean, for calibration inair it has to be dry. Clean contaminated sensors before calibration (seesection 5.2 CLEANING THE SENSOR SHAFT AND MEMBRANE).Calibration record /calibration historyMaintenance conditionThe result of a calibration is stored in the calibration record andcalibration history respectively and can be viewed afterwards (seerespective IQ SENSOR NET system operating manual).During calibration the sensor is in the so-called maintenance condition.This means all linked outputs retain their momentary state. Afterfinishing calibration the maintenance condition has to be switched offmanually. For more detailed information on the maintenance conditionplease refer to the respective IQ SENSOR NET system operatingmanual.4 - 1


Measuring / OperationTriOxmatic ® 700 IQ (SW)General course of acalibration on theIQ SENSOR NETGenerally, calibration on the IQ SENSOR NET is carried out as follows.System specific details are given in the respective IQ SENSOR NETsystem operating manual.1 Switch to the measured value display with m and select thesensor to be calibrated.2 Call up calibration with c.The next step switches on the maintenance condition for thesensor. A corresponding note appears on the display.3 Confirm the note with g.The maintenance condition is active.The menu-guided calibration routine starts.Follow the instructions on the display.After the calibration routine is finished, the measured valuedisplay appears again (the measured value flashes becausethe sensor is still in the maintenance condition).4 If the calibration was successful, bring the sensor into themeasuring position.5 Wait for a stable measured value.6 Switch off the maintenance condition.4.2.2 Calibration in water vapor-saturated air1 If necessary, clean the sensor and membrane and dry themembrane (see section 5.2 CLEANING THE SENSOR SHAFT ANDMEMBRANE).2 Start calibration and follow the instruction on the display untilthe instruction Bring sensor into calibrating position appears.Calibrating position2cm4 - 2


TriOxmatic ® 700 IQ (SW)Measuring / Operation3 Bring the sensor into the calibrating position.To do so, position the sensor approx. 2 cm above a watersurface, best in a narrow bucket or similar container with water.When doing so, no liquid should be brought on the membrane.NoteUp to this point, you can break off the calibration procedure at any timewith the e key. The system continues to work with the old calibrationdata. However, you have to switch off the maintenance condition in anycase.After starting the determination of calibration data with the g key (step4), you cannot break off the calibration any more.4 Continue with g.The calibration data is automatically determined anddocumented in the calibration record.5 Complete the calibration procedure according to theinstructions on the display.NoteIn the case of air temperatures under 5 °C do not calibrate the sensorin air but in air-saturated water that has a higher temperature. Youobtain air-saturated water by pouring water several times in and out oftwo vessels so that it sparkles.Then immerse the sensor in a vessel with air-saturated water(calibrating position) and calibrate as described above.Possible results of thecalibrationThe calibration data are evaluated by the system. A calibration canhave the following results:Display after thecalibrationMeasured value displayLog book entries(meaning/actions)Sensor was successfully calibrated.For the calibration data, see the calibrationhistory."----" Sensor could not be calibrated.Sensor blocked for measurement.– Service the sensor immediately(see operating manual).– View the calibration history.– Check the calibration conditions andcalibration standard.4 - 3


Measuring / OperationTriOxmatic ® 700 IQ (SW)4.3 Function checkThe function check is the simplest way of letting you know whether thesensor needs to be cleaned and calibrated.NoteThe function check can either be performed in water vapor-saturatedair or in air-saturated water.In the case of air temperatures under 5 °C we recommend to performthe function check not in air but in air-saturated water that has a highertemperature.NoteThe function check is carried out in the measuring mode, % saturation.When switching to the measuring mode % saturation, information onlinked outputs is lost.Function check in watervapor-saturated airProceed as follows:1 Take the sensor out of the measuring solution and position itapprox. 2 cm above a water surface, best in a narrow bucket orsimilar container with water. When doing so, no liquid shouldbe brought on the membrane.2cm2 Wait at least 15 minutes for the temperature to adapt.3 In the sensor setting table, select the measuring modeSaturation (unit %) and switch to the measured value displaywith m.4 Read the measured value and determine whether it lies withinthe range of precision required by the user.NoteIf there are large temperature differences between the sample and theair above the surface of the water, a balancing period of more than15 minutes may be required.4 - 4


TriOxmatic ® 700 IQ (SW)Measuring / OperationMeasured values withinthe required precisionMeasured valuesoutside the requiredprecisionIf the measured value lies within the range of precision required by theuser, no cleaning or recalibrating is necessary.Example:Required precision 5 %.Measured value in air: 97 % saturation (nominal value: 100%)⇒ No cleaning or recalibration is necessary as the measurement errorlies within the range of precision required by the user.If the measured value lies outside the range of precision required by theuser, clean the sensor shaft and membrane (see section 5.2) andcalibrate the sensor (see section 4.2).NoteIn the case of air temperatures under 5 °C do not perform the functioncheck in air but in air-saturated water that has a higher temperature.You obtain air-saturated water by pouring water several times in andout of two vessels so that it sparkles.Then submerse the sensor in a vessel with air-saturated water andperform the function check as described above.4 - 5


Measuring / OperationTriOxmatic ® 700 IQ (SW)4 - 6


TriOxmatic ® 700 IQ (SW)Maintenance, cleaning, disposal, and replacement5 Maintenance, cleaning, disposal, andreplacement5.1 General maintenance instructionsMaintenance conditionWe recommend to switch on the maintenance condition each timebefore removing the sensor from its measuring position. This avoidsunintended reactions of linked outputs. For more detailed informationon the maintenance condition please refer to the respectiveIQ SENSOR NET system operating manual.5.2 Cleaning the sensor shaft and membraneFor normal operation (e.g. municipal wastewater), cleaning andcalibration are strongly recommended:• in the case of pollution (according to visual check)• in the case of suspected incorrect measured values (usually too low)• if the measured value lies outside the range of precision required bythe user during the function check.Cleaning agentsContamination Cleaning agentsFor sludge and looselyadhering dirt, or biologicaldepositsSoft cloth or soft sponge, warmtapwater with detergentSalt and / or lime deposits Acetic acid (volume percentage =20 %),soft cloth or soft spongeNoteWe recommend not to unscrew the sensor from the sensor connectioncable while cleaning the sensor shaft and membrane. Otherwise,moisture and/or dirt can get into the plug connection where they cancause contact problems.If you need to disconnect the sensor from the sensor connection cable,please note the following points:• Before disconnecting the sensor from the SACIQ sensor connectioncable, remove any larger pieces of contamination from the sensor,particularly in the area of the plug connection (brush it off in a bucketof tap water, wash it off with a hose or wipe it off with a cloth).• Unscrew the sensor from the SACIQ sensor connection cable.5 - 1


Maintenance, cleaning, disposal, and replacementTriOxmatic ® 700 IQ (SW)• Place a protective cap on the sensor plug head connector and onthe SACIQ sensor connection cable each time so that no moistureor dirt can get on the contacting surfaces.• Connect the sensor with the SACIQ sensor connection cable againbefore polarization.Cleaning1 Pull the sensor out of the sample.2 Get rid of any coarse contamination on the sensor (e.g. brushit off in a bucket of tapwater, wash it down with a hose or washit off with a cloth).3 Unscrew the protective hood from the sensor.CautionAcetic acid irritates the eyes and the skin. When handling aceticacid, always wear protective gloves and protective goggles.4 Carefully clean the sensor shaft and membrane as specifiedin the CLEANING AGENTS section.CautionClean the membrane very carefully!Make sure the membrane is not damaged, e.g. by sharp-edgedobjects, by laying it on sharp stones, etc.5 - 2


TriOxmatic ® 700 IQ (SW)Maintenance, cleaning, disposal, and replacement5 Then, rinse it thoroughly with tapwater.6 Also clean and rinse the protective hood.7 Screw the protective hood back on again.8 If necessary, carefully dry the membrane with a lint-free papertowel.9 Calibrate the sensor (see section 4.2 CALIBRATION).5 - 3


Maintenance, cleaning, disposal, and replacementTriOxmatic ® 700 IQ (SW)5.3 Changing the electrolyte and membrane capWTW delivers the sensor ready for operation. The electrolyte solutionand membrane cap must only be replaced if:• the membrane is heavily contaminated and a calibration erroroccurs (log book message)• the membrane is damaged(log book message)• the electrolyte solution is depleted or the counter electrode ispoisoned (log book message)• after the gold working electrode and silver counter electrode havebeen cleanedNoteInformation on the contents and structure of the log book and how youcan call it up is given in the LOG BOOK chapter of the IQ SENSOR NETsystem operating manual.NoteUnscrew the membrane cap for maintenance purposes only. Afterdoing so, always use a new membrane cap!WarningContact with the sample can lead to danger to the user!Depending on the type of sample, suitable protective measuresmust be taken (protective clothing, protective goggles, etc.).CautionThe ELY/A electrolyte solution irritates eyes, skin and mucusmembranes. If it comes into contact with the eyes, rinsethoroughly with water and consult a doctor! During workingactivities, always wear suitable protective gloves and protectivegoggles/face shield! Follow the safety datasheet.5 - 4


TriOxmatic ® 700 IQ (SW)Maintenance, cleaning, disposal, and replacementNoteWe do not recommend to unscrew the sensor from the sensorconnection cable when changing the electrolyte solution or membranecap. Otherwise, moisture and/or dirt can get into the plug connection.If you would like to disconnect the sensor from the sensor connectioncable, please note the following points:• Before disconnecting the sensor from the SACIQ sensor connectioncable, remove any larger pieces of contamination from the sensor,particularly in the area of the plug connection (brush it off in a bucketof tap water, wash it off with a hose or wipe it off with a cloth).• Unscrew the sensor from the SACIQ sensor connection cable.• Place a protective cap on the sensor plug head connector and onthe SACIQ sensor connection cable each time so that no moistureor dirt can get on the contacting surfaces.• Connect the sensor with the SACIQ sensor connection cable againbefore polarization.Changing the electrolyteand membrane cap1 Pull the sensor out of the sample.2 Get rid of any coarse contamination on the sensor (e.g. brushit off in a bucket of tapwater, wash it down with a hose or washit off with a cloth).3 Unscrew the protective hood from the sensor.CautionThe ELY/A electrolyte solution irritates eyes, skin and mucusmembranes. If it comes into contact with the eyes, rinsethoroughly with water and consult a doctor! During workingactivities, always wear suitable protective gloves and protectivegoggles/face shield! Follow the safety datasheet.4 Unscrew the membrane cap (for disposal ofmembrane cap and electrolyte solution, see section 5.7).5 - 5


Maintenance, cleaning, disposal, and replacementTriOxmatic ® 700 IQ (SW)5 Rinse the sensor head with tapwater.6 Carefully rub the silver counter electrode with a paper toweland rinse it with deionized water.7 Thoroughly rinse the sensor head with electrolyte solution.5 - 6


TriOxmatic ® 700 IQ (SW)Maintenance, cleaning, disposal, and replacement8 Fill a new WP 600 membrane cap with ELY/A electrolytesolution(see section 5.8 MAINTENANCE EQUIPMENT AND REPLACEMENTPARTS).9 Throw away the first filling and fill the membrane cap oncemore with electrolyte solution.10 Remove any air bubbles by carefully tapping the membranecap.CautionThe ELY/A electrolyte solution irritates eyes, skin and mucusmembranes. If it comes into contact with the eyes, rinsethoroughly with water and consult a doctor! During working5 - 7


Maintenance, cleaning, disposal, and replacementTriOxmatic ® 700 IQ (SW)activities, always wear suitable protective gloves and protectivegoggles/face shield! Follow the safety datasheet.11 Screw the membrane cap onto the shaft while holding thesensor at an angle. Excess electrolyte solution is forced out ofthe ventilation area.Ventilation area (hold upward)12 The filling should be free of air bubbles as far as possible.However, small air bubbles do not cause any interference.13 After approx. 60 minutes polarization time, the sensor is readyfor operation.14 Recalibrate the sensor.PrecisionmeasurementsRecommendation: In order to carry out precision measurements, leavethe sensor to polarize for a longer period of time, e.g. overnight, andrecalibrate it the following day.NoteIn the following cases, you must refill the sensor once more:• If there are large air bubbles• If there are air bubbles on the gold working electrode• For measurements at high water pressure, even if small air bubblesare present. Otherwise, the membrane and, as a result, themeasuring characteristics could change.• If the sensor cannot be calibrated.5 - 8


TriOxmatic ® 700 IQ (SW)Maintenance, cleaning, disposal, and replacement5.4 Cleaning the electrodesThe counter electrode and reference electrode always show acoloration. This is necessary for the operation of the sensor.This is not contamination. Cleaning is only required in cases of slopesthat are too small or too large (the sensor cannot then be calibrated)that cannot be resolved by changing the membrane cap and electrolytesolution.CautionTo clean the electrodes, always unscrew the sensor from theSACIQ sensor connection cable. Otherwise, undesirableelectrochemical reactions can occur that may lead to thedestruction of the sensor.Preparatory activities1 Pull the sensor out of the sample and remove any coarsecontamination from the sensor (e.g. brush it off in a bucket oftapwater, wash it down with a hose or wash it off with a cloth).2 Unscrew the sensor from the SACIQ sensor connection cable.3 Place the protective cap on the sensor plug head connectorand on the SACIQ sensor connection cable each time so thatno moisture or dirt can get on the contacting surfaces.4 Unscrew the protective hood from the sensor.CautionThe ELY/A electrolyte solution irritates eyes, skin and mucusmembranes. If it comes into contact with the eyes, rinsethoroughly with water and consult a doctor! During workingactivities, always wear suitable protective gloves and protectivegoggles/face shield! Follow the safety datasheet.5 Unscrew the membrane cap (for disposal of membrane capand electrolyte solution, see section 5.7).5 - 9


Maintenance, cleaning, disposal, and replacementTriOxmatic ® 700 IQ (SW)6 Rinse the sensor head with tapwater.5.4.1 Cleaning the gold working electrode1 Moisten the gold working electrode and the SF 300 polishingstrip (see section 5.8 MAINTENANCE EQUIPMENT ANDREPLACEMENT PARTS) with deionized water.2 Using the rough side of the wet SF 300 polishing strip, polishoff any contamination from the gold working electrode usinglight pressure.CautionDo not use any conventional sandpaper or glass-fiber brushes.They could damage the electrode.5 - 10


TriOxmatic ® 700 IQ (SW)Maintenance, cleaning, disposal, and replacement3 Rinse the sensor head with deionized water.NoteCleaning the gold working electrode may already be sufficient to enablethe sensor to be calibrated again. For safety, however, we recommendto also clean the silver counter electrode (see section 5.4.2) and to usea new membrane cap afterwards.5.4.2 Cleaning the silver counter electrodeWe recommend to clean the silver counter electrode with the RA 600cleaning attachment that is available as an accessory (see section5.8 MAINTENANCE EQUIPMENT AND REPLACEMENT PARTS). This preventsthe reference electrode from coming into contact with the cleaningsolution. However, it is also possible to clean the silver counterelectrode in a beaker.CautionThe reference electrode must not come into contact with thecleaning solution under any circumstances. This could destroythe reference electrode and cause the sensor to becomedefective. However, no danger can arise if the RA 600 is usedcorrectly.5 - 11


Maintenance, cleaning, disposal, and replacementTriOxmatic ® 700 IQ (SW)Cleaning the silvercounter electrode withthe RA 600cleaning attachment1 Wipe the silver counter electrode with a lint free paper toweland carefully remove any loose deposits.2 Screw the cleaning attachment onto the sensor, instead of themembrane cap.CautionNever grease or lubricate the inner O-ring in the RA 600 cleaningattachment!3 Remove the screw cap from the safety cap of the cleaningattachment.5 - 12


TriOxmatic ® 700 IQ (SW)Maintenance, cleaning, disposal, and replacement4 Fill the screw cap with RL-AG/Oxi cleaning solution (seesection 5.8 MAINTENANCE EQUIPMENT AND REPLACEMENTPARTS).5 Screw the sensor with the safety cap onto the screw cap.6 Leave the cleaning solution for a maximum of 1 hour to takeeffect.7 Unscrew the screw cap.5 - 13


Maintenance, cleaning, disposal, and replacementTriOxmatic ® 700 IQ (SW)8 Thoroughly rinse the electrode unit with the safety cap on withdeionized water.9 Unscrew the safety cap.10 Rinse the sensor head and electrode unit several times withdeionized water.5 - 14


TriOxmatic ® 700 IQ (SW)Maintenance, cleaning, disposal, and replacement11 Water the sensor head and electrode unit in deionized waterfor at least an hour.12 Carefully shake off the drops of water.13 Fill a new WP 600 membrane cap and screw it on (see section5.3 CHANGING THE ELECTROLYTE AND MEMBRANE CAP).14 Connect the sensor to the sensor connection cable.15 After approx. 60 minutes, the sensor is ready for operation.16 Recalibrate the sensor.5 - 15


Maintenance, cleaning, disposal, and replacementTriOxmatic ® 700 IQ (SW)Cleaning the silvercounter electrode in abeaker1 Wipe the silver counter electrode with a lint free paper toweland carefully remove any loose deposits.2 Fill a beaker (150 ml, high-sided form) with approx. 25 ml RL-AG/Oxi cleaning solution.CautionThe reference electrode must not come into contact with thecleaning solution under any circumstances. This could destroythe reference electrode and cause the sensor to becomedefective.3 Clamp the sensor in a stand.Submerse the electrode unit in the RL-AG/Oxi cleaningsolution to just above the silver counter electrode.5 - 16


TriOxmatic ® 700 IQ (SW)Maintenance, cleaning, disposal, and replacement4 Leave the cleaning solution to work for 1 hour.5 Rinse the sensor head and electrode unit several times withdeionized water.6 Water the sensor head and electrode unit for at least an hourin deionized water.7 Carefully shake off the drops of water.8 Fill a new WP 600 membrane cap and screw it on (see section5.3 CHANGING THE ELECTROLYTE AND MEMBRANE CAP).9 Connect the sensor to the sensor connection cable.5 - 17


Maintenance, cleaning, disposal, and replacementTriOxmatic ® 700 IQ (SW)10 After approx. 60 minutes polarization time, the sensor is readyfor operation.11 Recalibrate the sensor.5.5 Checking the sensor for freedom from zero-currentThe sensor is zero-current free. Therefore, calibration of the sensor inoxygen-free solution is not provided.However, it is possible to check the freedom from zero-currentaccording to DIN EN 25814/ISO 5814. This is advisable if there is anysuspicion of a malfunction.Test solution 1 g/l aqueous sodium sulfite solution, Na 2 SO 3(addition of 1 mg/l of a cobalt(II) salt accelerates the removal of oxygenfrom the solution.)The sensor should be in operation for at least 1 hour before theinspection.Test procedure1 Clean the sensor (see section 5.2 CLEANING THE SENSOR SHAFTAND MEMBRANE).2 In the sensor setting table, select the measuring modeSaturation (unit %) and switch to the measured value displaywith m.3 Submerse the sensor in the test solution.4 Leave the sensor in the test solution for an hour.5 Read the measured value.Test criterionThe sensor is OK if < 2 % oxygen saturation is displayed as themeasured value.If the measured value is > 2 %:• Clean the gold working electrode (see section 5.4.1 CLEANING THEGOLD WORKING ELECTRODE) and change the electrolyte andmembrane cap (see section 5.3 CHANGING THE ELECTROLYTE ANDMEMBRANE CAP).• If necessary, also clean the silver counter electrode (see section5.4.2 CLEANING THE SILVER COUNTER ELECTRODE).5 - 18


TriOxmatic ® 700 IQ (SW)Maintenance, cleaning, disposal, and replacement5.6 StorageStore the clean and dry sensor with its protective cap in place.Observe the allowed temperature range (see chapter 7 TECHNICALDATA); the sensor may be stored in any position.5.7 DisposalWe recommend to dispose of the sensor without the membrane cap aselectronic refuse.CautionThe ELY/A electrolyte solution irritates eyes, skin and mucusmembranes. If it comes into contact with the eyes, rinsethoroughly with water and consult a doctor! During workingactivities, always wear suitable protective gloves and protectivegoggles/face shield! Follow the safety datasheet.Disposal of themembrane cap1 Before disposing of the membrane cap, unscrew it and rinsewith water.2 Dispose of the membrane cap in the household refuse.NoteTo dispose of the chemicals, follow the corresponding safetydatasheets. The safety datasheets can be obtained from WTW.5 - 19


Maintenance, cleaning, disposal, and replacementTriOxmatic ® 700 IQ (SW)5.8 Maintenance equipment and replacement partsDescription Model Order no.Accessory kit(2 replacement membrane caps,50 ml electrolyte solution, 50 mlcleaning solution, polishing film)Replacement membrane caps(2 pieces)Electrolyte solution(1 bottle x 50 ml)Cleaning solution forsilver counter electrode(1 bottle x 50 ml)ZBK 600 202 620WP 600/2 202 850ELY/A 205 212RL-Ag/Oxi 205 200Polishing film SF 300 203 680Cleaning attachment for cleaning thecounter electrodeRA 600 202 510NoteInformation on other IQ SENSOR NET accessories is given in the WTWcatalog and on the Internet.5 - 20


TriOxmatic ® 700 IQ (SW)What to do if...6 What to do if...The sensor is in the airand the display shows0.0 mg/l or 0% O 2Cause– No electrolyte in the membranecapRemedy– Change the WP 600membrane cap (see section5.3)The sensor cannot becalibratedCauseRemedy– Contaminated membrane cap – Clean the outside of thesensor (according tooperating manual), wait atleast 15 minutes andrecalibrate it.– In the case of contaminationthat cannot be removed:Exchange the membranecap and electrolyte.If the sensor still cannotbe calibrated afterchanging the electrolyteand membrane capCause– Contaminated electrodes orsensor toxificationRemedy– Clean the electrodes (seesection 5.4)Mechanical damage tothe sensorCauseRemedy– Return the sensorMeasured values toolowCause– Membrane contaminated– Sensor has not been calibratedfor a long time– Membrane does not fit snugly onthe gold working electrodeRemedy– Clean the outside of thesensor, then recalibrate (seesection 5.2 and section 4.2)– Change the membrane cap,then recalibrate (see section5.3 and section 4.2)6 - 1


What to do if...TriOxmatic ® 700 IQ (SW)Measured valuefluctuating heavilyCauseRemedy– Membrane cap loose – Screw the membrane captight– Membrane does not fit snugly onthe gold working electrode– Change the membrane cap,then recalibrate (see section5.3 and section 4.2)Measured values toohighCause– Sensor not yet completelypolarized– Sensor has not been calibratedfor a long timeRemedy– Wait for completepolarization (at least 1 hour)– Clean the outside of thesensor, then recalibrate(section 5.2 and section 4.2)Incorrect temperaturedisplayCauseRemedy– Temperature sensor defective – Return the sensor6 - 2


TriOxmatic ® 700 IQ (SW)Technical data7 Technical data7.1 General dataMeasuring principleMembrane-covered amperometric sensor with potentiostaticallyoperated 3-electrode system;Integrated microprocessor electronics, shielded 2-wire connection forpower and data transmission.DimensionsWeight (without sensorconnection cable)TriOxmatic 700 IQTriOxmatic 700 IQ SWapprox. 660 gapprox. 1170 gMembraneMaterialFluorine plasticThickness 50 µmElectrolyteELY/ATemperaturemeasurementVia integrated NTC, - 5 °C ... + 60 °CAccuracy± 0.5 KResolution0.1 K7 - 1


Technical dataTriOxmatic ® 700 IQ (SW)TemperaturecompensationIMT (integrated membrane temperature compensation)MaterialShaft:– TriOxmatic 700 IQTriOxmatic 700 IQ SWMembrane capMembraneSensor headInsulatorPlug head connector housingPlug, 3-poleProtective hoodVA steel 1.4571Polyoxymethylene (POM)Polyoxymethylene (POM)Fluorine plasticPolyoxymethylene (POM)Epoxy, polyetheretherketone (PEEK)Polyoxymethylene (POM)ETFE (blue) Tefzel®Polyoxymethylene (POM)Instrument safetyApplicable norms – EN 61010-1– UL 3111-1– CAN/CSA C22.2 No. 1010.17.2 Electrical dataNominal voltagePower consumptionProtective classMax. 24VDCvia the IQ SENSOR NET (for moredetails, see chapter TECHNICAL DATAof the IQ SENSOR NET systemoperating manual)0.2 WIII7 - 2


TriOxmatic ® 700 IQ (SW)Technical data7.3 Measurement conditionsMeasuring rangeD. O. concentration 0 ... 60 mg/lD. O. saturation 0 ... 600 %(≈ 0 ... 1200 mbar pO 2 )ResolutionD. O. concentration 0.1 mg/lD. O. saturation 1 %Measurement in water According to solubility function DIN 408 / Part 22Measurement in saltcontaminatedwastewaterSalinity input from 2.0 ... 70.0;corresponds to 3.4 mS/cm ... 86.2 mS/cm at T REF 20 °C(salinity measurement in the range 2 ... 42 according to IOT =International Oceanographic Tables)Polarization timeIn the case of re-commissioning or change ofelectrolyte:In the case of short polarization interruptions(depending on the duration of the interruption):At least60 minutes15 to 60 minutesTemperature rangeMeasuring medium 0 °C ... + 60 °CStorage/transport - 5 °C ... + 65 °CCalibration proceduresAir calibration (water calibration with air temperatures under 5 °C) withAutoRead proceduret min = 60 s; t max = 900 sAir pressurecompensationManuallyAutomaticallyEnter air pressure or altitudeOnly with the option of air pressuremeasurement of the MIQ/MC-Amodule7 - 3


Technical dataTriOxmatic ® 700 IQ (SW)Automatic sensormonitoring(SensCheck function)SensReg(electrolyte solution depleted)TriOxmatic700 IQYesTriOxmatic700 IQ SWYesSensLeck(membrane cap leaking)Yes No ** Note: The leak recognition can be selected in the setting table of theSensor monitoring relay function but the function is not active.Pressure resistanceSensor including connection cable:Max. allowed overpressure10 6 Pa (10 bar)The sensor meets all requirements according to article 3(3) of 97/23/EG ("pressure equipment directive").Type of protectionSensor including SACIQsensor connection cableSensor plug head connectorwithout sensor connectioncableIP 68, 10 barIP 67Depth of immersionmin. 10 cm; max. 100 m depthConnection techniqueOperating positionApproach flowConnection via SACIQ sensor connection cableAny, if the minimum approach flow is guaranteed≥ 5 cm/s (at 1 % measurement accuracy)7.4 Characteristic data on deliveryZero signal< 0.2 % of the saturation valueResponse time at 25 °Ct 90 (90 % of the final value display) after: < 180 sOwn consumption 0.0059 µg h -1 (mg/l) - 1 at 20 °CDriftWorking life< 1 % per month during permanent polarizationapprox. 5 years per electrolyte filling(theoretical electrolyte reserve for operation under air saturation)7 - 4


TriOxmatic ® 700 IQ (SW)Indexes8 Indexes8.1 Explanation of the messagesThis chapter contains a list of all the message codes and relatedmessage texts that can occur in the log book of theIQ SENSOR NETTriOxmatic ® 700 IQ (SW) system for the sensor.NoteInformation on the contents and structure of the log book and how youcan call it up is given in the LOG BOOKIQ SENSOR NET chapter of thesystem operating manual.8.1.1 Error messagesMessage codeEA13xxEA23xxEA33xxEC43xxEC53xxESB33xMessage textMeas. range exceeded or undercut* Check process* Select other meas. rangeSensor temperature too high!* Check process and applicationSensor temperature too low!* Check process and applicationSensor could not be calibrated,sensor blocked for measurementCause: instable signal* Check polarization time and temperature adjustment* Check calibration conditions* Read calibration history* Repeat calibration, if unsuccessful: service the sensor (seeoperating manual)Sensor could not be calibrated,sensor blocked for measurementCause: slope error* Check polarization time and temperature adjustment* Check calibration conditions* Read calibration history* Repeat calibration, if unsuccessful: service the sensor (seeoperating manual)SensReg: Electrolyte supply is depleted* Change electrolyte solution and membrane head immediately (seeoperating manual)8 - 1


IndexesTriOxmatic ® 700 IQ (SW)Message codeESC33xES13xxMessage textSensLeck: Membrane head damaged* Change electrolyte solution and membrane head immediately (seeoperating manual)Note: This information is transmitted with the TriOxmatic 700 IQ only.Component hardware defective* Contact WTW8.1.2 Info messagesMessage codeIC13xxIS133xMessage textSensor has been successfully calibrated* For calibration data, see calibration historySensReg: Electrolyte supply is almost depleted* Change electrolyte solution and membrane head as soon aspossible (see operating manual)8.2 Status infoThe status info is a coded piece of information on the current status ofa sensor. Each sensor sends this status info to the controller. Thestatus info of sensors consists of 32 bits, each of which can have thevalue 0 or 1.Status info,general structure0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 151 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 (general)0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 (internal16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31The bits 0 - 15 are reserved for general information.The bits 16 - 21 are reserved for internal service information.8 - 2


TriOxmatic ® 700 IQ (SW)IndexesYou obtain the status info:• via a manual query in the Einstellungen/Settings/Service/ menu Listof all components (see system operating manual)• via an automated query– from a superordinate process control (e. g. when connected tothe Profibus)– from the IQ Data Server (see IQ SENSOR NET Software Packoperating manual)NoteThe evaluation of the status info, e.g. in the case of an automatedquery, has to be made individually for each bit.Status infoTriOxmatic ® 700 IQ (SW)Status bitBit 0Bit 1Bit 2 *ExplanationComponent hardware defectiveSensReg: Electrolyte supply is depletedSensLeck: Membrane head damagedBit 3-31 -* Note:This information is transmitted with the TriOxmatic 700 IQ only.8 - 3


IndexesTriOxmatic ® 700 IQ (SW)8 - 4