SensoLyt ® 700 (SW)Accuracy when going topressThe use of advanced technology and the high quality standard of ourinstruments are the result of continuous development. This may resultin differences between this operating manual and your instrument.Also, we cannot guarantee that there are absolutely no errors in thismanual. Therefore, we are sure you will understand that we cannotaccept any legal claims resulting from the data, figures or descriptions.NoteThe latest version of the present operating manual can be found on theInternet under www.WTW.com.Copyright© Weilheim 2005, WTW GmbHReprinting - even as excerpts - is only allowed with the explicit writtenauthorization of WTW GmbH, Weilheim.Printed in Germany.26


SensoLyt ® 700 (SW)ContentsSensoLyt ® 700 (SW) - Contents1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 281.1 Structure of the SensoLyt ® 700 (SW) pH/ORP armature 281.2 Recommended fields of application . . . . . . . . . . . . . . . . 282 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 292.1 General information on safety . . . . . . . . . . . . . . . . . . . . 292.2 Authorized use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 302.3 User qualifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 302.4 General safety instructions . . . . . . . . . . . . . . . . . . . . . . . 303 Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 323.1 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 323.2 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 323.3 Commissioning / Getting the instrument ready formeasuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 334 Measuring / Operation . . . . . . . . . . . . . . . . . . . . . . . . . . 364.1 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 364.2 Measuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 365 Maintenance and changing the electrode . . . . . . . . . . 375.1 General maintenance instructions . . . . . . . . . . . . . . . . . 375.2 Replacing the combination electrode . . . . . . . . . . . . . . . 375.3 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 415.4 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 416 Replacement parts and accessories . . . . . . . . . . . . . . 426.1 Combination electrodes . . . . . . . . . . . . . . . . . . . . . . . . . 426.2 General accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . 427 What to do if... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 438 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 448.1 Measurement characteristics . . . . . . . . . . . . . . . . . . . . . 448.2 Application characteristics . . . . . . . . . . . . . . . . . . . . . . . 448.3 General data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 468.4 Electrical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48ba15307e04 12/200527


OverviewSensoLyt ® 700 (SW)1 Overview1.1 Structure of the SensoLyt ® 700 (SW) pH/ORParmature1 2 3 4 5Fig. 1-1 Structure of the pH/ORP armature (example, SensoLyt ® 700)1 Protective hood2 Temperature sensor3 Electrode (not contained in the scope of delivery)4 Electrode receptacle5 Armature shaftNoteThe pH combination electrodes that can be used are available asaccessories (see chapter 6 REPLACEMENT PARTS AND ACCESSORIES).Impedance converterGlass breakagemonitoringAn impedance converter integrated in the armature shaft converts thehigh-impedance measurement signal of the electrode into a lowimpedancesignal. This ensures an interference-free signaltransmission even over greater distances.The sensor is equipped with a SensCheck function for monitoring glassbreakage.1.2 Recommended fields of applicationSensoLyt 700SensoLyt 700 SWIn conjunction with the SensoLyt ® SEA, SensoLyt ® DWA andSensoLyt ® ECA pH combination electrodes as well as the SensoLyt ®PtA ORP combination electrode, the SensoLyt ® 700 (SW) pH/ORParmature is suitable for stationary pH or ORP measurement in thefollowing fields:Stationary measurements in water/wastewater applications.Stationary measurements in seawater and brackish water,aquaculture.28 ba15307e04 12/2005


SensoLyt ® 700 (SW)Safety2 Safety2.1 General information on safetyThese safety instructions contain all instructions that have to befollowed for a safe operation of the SensoLyt ® 700 (SW) pH/ORParmature. Before starting any work with the SensoLyt ® 700 (SW),carefully read the safety instructions and strictly follow all protectivemeasures mentioned.Always keep this operating manual in the vicinity of the armature.General safetyinstructionsSafety instructions in this operating manual can be recognized by thewarning symbol (triangle) in the left column. The signal word (e. g."Caution") indicates the level of the danger:Warningindicates instructions that must be followed precisely in order toprevent serious dangers to persons.Cautionindicates instructions that must be followed precisely in order toavoid slight injuries or damage to the instrument or theenvironment.Other labelsNoteindicates notes that draw your attention to special features.Noteindicates cross-references to other documents, e.g. operatingmanuals.ba15307e04 12/200529


SafetySensoLyt ® 700 (SW)2.2 Authorized useThe authorized use of the SensoLyt ® 700 (SW) consists of thestationary measurement of pH or ORP and temperature in conjunctionwith a pH or ORP combination electrode. The technical specificationsaccording to chapter 8 TECHNICAL DATA must be observed. Onlyoperation according to the instructions given in this operating manual isconsidered to be authorized.Any other use is considered to be unauthorized. Unauthorized useinvalidates any claims with regard to the guarantee.2.3 User qualificationsCalibrating the SensoLyt ® 700 (SW) requires the handling ofchemicals. Thus, we assume that the relevant personnel are familiarwith the necessary precautions to take when dealing with chemicals asa result of their professional training and experience.2.4 General safety instructionsFunction andoperational safetyThe armature left the factory in a safe and secure technical condition.The failure-free function and operational safety of the armature is onlyguaranteed if the generally applicable safety measures and the specialsafety instructions in this operating manual are followed during its use.The smooth functioning and operational safety of the armature can onlybe guaranteed under the environmental conditions that are specified inchapter 8 TECHNICAL DATA.The specified temperature (chapter 8 TECHNICAL DATA) must bemaintained during the application and transport of the armature.Protect the armature, particularly against frost or overheating.30 ba15307e04 12/2005


SensoLyt ® 700 (SW)SafetySafe operationIf safe operation is no longer possible, the armature must be taken outof operation and secured against inadvertent operation.Safe operation is no longer possible if the armature:• has been damaged in transport• has been stored under adverse conditions for a lengthy period oftime• is visibly damaged• no longer operates as described in this manual.If you are in any doubt, contact the supplier of your armature.Obligations of theoperatorThe operator of the armature must ensure that the following rules andregulations are followed when dealing with hazardous substances:• EEC directives for protective labor legislation• National protective labor legislation• Safety regulations• Safety data sheets of the chemical manufacturer.CautionAll changes of the SensoLyt ® 700 (SW) that exceed the workdescribed in this operating manual are not allowed. Repair workmay only be carried out by WTW Weilheim.ba15307e04 12/200531


CommissioningSensoLyt ® 700 (SW)3 Commissioning3.1 Scope of delivery• SensoLyt ® 700 (SW) pH/ORP armatureThe armature is equipped with a protective hood and protectioncaps• Operating manual.3.2 InstallationCautionThe SensoLyt ® 700 (SW) pH/ORP armature may only besubmersed in conjunction with a mounted combination electrode.When exchanging the electrode, no moisture may get into thearmature, as otherwise the armature will be destroyed. Whichelectrodes can be used together with the SensoLyt ® 700 (SW) pH/ORP armature is given in section 6.1 COMBINATION ELECTRODES.Connection to themeasuring transmitterThe connection cable of the SensoLyt ® 700 (SW) is ready to beconnected to the terminal strip of a measuring transmitter with highimpedancepH/ORP input. For all further information please refer to theoperating manual of the measuring transmitter. The assignment of thecable wires of the SensoLyt ® 700 (SW) can be found in the chapter 8TECHNICAL DATA of this operating manual.NoteDo not suspend the sensor on the sensor connection cable. Use anarmature or electrode holder. Information on this and otheraccessories is given in the WTW catalog and on the Internet.32 ba15307e04 12/2005


SensoLyt ® 700 (SW)Commissioning3.3 Commissioning / Getting the instrument ready formeasuringNoteA KCI-filled plastic cap is mounted on the tip of the sensor to keep theelectrode active during storage (or during longer pauses in measuring).The cap must be removed for measuring.Mounting thecombination electrode1 Unscrew the protective hood from the armature.2 Pull off the blind plug from the plug head socket of thearmature.3 Screw off the protective cap of the plug head connector of theelectrode.ba15307e04 12/200533


CommissioningSensoLyt ® 700 (SW)4 Screw the electrode into the plug head socket of the armature.5 Push the unit into the armature up to the stop.CautionPush the connected electrode into the armature right up to thestop so that the connection is watertight.6 Pull the KCI-filled plastic cap off the electrode for measuring.34 ba15307e04 12/2005


SensoLyt ® 700 (SW)Commissioning7 Screw the protective hood onto the armature.8 Make the settings for the electrode on the measuringtransmitter. Subsequently, calibrate the measuring system(see section 4.1 CALIBRATION).ba15307e04 12/200535


Measuring / OperationSensoLyt ® 700 (SW)4 Measuring / OperationNoteCalibrate the measuring system after the initial commissioning and atregular intervals (depending on the application).4.1 CalibrationWhy calibrate?When to calibrate?Calibration proceduresDuring the operation of a pH electrode, the slope and asymmetry of theelectrode changes with time. The calibration procedure determines thecurrent slope and asymmetry of the electrode.Calibrate before measuring and at regular intervals (depending on theapplication).The available calibration procedures depend on the measuringtransmitter used. The individual steps for the calibration arecomprehensively described in the operating manual of the measuringtransmitter.4.2 MeasuringWarningContact with the sample can lead to danger to the user!Depending on the type of sample, suitable protective measuresmust be taken (protective clothing, protective goggles, etc.).Please pay attention to:• the minimum immersion depth of the armature (> 40 mm)• the measuring range of the electrode used (see operating manual ofthe electrode).36 ba15307e04 12/2005


SensoLyt ® 700 (SW)Maintenance and changing the electrode5 Maintenance and changing the electrode5.1 General maintenance instructionsThe SensoLyt ® 700 (SW) pH/ORP armature works maintenance-free.Please read the maintenance of the electrode in the relevant operatingmanual of the electrode.5.2 Replacing the combination electrodeWarningContact with the sample can lead to danger to the user!Depending on the type of sample, suitable protective measuresmust be taken (protective clothing, protective goggles, etc.).CautionIf the glass of the pH electrode breaks, there is a danger of cutsfrom the splinters of glass!If it is necessary to replace an electrode, proceed as follows:1 Unscrew the protective hood from the armature.ba15307e04 12/200537


Maintenance and changing the electrodeSensoLyt ® 700 (SW)2 Use the protective hood as a tool to lever out the electrode.3 Carefully pull out the electrode until the plug head screwedfitting can be seen.4 Unscrew the combination electrode from the plug head socket(for disposal, see section 5.4).38 ba15307e04 12/2005


SensoLyt ® 700 (SW)Maintenance and changing the electrode5 Screw in a new electrode.6 Push the unit into the armature up to the stop.CautionPush the connected electrode into the armature right up to thestop so that the connection is watertight.7 Pull the KCI-filled plastic cap off the electrode for measuring.8 Screw the protective hood onto the armature.ba15307e04 12/200539


Maintenance and changing the electrodeSensoLyt ® 700 (SW)9 Calibrate the measuring system (see section 4.1 CALIBRATION).40 ba15307e04 12/2005


SensoLyt ® 700 (SW)Maintenance and changing the electrode5.3 CleaningFor normal operation (e.g. municipal wastewater), exterior cleaningand calibration are strongly recommended:• in the case of pollution (according to visual check)• if there is a suspicion of erroneous measured values• if the measured value lies outside the range of precision required bythe user during the function check.Exterior cleaningContaminationFor sludge and loosely adheringdirt, or biological depositsSalt and / or lime depositsCleaning agentsSoft cloth or soft sponge, warmtapwater with detergentAcetic acid (volume percentage= 20 %),soft cloth or soft spongeNotePlease also observe the instructions on cleaning the electrode in therespective operating manual.5.4 DisposalArmatureCombination electrodesWe recommend disposing of the armature as electronic refuse.If no official regulations apply to the contrary, used and defectiveelectrodes can be treated as household waste.ba15307e04 12/200541


Replacement parts and accessoriesSensoLyt ® 700 (SW)6 Replacement parts and accessories6.1 Combination electrodesCautionThe pressure resistance of the operable pH/ORP armature can berestricted by the pressure resistance of the electrode (see chapter8 TECHNICAL DATA). When selecting the electrode please make sureit is suitable for the pressure range and temperature rangeplanned.pH combinationelectrodesModelSensoLyt ® SEASensoLyt ® DWASensoLyt ® ECASensoLyt ® SEA-HPOrder no.109 115109 119109 117109 118ORP combinationelectrodeSensoLyt ® PtA 109 1256.2 General accessoriesTechnical buffersolutions forpH calibrationBuffer (bottles of 1 liter) pH value Order no.TEP 4TEP 7TEP 104,017,010,0108 700108 702108 704Protective hoodModelOrder no.SensoLyt ® 700 SK 109 194NoteInformation on further accessories is given in the WTW catalog and onthe Internet.42 ba15307e04 12/2005


SensoLyt ® 700 (SW)What to do if...7 What to do if...Measurement deliversno or wrong measuredvaluesCauseRemedy– Armature not connected – Check connection to meter– Electrode not connected ordefective– Check electrode andelectrode connection– Watering cap still on theelectrode– No or wrong calibrationperformed– Pull off watering cap andcalibrate– Calibrate– Electrode contaminated – Clean electrode– Liquid has penetrated thearmature– Armature defective, return toWTW– Instrument setting incorrect – Correct instrument settingSystem cannot becalibratedCauseRemedy– Slope of the electrode too low – Replace electrode– Asymmetry of the electrode toohigh– Armature is operated with ORPelectrode– Replace electrode– Use pH electrodeba15307e04 12/200543


Technical dataSensoLyt ® 700 (SW)8 Technical data8.1 Measurement characteristicsMeasuring principleMeasuring rangePotentiometric measurement using a combination electrode. Signalamplifier integrated in the sensor for low impedance signaltransmissionDepending on the built-in electrodeTemperaturemeasurementSensor accuracyResponse time± 0.3 Kt 99 (99 % of the final valuedisplay after) < 15 s8.2 Application characteristicsTemperature rangeMeasuring mediumStorage/transport0 °C ... + 60 °C (32 ... 140 °F)- 5 °C ... + 65 °C (23 ... 149 °F)Allowed pH range of thetest sample4 ... 12Pressure resistanceMax. allowed overpressure (sensor including connection cable):with installed combinationelectrode,SensoLyt ® SEA, DWA, PtAwith installed combinationelectrode,SensoLyt ® ECAwith installed combinationelectrode,SensoLyt ® SEA-HP10 6 Pa (10 bar) *6 x10 5 Pa (6 bar) *10 6 Pa (10 bar) *** temperature dependent (see safety instruction below)** in the entire temperature rangeThe SensoLyt ® 700 (SW) meets the requirements according to article3(3) of the directive, 97/23/EC ("pressure equipment directive").44 ba15307e04 12/2005


SensoLyt ® 700 (SW)Technical dataImmersion depthwith installed combinationelectrode,SensoLyt ® SEA, DWA, PtAwith installed combinationelectrode,SensoLyt ® ECAwith installed combinationelectrode,SensoLyt ® SEA-HPmin. 40 mm; max. 100 m * or tocable endmin. 40 mm; max. 60 m * or tocable endmin. 40 mm; max. 100 m ** or tocable end* temperature dependent (see safety instruction below)** in the entire temperature rangeCautionThe pressure resistance of the operable pH/ORP armature can bereduced by the pressure resistance of the combination electrode.When selecting the combination electrode make sure it is suitablefor the intended pressure and temperature range.Type of protectionArmature with integratedelectrode including connectioncableIP 68, 10 bar (10 6 Pa)Plug connector IP 65Operating positionAnyFields of applicationSensoLyt 700SensoLyt 700 SWStationary measurements inwater/wastewater applicationsStationary measurements inseawater and brackish water,aquacultureAutomatic sensormonitoring(SensCheck function)Function for glass breakage monitoring of the pH electrode by themeasuring transmitterba15307e04 12/200545


Technical dataSensoLyt ® 700 (SW)8.3 General dataDimensionsWeight (without sensorconnection cable andcombination electrode)SensoLyt 700SensoLyt 700 SWapprox. 320 gapprox. 880 gElectrodes that can beintegratedpH combination electrodesORP electrodesSensoLyt ® SEA, SEA-HP, DWA,ECASensoLyt ® PtAConnection techniqueConnection cable permanently mounted on the sensor.Connection to the measuring transmitter via 7-pole screw plug.46 ba15307e04 12/2005


SensoLyt ® 700 (SW)Technical dataMaterialShaftShaft screw:– SensoLyt 700SensoLyt 700 SWProtective hoodElectrode receptacleTemperature sensor:– SensoLyt 700SensoLyt 700 SWClosing headProtection ringCable gland:– SensoLyt 700SensoLyt 700 SWCable sheathPOMV4A stainless steel 1.4571Titan Grade 2PVCPOMV4A stainless steel 1.4571V4A stainless steel 1.4571, KTLcoatedPOMPOMV4A stainless steel 1.4571Titan Grade 2PURConnection cableLength:– SensoLyt 700SensoLyt 700 SWDiameterSmallest allowedbend radius7 m (special lengths on request)15 m8.6 mmPermanent bend: 130 mmShort time bend: 80 mmInstrument safetyApplicable norms – EN 61010-1– UL 3111-1– CAN/CSA C22.2 No. 1010.1ElectromagneticcompatibilityTest certificatesaccording to EN 61326 class B and FCC class AUL, cULba15307e04 12/200547


Technical dataSensoLyt ® 700 (SW)8.4 Electrical dataPin assignment Pin Assignment Voltage Max. current1Ua (output signal)------20 V------3Ub++ 10 V< 1 mA4Ub-- 10 V< 1 mA5NTC< 3.5 V< 0.15 mA6NTC< 3.5 V< 0.15 mA7---------Plug from the front:342751648 ba15307e04 12/2005


SensoLyt ® 700 (SW)Technical databa15307e04 12/200549


Wissenschaftlich-Technische Werkstätten GmbHDr.-Karl-Slevogt-Straße 1D-82362 WeilheimGermanyTel: +49 (0) 881 183-0+49 (0) 881 183-100Fax: +49 (0) 881 183-420E-Mail: Info@WTW.comInternet: http://www.WTW.com