Operating Instructions - VEGADIS 62 -
Operating Instructions - VEGADIS 62 -
Operating Instructions - VEGADIS 62 -
- No tags were found...
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
5 Connecting to power supply5.3 Wiring plan3( + ) 1 2(-) ( + ) 3 4(-)1 2Fig. 8: Wiring plan <strong>VEGADIS</strong> <strong>62</strong>1 To the sensor2 For power supply3 Coupling for the connection cable to the indicating and adjustment module5.4 Installation HART standardThe following illustrations show in a simplified way the use of<strong>VEGADIS</strong> <strong>62</strong> in conjunction with a HART sensor. The HARTcommunication resistance in the signal cable is always required for alow impedance power supply. It is absolutely necessary to install itbetween voltage supply and the <strong>VEGADIS</strong> <strong>62</strong>.Note:If the instrument is powered via a VEGAMET signal conditioninginstrument, then the communication resistor is already included in thesignal conditioning instrument. A communication resistor in the signalcable must not be installed.Overview=~1234Fig. 9: <strong>VEGADIS</strong> <strong>62</strong> in conjunction with an individual sensor1 Sensor2 <strong>VEGADIS</strong> <strong>62</strong>3 HART resistor > 150 Ω (with low impedance power supply necessary)4 Voltage supply/Processing16 <strong>VEGADIS</strong> <strong>62</strong>36469-EN-100514