Operating manualSensoLyt ® 700 IQSensoLyt ® 700 IQ SWSensoLyt 700 IQSensoLyt 700 IQ SWIQ SENSOR NET pH/ORP sensorba15324e06 12/2005


SensoLyt ® 700 IQ (SW)Accuracy when going topressThe use of advanced technology and the high quality standard of ourinstruments are the result of continuous development. This may resultin differences between this operating manual and your instrument.Also, we cannot guarantee that there are absolutely no errors in thismanual. Therefore, we are sure you will understand that we cannotaccept any legal claims resulting from the data, figures or descriptions.NoteThe latest version of the present operating manual can be found on theInternet under www.WTW.com.Copyright© Weilheim 2005, WTW GmbHReprinting - even as excerpts - is only allowed with the explicit writtenauthorization of WTW GmbH, Weilheim.Printed in Germany.ba15324e06 12/2005


SensoLyt ® 700 IQ (SW)ContentsSensoLyt ® 700 IQ (SW) - Contents1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-11.1 How to use this component operating manual . . . . . . . . 1-11.2 Structure of the SensoLyt ® 700 IQ (SW) . . . . . . . . . . . . 1-21.3 Recommended fields of application . . . . . . . . . . . . . . . . 1-22 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12.1 Authorized use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-22.2 General safety instructions . . . . . . . . . . . . . . . . . . . . . . . 2-23 Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13.1 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13.2 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13.3 Commissioning / Getting the instrument ready formeasuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-33.4 SensoLyt ® 700 IQ (SW) setting table . . . . . . . . . . . . . . . 3-64 Measuring / Operation . . . . . . . . . . . . . . . . . . . . . . . . . 4-14.1 Measuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14.2 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14.2.1 General information on calibration . . . . . . . . . . 4-14.2.2 Calibration with CAL TEC AUTO . . . . . . . . . . . 4-34.2.3 Calibration with CAL CON 2P . . . . . . . . . . . . . . 4-44.2.4 Calibration with CAL CON 1P . . . . . . . . . . . . . . 4-54.2.5 Calibration result . . . . . . . . . . . . . . . . . . . . . . . . 4-65 Maintenance and changing the electrode . . . . . . . . . 5-15.1 General maintenance instructions . . . . . . . . . . . . . . . . . 5-15.2 Replacing the combination electrode . . . . . . . . . . . . . . . 5-25.3 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-46 Replacement parts and accessories . . . . . . . . . . . . . 6-16.1 Combination electrodes . . . . . . . . . . . . . . . . . . . . . . . . . 6-16.2 General accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-17 What to do if... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-18 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-18.1 Measurement characteristics . . . . . . . . . . . . . . . . . . . . . 8-1ba15324e06 12/20050 - 1


ContentsSensoLyt ® 700 IQ (SW)8.2 Application characteristics . . . . . . . . . . . . . . . . . . . . . . .8-18.3 General data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-38.4 Electrical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-49 Indexes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-19.1 Explanation of the messages . . . . . . . . . . . . . . . . . . . . .9-19.1.1 Error messages . . . . . . . . . . . . . . . . . . . . . . . . .9-19.1.2 Info messages . . . . . . . . . . . . . . . . . . . . . . . . . .9-19.2 Status info . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-20 - 2 ba15324e06 12/2005


SensoLyt ® 700 IQ (SW)Overview1 Overview1.1 How to use this component operating manualStructure of theIQ SENSOR NET operatingmanualFig. 1-1Structure of the IQ SENSOR NET operating manualThe IQ SENSOR NET operating manual has a modular structure like theIQ SENSOR NET system itself. It consists of a system operating manualand the operating manuals of all the components used.Please file this component operating manual in the ring binder of thesystem operating manual.ba15324e06 12/20051 - 1


OverviewSensoLyt ® 700 IQ (SW)1.2 Structure of the SensoLyt ® 700 IQ (SW)1 2 3 4 5Fig. 1-2Structure of the pH/ORP sensor (example, SensoLyt ® 700 IQ)1 Protective hood2 Temperature sensor3 Combination electrode (not contained in the scope of delivery)4 Electrode holder5 Sensor shaftNoteThe pH combination electrodes that can be used are available asaccessories (see chapter 6 REPLACEMENT PARTS AND ACCESSORIES).Screening of the pH/ORP sensorGlass breakagemonitoringThe combination electrode and the SensoLyt ® 700 IQ pH/ORP sensortogether with the IQ SENSOR NET system form a measuring system thatis protected to a high degree against low and high frequencyinterference as well as against the indirect effects of lightning strikes.The sensor is equipped with a SensCheck function for monitoring glassbreakage.1.3 Recommended fields of applicationIn conjunction with the SensoLyt ® SEA, SensoLyt ® DWA andSensoLyt ® ‚ ECA pH combination electrodes as well as the SensoLyt ®PtA ORP combination electrode, the SensoLyt ® 700 IQ pH/ORPsensor is suitable for stationary pH or ORP measurement in thefollowing ranges.SensoLyt 700 IQSensoLyt 700 IQ SWStationary measurements in water/wastewater applications.Stationary measurements in seawater and brackish water,aquaculture.1 - 2 ba15324e06 12/2005


SensoLyt ® 700 IQ (SW)Safety2 SafetyThis component operating manual contains special instructions thatmust be followed in the operation of the SensoLyt ® 700 IQ (SW) pH/ORP sensor. Thus, it is essential to read this component operatingmanual before carrying out any work using this sensor. In addition tothis manual, the SAFETY chapter of the IQ SENSOR NET systemoperating manual must be followed.Always keep this component operating manual together with thesystem operating manual and any other component operating manualsin the vicinity of the IQ SENSOR NET system.Special userqualificationsGeneral safetyinstructionsThe pH/ORP sensor was developed for applications in onlinemeasurement - essentially in the field of wastewater treatment. Thus,we assume that the operators are familiar with the necessaryprecautions to take when dealing with chemicals as a result of theirprofessional training and experience.Safety instructions in this operating manual can be recognized by thewarning symbol (triangle) in the left column. The signal word (e. g."Caution") indicates the level of the danger:Warningindicates instructions that must be followed precisely in order toprevent serious dangers to persons.Cautionindicates instructions that must be followed precisely in order toavoid slight injuries or damage to the instrument or theenvironment.Other labelsNoteindicates notes that draw your attention to special features.Noteindicates cross-references to other documents, e.g. operatingmanuals.ba15324e06 12/20052 - 1


SafetySensoLyt ® 700 IQ (SW)2.1 Authorized useThe authorized use of the SensoLyt ® 700 IQ (SW) comprises its use asa pH/ORP sensor together with a pH combination electrode or ORPcombination electrode in the IQ SENSOR NET.The technical specifications according to chapter 8 TECHNICAL DATAmust be observed. Only operation according to the instructions in thisoperating manual is authorized.Any other use is considered to be unauthorized. Unauthorized useinvalidates any claims with regard to the guarantee.CautionOnly connect and operate the sensor together with IQ SENSOR NETaccessories.2.2 General safety instructionsThe sensor left the factory in a safe and secure technical condition.Function andoperational safetyThe failure-free function and operational safety of thesensor is only guaranteed if the generally applicable safety measuresand the special safety instructions in this operating manual are followedduring its use.The failure-free function and operational safety of the sensor is onlyguaranteed under the environmental conditions that are specified inchapter 8 TECHNICAL DATA.The specified temperature (chapter 8 TECHNICAL DATA) must bemaintained during the operation and transport of the sensor. Protectthe sensor, particularly against frost or overheating.CautionThe sensor may only be opened by specialists authorized byWTW.2 - 2 ba15324e06 12/2005


SensoLyt ® 700 IQ (SW)SafetySafe operationIf safe operation is no longer possible, the sensor must be taken out ofoperation and secured against inadvertent operation.Safe operation is no longer possible if the sensor:• has been damaged in transport• has been stored under adverse conditions for a lengthy period oftime• is visibly damaged• no longer operates as described in this manual.If you are in any doubt, contact the supplier of your sensor.Obligations of theoperatorThe operator of the sensor must ensure that the following rules andregulations are followed when dealing with hazardous substances:• EEC directives for protective labor legislation• National protective labor legislation• Safety regulations• Safety data sheets of the chemical manufacturer.ba15324e06 12/20052 - 3


SafetySensoLyt ® 700 IQ (SW)2 - 4 ba15324e06 12/2005


SensoLyt ® 700 IQ (SW)Commissioning3 Commissioning3.1 Scope of delivery• SensoLyt ® 700 IQ (SW)• The sensor is fitted with a protective hood and protective caps• Operating manual.3.2 InstallationConnection cableA sensor connection cable of the SACIQ or SACIQ SW type is requiredto connect the sensor. The cable is available in different lengths.Compared to the standard model SACIQ, the SACIQ SW sensorconnection cable is optimized regarding its corrosion resistance inseawater and brackish water and adapted for use in conjunction withthe SensoLyt ® 700 IQ SW. Information on this and otherIQ SENSOR NET accessories is given in the WTW catalog and on theInternet.NoteHow to connect the SACIQ (SW) sensor connection cable to theterminal strip of an MIQ module is described in chapter 3 INSTALLATIONof the IQ SENSOR NET system operating manual.CautionThe SensoLyt ® 700 IQ pH/ORP sensor unit may only be immersedin conjunction with a mounted combination electrode. Moisturemust be prevented from penetrating the pH/ORP sensor duringthe replacement of the electrode as, otherwise, the sensor couldbe destroyed. Which electrodes can be used in conjunction withthe SensoLyt ® 700 IQ pH/ORP sensor unit is given in section6.1 COMBINATION ELECTRODES.Are the plugconnections dry?Before connecting the sensor and sensor connection cable, pleasemake sure that the plug connections are dry. If moisture gets into theplug connections, first dry the plug connections (dab them dry or blowthem dry using compressed air).NoteDo not suspend the sensor on the sensor connection cable. Use anarmature or electrode holder. Information on this and other IQ SENSORNET accessories is given in the WTW catalog and on the Internet.ba15324e06 12/20053 - 1


CommissioningSensoLyt ® 700 IQ (SW)Connecting the sensorto the sensorconnection cable1 Take the protective caps off the plug connections of the sensorand the SACIQ (SW) sensor connection cable and keep themsafe.2 Plug the jack of the SACIQ (SW) sensor connection cable ontothe plug head connector of the sensor. At the same time, rotatethe socket so that the pin in the plug head connector (1) clicksinto one of the two holes in the jack.3 Then, screw the coupling ring (2) of the sensor connectioncable onto the sensor up to the stop.SACIQ21Fig. 3-1Connecting the sensor3 - 2 ba15324e06 12/2005


SensoLyt ® 700 IQ (SW)Commissioning3.3 Commissioning / Getting the instrument ready formeasuringNoteA KCI-filled plastic cap is mounted on the tip of the sensor to keep thecombination electrode active during storage (or during longer pausesin measuring). The cap must be removed for measuring.Mounting thecombination electrode1 Unscrew the protective hood from the sensor.2 Pull off the blind plug from the plug head socket of the sensor.3 Screw off the protective cap of the plug head connector of thecombination electrode.ba15324e06 12/20053 - 3


CommissioningSensoLyt ® 700 IQ (SW)4 Screw the combination electrode into the plug head socket ofthe sensor.5 Push the unit into the sensor up to the stop.CautionPush the connected combination electrode into the sensor rightup to the stop so that the connection is watertight. Leaks couldlead to the destruction of the sensor.6 Pull the KCI-filled plastic cap off the combination electrode formeasuring.3 - 4 ba15324e06 12/2005


SensoLyt ® 700 IQ (SW)Commissioning7 Screw the protective hood onto the sensor.8 Carry out the settings for the sensor on the terminal of themeasuring system (see section 3.4 SENSOLYT® 700 IQ (SW)SETTING TABLE).9 Calibrate the sensor and the combination electrode with themeasuring system (see section 4.2 CALIBRATION).ba15324e06 12/20053 - 5


CommissioningSensoLyt ® 700 IQ (SW)3.4 SensoLyt ® 700 IQ (SW) setting tableMenu item Possible settings ExplanationsMeasuring mode• mV• pHUnit of the measured values on themeasured value display.Temperature mode • °C• °FUnit of the measured temperature value(Celsius, Fahrenheit).Calibrationprocedure(only in pHmeasuring mode)• CAL TEC AUTO• CAL CON 2P• CAL CON 1P• Simplified 2-point calibration using anytwo different WTW technical buffersolutions. The nominal values of thebuffer solutions are stored in thesensor. This makes the manual entry ofthe nominal values redundant.• 2-point calibration using any twodifferent buffer solutions. The nominalvalues of the buffer solutions must beentered.• 1-point calibration using any singlebuffer solution. The nominal value ofthe buffer solution must be entered.ORP shift(only in mVmeasuring mode)Temperatureadjustment-100 mV ... +100 mV You can set the ORP zero point here.-1.5 K ... +1.5 K The temperature compensation functionenables the temperature sensor to bebalanced against a reference temperaturemeasurement (displacement of the zeropoint by ±1.5 K).Notes:• Due to the thermal capacity of thesensor, it is necessary to place it in acontainer with at least 2 liters of water.• Leave the sensor in this container for atleast 15 minutes, or in the case oftemperature differences between thewater and sensor > 10 K for at least 1hour, while stirring occasionally. Thencarry out the balancing procedure.3 - 6 ba15324e06 12/2005


SensoLyt ® 700 IQ (SW)CommissioningSave and quitQuitThe system confirms the saving of thesettings and the display switches to thenext higher level.The display switches to the next higherlevel without saving the new settings.Carrying out settingsUsing s, switch from the measured value display to the main menu ofthe settings. Then navigate to the setting menu (setting table) of thesensor. The exact procedure is given in the relevant IQ SENSOR NETsystem operating manual.ba15324e06 12/20053 - 7


CommissioningSensoLyt ® 700 IQ (SW)3 - 8 ba15324e06 12/2005


SensoLyt ® 700 IQ (SW)Measuring / Operation4 Measuring / Operation4.1 MeasuringWarningContact with the sample can lead to danger to the user!Depending on the type of sample, suitable protective measuresmust be taken (protective clothing, protective goggles, etc.).NoteCalibrate the combination electrode with the sensor and the measuringsystem before measuring and at regular intervals (depending on theapplication).NotePlease pay attention to:• the minimum immersion depth of the sensor (> 40 mm)• the measuring range of the electrode used (see operating manual ofthe electrode).4.2 Calibration4.2.1 General information on calibrationWhy calibrate?When to calibrate?Calibration procedureDuring the operation of a pH electrode, the slope and asymmetry of theelectrode changes with time. The calibration procedure determines thecurrent slope and asymmetry of the electrode.Calibrate before measuring and at regular intervals (depending on theapplication).The CAL TEC AUTO calibration procedure enables a fully automaticcalibration using WTW technical buffer solutions. Ordering informationon WTW technical buffer solutions is given in chapter 6 REPLACEMENTPARTS AND ACCESSORIES.The CAL CON 2P calibration procedure enables conventional twopointcalibration using any two different buffer solutions.The CAL CON 1P calibration procedure enables conventional singlepointcalibration with any single buffer solution.Calibration record /calibration historyThe result of a calibration is stored in the calibration record andcalibration history respectively and can be viewed afterwards (seerespective IQ SENSOR NET system operating manual).ba15324e06 12/20054 - 1


Measuring / OperationSensoLyt ® 700 IQ (SW)Maintenance conditionGeneral course of acalibration on theIQ SENSOR NETDuring calibration the sensor is in the so-called maintenance condition.This means all linked outputs retain their momentary state. Afterfinishing calibration the maintenance condition has to be switched offmanually. For more detailed information on the maintenance conditionplease refer to the respective IQ SENSOR NET system operatingmanual.Generally, calibration on the IQ SENSOR NET is carried out as follows.System specific details are given in the respective IQ SENSOR NETsystem operating manual.NoteBefore starting make sure the correct calibration procedure is set (seesection 3.4 SENSOLYT® 700 IQ (SW) SETTING TABLE).1 Switch to the measured value display with m and select thesensor to be calibrated.2 Call up calibration with c.The next step switches on the maintenance condition for thesensor. A corresponding note appears on the display.3 Confirm the note with g.The maintenance condition is active.The menu-guided calibration routine starts.Follow the instructions on the display.After the calibration routine is finished, the measured valuedisplay appears again (the measured value flashes becausethe sensor is still in the maintenance condition).4 If the calibration was successful, bring the sensor into themeasuring position.5 Wait for a stable measured value.6 Switch off the maintenance condition.4 - 2 ba15324e06 12/2005


SensoLyt ® 700 IQ (SW)Measuring / Operation4.2.2 Calibration with CAL TEC AUTOMessages displayed during calibration with CAL TEC AUTODisplay* Have any two WTW technicalbuffer solutions ready.* Rinse the sensor.* Immerse the sensor in the firstbuffer solution.* Wait for a stable measuredvalue.* Rinse the sensor.* Immerse the sensor in thesecond buffer solution.* Wait for a stable measuredvalue.Successfully calibrated.End of the CAL TEC AUTOcalibration.ExplanationYou can use any two differentWTW technical buffer solutions todo this.Confirm with g.Follow the instructions on thedisplay. As soon as a stablemeasured value is reached, thenext display appears.Follow the instructions on thedisplay. As soon as a stablemeasured value is reached, thenext display appears.The values determined for Slopeand Asymmetry potential aredisplayed.The calibration is complete.Confirm with g.The display returns to themeasured value display.ba15324e06 12/20054 - 3


Measuring / OperationSensoLyt ® 700 IQ (SW)4.2.3 Calibration with CAL CON 2PMessages displayed during two-point calibration with CAL CON 2PDisplay* Have any two buffer solutionsready.* Rinse the sensor.* Immerse the sensor in the firstbuffer solution.* Wait for a stable measuredvalue.* Enter the pH value of the firstbuffer solution.Confirm entered data andcontinue calibration* Rinse the sensor.* Immerse the sensor in thesecond buffer solution.* Wait for a stable measuredvalue.* Enter the pH value of thesecond buffer solution.Confirm entered data andcontinue calibrationSuccessfully calibrated.End of the CAL CON 2Pcalibration.ExplanationYou can use any two differentbuffer solutions the pH value ofwhich is known at the currenttemperature.Follow the instructions on thedisplay. As soon as a stablemeasured value is reached, thenext display appears.Enter the pH value with d.Confirm with g. Themeasurement of the buffersolution begins.Follow the instructions on thedisplay. As soon as a stablemeasured value is reached, thenext display appears.Enter the pH value with d.Confirm with g. Themeasurement of the buffersolution begins.The values determined for Slopeand Asymmetry potential aredisplayed.The calibration is complete.Confirm with g.The display returns to themeasured value display.4 - 4 ba15324e06 12/2005


SensoLyt ® 700 IQ (SW)Measuring / Operation4.2.4 Calibration with CAL CON 1PMessages displayed during the single-point calibration withCAL CON 1PDisplayExplanation* Have any buffer solution ready. You can use any buffer solutionthe pH value of which is known atthe current temperature.* Rinse the sensor.* Immerse the sensor in the buffersolution.* Wait for a stable measuredvalue.* Enter the pH value of the buffersolution.Confirm entered data andcontinue calibrationSuccessfully calibrated.End of the CAL CON 1Pcalibration.Follow the instructions on thedisplay. As soon as a stablemeasured value is reached, thenext display appears.Enter the pH value with d.Confirm with g. Themeasurement of the buffersolution begins.The values determined for Slopeand Asymmetry potential aredisplayed.The calibration is complete.Confirm with g.The display returns to themeasured value display.ba15324e06 12/20054 - 5


Measuring / OperationSensoLyt ® 700 IQ (SW)4.2.5 Calibration resultCalibration evaluationAfter calibrating the system automatically evaluates the calibration dataand current state of the sensor. The asymmetry and slope areevaluated separately. The values must be within the following ranges:Slope: -50 ... -62 mV/pHAsymmetry: -45 mV ... +45 mVIf one of both values is outside the specified range, the calibration isevaluated as unsuccessful, i. e. the sensor could not be calibrated.A calibration can have the following results:Possible results of thecalibrationDisplay after thecalibrationMeasured value displayLog book entries(meaning/actions)Sensor was successfully calibrated.For the calibration data, see the calibrationhistory."----" Sensor could not be calibrated.Sensor blocked for measurement.– Service the sensor immediately(see operating manual).– View the calibration history.– Check the calibration conditions andcalibration standard.NoteInformation on the contents and structure of the log book, and how tocall it up, is given in the LOG BOOK chapter of the IQ SENSOR NETsystem operating manual.4 - 6 ba15324e06 12/2005


SensoLyt ® 700 IQ (SW)Maintenance and changing the electrode5 Maintenance and changing the electrode5.1 General maintenance instructionsThe SensoLyt ® 700 IQ (SW) pH/ORP sensor operates maintenancefree.NotePlease read the maintenance of the combination electrode in therelevant operating manual of the electrode.WarningContact with the sample can lead to danger to the user!Depending on the type of sample, suitable protective measuresmust be taken (protective clothing, protective goggles, etc.).CautionIf the glass of the pH electrode breaks, there is a danger of cutsfrom the splinters of glass!NoteWe do not recommend unscrewing the sensor from the sensorconnection cable when changing the electrode. Otherwise, moistureand/or dirt can get into the plug connection where they can causecontact problems.If you would like to disconnect the sensor from the sensor connectioncable, please note the following points:• Before disconnecting the sensor from the SACIQ (SW) sensorconnection cable, remove any larger pieces of contamination fromthe sensor, particularly in the area of the plug connection (brush itoff in a bucket of tapwater, wash it off with a hose or wipe it off witha cloth).• Unscrew the sensor from the SACIQ (SW) sensor connection cable.• Always place a protective cap on the plug head of the sensor and onthe SACIQ (SW) sensor connection cable so that no moisture or dirtcan get into the contacting surfaces.• In corrosive environments close the dry socket of the sensorconnection cable with the SACIQ-Plug protective screw cap in orderto protect the electrical contacts from corrosion. The protective capis available as an accessory (see section 6.2 GENERALACCESSORIES). It is included in the standard scope of delivery of theSACIQ SW sensor connection cable.ba15324e06 12/20055 - 1


Maintenance and changing the electrodeSensoLyt ® 700 IQ (SW)5.2 Replacing the combination electrodeIf it is necessary to replace the combination electrode, proceed asfollows:1 Unscrew the protective hood from the sensor.2 Use the protective hood as a tool to lever out the combinationelectrode.3 Carefully pull out the combination electrode until the plug headscrewed fitting can be seen.5 - 2 ba15324e06 12/2005


SensoLyt ® 700 IQ (SW)Maintenance and changing the electrode4 Unscrew the combination combination electrode from the plughead socket (for disposal, see section 5.3).5 Screw in a new combination electrode.6 Push the unit into the sensor up to the stop.7 Pull the KCI-filled plastic cap off the combination electrode formeasuring.ba15324e06 12/20055 - 3


Maintenance and changing the electrodeSensoLyt ® 700 IQ (SW)8 Screw the protective hood onto the sensor.9 Calibrate the sensor and the electrode with the measuringsystem (see section 4.2 CALIBRATION).5.3 DisposalSensorCombination electrodesWe recommend disposing of the sensor as electronic refuse.If no official regulations apply to the contrary, used and defectivecombination electrodes can be treated as household waste.5 - 4 ba15324e06 12/2005


SensoLyt ® 700 IQ (SW)Replacement parts and accessories6 Replacement parts and accessories6.1 Combination electrodespH combinationelectrodesModelSensoLyt ® SEASensoLyt ® DWASensoLyt ® ECASensoLyt ® SEA-HPOrder no.109 115109 119109 117109 118ORP combinationelectrodeSensoLyt ® PtA 109 1256.2 General accessoriesTechnical buffersolutions forpH calibrationBuffer (bottles of 1 liter) pH value Order no.TEP 4TEP 7TEP 10TEP 10 Trace4.017.010.010.01108 700108 702108 704*108 703*** for SensoLyt ® 700 IQ (SW) to software version 2.02** for SensoLyt ® 700 IQ (SW) from software version 2.10Protective hoodModelOrder no.SensoLyt ® 700 SK 109 194Protective screw cap forsensor connection cableModelOrder no.SACIQ-Plug 480 065NoteInformation on other IQ SENSOR NET accessories is given in the WTWcatalog and on the Internet.ba15324e06 12/20056 - 1


Replacement parts and accessoriesSensoLyt ® 700 IQ (SW)6 - 2 ba15324e06 12/2005


SensoLyt ® 700 IQ (SW)What to do if...7 What to do if...No measured valueCauseRemedy– Sensor not connected – Connect the sensor– Unknown – Look in the log bookMeasurement does notfunctionCause– Watering cap still on theelectrodeRemedy– Pull off watering cap andcalibrate– Electrode not connected – Connect electrode– Liquid has penetrated thesensor– Sensor defective, send it back– Sensor not connected – Connect the sensor– Instrument setting incorrect – Correct instrument settingSystem cannot becalibratedCause– Slope of the electrode notwithin tolerance (see section4.2.5)Remedy– Condition the electrode– If the slope is still outside thetolerance: replace electrode– Slope of the electrode too low – Replace electrode– Asymmetry of the electrodetoo high– Sensor is operated with ORPelectrode– Replace electrode– Use pH electrodeba15324e06 12/20057 - 1


What to do if...SensoLyt ® 700 IQ (SW)Measurement providesimplausible measuredvaluesCauseRemedy– No calibration performed – Calibrate– Electrode not connected ordefective– Check electrode and electrodeconnection– Electrode contaminated – Clean electrode– Liquid has penetrated thesensor– Sensor defective, send it back– Instrument setting incorrect – Correct instrument setting(Measuring mode pH or mV)7 - 2 ba15324e06 12/2005


SensoLyt ® 700 IQ (SW)Technical data8 Technical data8.1 Measurement characteristicsMeasuring principlePotentiometric measurement using a combination electrode;Integrated microprocessor electronics, shielded 2-wire connection forpower and data transmission.Measuring rangepHORP0.00 ... 14.00 pH (depending on the electrode)-2000 mV ... +2000 mV(depending on the electrode)ResolutionpHORP0.01 pH1 mVTemperaturemeasurementTemperature sensorMeasuring rangeAccuracyResolutionResponse time t 99 of thetemperature sensorintegrated NTC- 5 °C ... + 60 °C (23 ... 140 °F)± 0.5 K0.1 K< 15 sTemperaturecompensationin the range 0 °C ... 60 °C (32 ... 140 °F)8.2 Application characteristicsTemperature rangeMeasuring mediumStorage/transport0 °C ... + 60 °C (32 ... 140 °F)- 5 °C ... + 65 °C (23 ... 149 °F)Allowed pH range of thetest sample4 ... 12ba15324e06 12/20058 - 1


Technical dataSensoLyt ® 700 IQ (SW)Pressure resistanceMax. allowed overpressure (sensor including connection cable):with installed combination electrode,SensoLyt ® SEA, DWA, PtAwith installed combination electrode,SensoLyt ® ECAwith installed combination electrode,SensoLyt ® SEA-HP10 6 Pa (10 bar) *6 x10 5 Pa (6 bar) *10 6 Pa (10 bar) *** temperature dependent (see safety instruction below)** in the entire temperature rangeThe SensoLyt ® 700 IQ (SW) meets the requirements according toarticle 3(3) of the directive, 97/23/EC ("pressure equipment directive").Immersion depthwith installed combination electrode,SensoLyt ® SEA, DWA, PtAwith installed combination electrode,SensoLyt ® ECAwith installed combination electrode,SensoLyt ® SEA-HPmin. 40 mm; max. 100 m *min. 40 mm; max. 60 m *min. 40 mm; max. 100 m *** temperature dependent (see safety instruction below)** in the entire temperature rangeCautionThe pressure resistance of the operable pH/ORP armature can bereduced by the pressure resistance of the combination electrode.When selecting the combination electrode make sure it is suitablefor the intended pressure and temperature range.Type of protectionSensor with installed combinationelectrode and including SACIQ (SW)sensor connection cableSensor plug head connector withoutsensor connection cable (sensor withinstalled combination electrode)IP 68, 10 bar (10 6 Pa)IP 67Operating positionanyFields of applicationSensoLyt 700 IQSensoLyt 700 IQ SWStationary measurements in water/wastewater applicationsStationary measurements in seawaterand brackish water, aquaculture8 - 2 ba15324e06 12/2005


SensoLyt ® 700 IQ (SW)Technical data8.3 General dataDimensionsWeight (without sensorconnection cable andcombination electrode)SensoLyt 700 IQSensoLyt 700 IQ SWapprox. 970 gapprox. 1800 gElectrodes that can beintegratedpH combination electrodesORP combination electrodesSensoLyt ® SEA, SEA-HP, DWA,ECASensoLyt ® PtAConnection techniqueConnection via the SACIQ or SACIQ SW sensor connection cableMaterialShaft– SensoLyt 700 IQSensoLyt 700 IQ SWProtective hoodElectrode holderTemperature sensor:– SensoLyt 700 IQSensoLyt 700 IQ SWV4A stainless steel 1.4571POMPVCPOMV4A stainless steel 1.4571V4A stainless steel 1.4571, KTLcoatedPlug head housingPOMPlug, 3-pole ETFE (blue) Tefzel ®ba15324e06 12/20058 - 3


Technical dataSensoLyt ® 700 IQ (SW)Automatic sensormonitoring(SensCheck function)Function for monitoring glass breakage in the pH electrodeInstrument safetyApplicable norms – EN 61010-1– UL 3111-1– CAN/CSA C22.2 No. 1010.18.4 Electrical dataNominal voltagePower consumptionProtective classMax. 24VDCvia the IQ SENSOR NET (for moredetails, see chapter TECHNICALDATA of the IQ SENSOR NETsystem operating manual)0.2 WIII8 - 4 ba15324e06 12/2005


SensoLyt ® 700 IQ (SW)Indexes9 Indexes9.1 Explanation of the messagesThis chapter contains a list of all the message codes and relatedmessage texts that can occur in the log book of the IQ SENSOR NETsystem for the SensoLyt ® 700 IQ (SW) sensor.NoteInformation on the contents and structure of the log book, and how tocall it up, is given in the LOG BOOK chapter of the IQ SENSOR NETsystem operating manual.9.1.1 Error messagesMessage codeEA13xxEA23xxEA33xxEC13xxESA31xES13xxMessage textMeas. range exceeded or undercut* Check process* Select other meas. rangeSensor temperature too high!* Check process and applicationSensor temperature too low!* Check process and applicationSensor could not be calibrated,Sensor blocked for measurement* Check calibration conditions and calibration standard* View calibration history* Service sensor immediately(see operating manual)SensCheck: pH electrode defective, glass broken* Replace pH electrodeComponent hardware defective* Contact WTW9.1.2 Info messagesMessage codeIC13xxMessage textSensor has been successfully calibrated* For calibration data, see calibration historyba15324e06 12/20059 - 1


IndexesSensoLyt ® 700 IQ (SW)9.2 Status infoThe status info is a coded piece of information on the current status ofa sensor. Each sensor sends this status info to the controller. Thestatus info of sensors consists of 32 bits, each of which can have thevalue 0 or 1.0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15Status info,general structure1 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 (general)0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 (internal16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31The bits 0 - 15 are reserved for general information.The bits 16 - 21 are reserved for internal service information.You obtain the status info:• via a manual query in the Einstellungen/Settings/Service/List of allcomponents menu (see system operating manual)• by an automated query– of a superordinate process control (e. g. when connected to theProfibus)– from the IQ Data Server (see IQ SENSOR NET Software Packoperating manual)NoteThe evaluation of the status info, e.g. in the case of an automatedquery, has to be made individually for each bit.Status infoSensoLyt ® 700 IQ (SW)Status bitBit 0Bit 1ExplanationComponent hardware defectiveSensCheck: pH electrode defective, glass brokenBit 2-31 -9 - 2 ba15324e06 12/2005