<strong>1066</strong> Instruction Manual Section 4: WiringLIQ_MAN_<strong>1066</strong> September 2013FIGURE 4-8. Chlorine, oxygen, ozone sensor wiring to <strong>1066</strong> printed circuit board (<strong>1066</strong>-CL, <strong>1066</strong>-DO,<strong>1066</strong>-OZ)HINGE SIDE OF FRONT PANELOUTPUT 2TB7+24V(OUTPUT1)SENSOR WIRINGLOOP PWR TB5 TB3 TB2+24VGND GNDTB6THUMANODECATHODERTNSNSRTN IN+V-VREFSHLDGNDSOLSHLDpHTB4TB1<strong>1066</strong> CIRCUIT BOARDASSY 24359- 00CHLORINE, OXYGEN, OZONE SENSOR WIRING(FOLLOW RECOMMENDED ORDER)1) TB5/ANODE& CATHODE2) TB3/RTDANODECATHODERETURNSENSERTD IN3) TB2/ SOLUTIONGROUNDNO CONNECTIONNO CONNECTIONSOLUTION GROUNDNOTE:A) TB1, TB4, TB6 AND TB7 NOT USED FOR OXYGEN AND OZONE SENSOR WIRING.B) TB1, TB2 AND TB4 MAY BE USED FOR pH SENSOR WIRING IF FREE CHLORINE MEASEMENT REQUIRESLIVE pH INPUT.<strong>1066</strong> AMPSENSOR WIRING40<strong>1066</strong>11 AWiring 23