Alert - 2 SeriesDIP SWITCH SETTINGS FOR LEFT & RIGHT TRANSDUCERHigh Pressure mode for Amico Model #: M2-X-MAN-07B (2500 Psi)Set Dip switches # 3 & 4 to the ON position.This will enable gas cylinders to be attached to the left bank of the manifold.Liquid Mode for Amico Model #: M2-X-MAN-07C (500 Psi)Set Dip switch # 3 to the OFF position.Set Dip switch # 4 to the ON position.This will enable liquid cylinders to be attached to the left bank of the manifoldDip Switch # 7 is used for the heater control leave ON position.Dip Switch # 8 should remain in the ON position and should never be tampered with.High Pressure mode for Amico Model #: M2-X-MAN-07B (2500 Psi)Set Dip switches # 5 & 6 to the ON position.This will enable gas cylinders to be attached to the right bank of the manifold.Liquid Mode for Amico Model #: M2-X-MAN-07C (500 Psi)Set Dip switch # 5 to the OFF position.Set Dip switch # 6 to the ON position.This will enable liquid cylinders to be attached to the right bank of the manifoldDip Switch # 7 is not used. Leave in the ON position.Dip Switch # 8 should remain in the ON position and should never be tampered with.Page: 19