18.10.2016 Views

UT Testing-Section 4 Calibration Methods

UT Testing-Section 4 Calibration Methods

UT Testing-Section 4 Calibration Methods

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>Section</strong> 4: <strong>Calibration</strong> <strong>Methods</strong>


Content: <strong>Section</strong> 4: <strong>Calibration</strong> <strong>Methods</strong><br />

4.1: <strong>Calibration</strong> <strong>Methods</strong><br />

4.2: The <strong>Calibration</strong>s<br />

4.2.1: Distance Amplitude Correction (DAC)<br />

4.2.2: Finding the probe index<br />

4.2.3: Checking the probe angle<br />

4.2.4: <strong>Calibration</strong> of shear waves for range V1 Block<br />

4.2.5: Dead Zone<br />

4.2.7: Transfer Correction<br />

4.2.8: Linearity Checks (Time Base/ Equipment Gain/ Vertical Gain)<br />

4.2.9: TCG-Time Correction Gain<br />

4.3: Curvature Correction<br />

4.4: <strong>Calibration</strong> References & Standards<br />

4.5: Questions & Answers<br />

4.6: Video Time


4.1: <strong>Calibration</strong> <strong>Methods</strong><br />

<strong>Calibration</strong> refers to the act of evaluating and adjusting the precision and<br />

accuracy of measurement equipment. In ultrasonic testing, several forms of<br />

calibration must occur. First, the electronics of the equipment must be<br />

calibrated to ensure that they are performing as designed. This operation is<br />

usually performed by the equipment manufacturer and will not be discussed<br />

further in this material. It is also usually necessary for the operator to perform<br />

a "user calibration" of the equipment. This user calibration is necessary<br />

because most ultrasonic equipment can be reconfigured for use in a large<br />

variety of applications. The user must "calibrate" the system, which includes<br />

the equipment settings, the transducer, and the test setup, to validate that the<br />

desired level of (1) precision and (2) accuracy are achieved. The term<br />

calibration standard is usually only used when an absolute value is measured<br />

and in many cases, the standards are traceable back to standards at the<br />

National Institute for Standards and Technology.


<strong>Calibration</strong>s


In ultrasonic testing, there is also a need for reference standards. Reference<br />

standards are used to establish a general level of consistency in<br />

measurements and to help interpret and quantify the information contained in<br />

the received signal. Reference standards are used to validate that the<br />

equipment and the setup provide similar results from one day to the next and<br />

that similar results are produced by different systems. Reference standards<br />

also help the inspector to estimate the size of flaws. In a pulse-echo type<br />

setup, signal strength depends on both the size of the flaw and the distance<br />

between the flaw and the transducer. The inspector can use a reference<br />

standard with an artificially induced flaw of known size and at approximately<br />

the same distance away for the transducer to produce a signal. By comparing<br />

the signal from the reference standard to that received from the actual flaw,<br />

the inspector can estimate the flaw size.


This section will discuss some of the more common calibration and reference<br />

specimen that are used in ultrasonic inspection. Some of these specimens<br />

are shown in the figure above. Be aware that there are other standards<br />

available and that specially designed standards may be required for many<br />

applications. The information provided here is intended to serve a general<br />

introduction to the standards and not to be instruction on the proper use of the<br />

standards.


Introduction to the Common Standards<br />

<strong>Calibration</strong> and reference standards for ultrasonic testing come in many<br />

shapes and sizes. The type of standard used is dependent on the NDE<br />

application and the form and shape of the object being evaluated. The<br />

material of the reference standard should be the same as the material being<br />

inspected and the artificially induced flaw should closely resemble that of the<br />

actual flaw. This second requirement is a major limitation of most standard<br />

reference samples. Most use drilled holes and notches that do not closely<br />

represent real flaws. In most cases the artificially induced defects in reference<br />

standards are better reflectors of sound energy (due to their flatter and<br />

smoother surfaces) and produce indications that are larger than those that a<br />

similar sized flaw would produce. Producing more "realistic" defects is cost<br />

prohibitive in most cases and, therefore, the inspector can only make an<br />

estimate of the flaw size. Computer programs that allow the inspector to<br />

create computer simulated models of the part and flaw may one day lessen<br />

this limitation.


The IIW Type <strong>Calibration</strong> Block


The IIW Type <strong>Calibration</strong> Block


The IIW Type 2 <strong>Calibration</strong> Block


The IIW Type I <strong>Calibration</strong> Block


EN12223:1999 <strong>Calibration</strong> Block


The IIW Phase Array <strong>Calibration</strong> Block


The IIW <strong>Calibration</strong> Block<br />

1 st Check Index / Check Range


The IIW <strong>Calibration</strong> Block<br />

2 nd Check Angle


The IIW <strong>Calibration</strong> Block<br />

2 nd Check Angle


Find probe angle<br />

Find Index/Range/Resolution


The IIW Phase Array <strong>Calibration</strong> Block<br />

3 rd Check Resolution


V2 <strong>Calibration</strong> Block


The IIW 2 <strong>Calibration</strong> Block<br />

Check focal point<br />

Check probe angle<br />

Check range<br />

Can not Check resolution


<strong>Calibration</strong> Blocks


<strong>Calibration</strong> Blocks- Area Amplitude Block


The standard shown in the above figure is commonly known in the US as an<br />

IIW type reference block. IIW is an acronym for the International Institute of<br />

Welding. It is referred to as an IIW "type" reference block because it was<br />

patterned after the "true" IIW block but does not conform to IIW requirements<br />

in IIS/IIW-23-59. "True" IIW blocks are only made out of steel (to be precise,<br />

killed, open hearth or electric furnace, low-carbon steel in the normalized<br />

condition with a grain size of McQuaid-Ehn #8) where IIW "type" blocks can<br />

be commercially obtained in a selection of materials. The dimensions of "true"<br />

IIW blocks are in metric units while IIW "type" blocks usually have English<br />

units. IIW "type" blocks may also include additional calibration and references<br />

features such as notches, circular groves, and scales that are not specified by<br />

IIW. There are two full-sized and a mini versions of the IIW type blocks. The<br />

Mini version is about one-half the size of the full-sized block and weighs only<br />

about one-fourth as much. The IIW type US-1 block was derived the basic<br />

"true" IIW block and is shown below in the figure on the left. The IIW type US-<br />

2 block was developed for US Air Force application and is shown below in the<br />

center. The Mini version is shown on the right.


IIW Blocks- US-1<br />

IIW Type US-1


IIW Blocks- IIW Type US-2


IIW Blocks- IIW Type Mini


V1/5, A2 Block


IIW type blocks are used to calibrate instruments for both angle beam and<br />

normal incident inspections. Some of their uses include setting metal-distance<br />

and sensitivity settings, determining the sound exit point and refracted angle<br />

of angle beam transducers, and evaluating depth resolution of normal beam<br />

inspection setups. Instructions on using the IIW type blocks can be found in<br />

the annex of American Society for <strong>Testing</strong> and Materials Standard E164,<br />

Standard Practice for Ultrasonic Contact Examination of Weldments.<br />

The Miniature Angle-Beam or ROMPAS <strong>Calibration</strong> Block


DSC Block, Mini block, Rompas Block are all mini blocks.<br />

ROMPAS <strong>Calibration</strong> Block<br />

AWS Shear Wave<br />

Distance/Sensitivity<br />

<strong>Calibration</strong> (DSC) Block


A block that closely resembles the miniature angle-beam block and is used in<br />

a similar way is the DSC AWS Block. This block is used to determine the<br />

beam exit point and refracted angle of angle-beam transducers and to<br />

calibrate distance and set the sensitivity for both normal and angle beam<br />

inspection setups. Instructions on using the DSC block can be found in the<br />

annex of American Society for <strong>Testing</strong> and Materials Standard E164,<br />

Standard Practice for Ultrasonic Contact Examination of Weldments.


A block that closely resembles the miniature angle-beam block and is used in<br />

a similar way is the DSC AWS Block. This block is used to determine the<br />

beam exit point and refracted angle of angle-beam transducers and to<br />

calibrate distance and set the sensitivity for both normal and angle beam<br />

inspection setups. Instructions on using the DSC block can be found in the<br />

annex of American Society for <strong>Testing</strong> and Materials Standard E164,<br />

Standard Practice for Ultrasonic Contact Examination of Weldments.


DSC AWS Block


<strong>Calibration</strong> Range Using DSC AWS Block<br />

www.youtube.com/embed/TEQ8Qrz4D-A


AWS Shear Wave Distance <strong>Calibration</strong> (DC) Block


AWS Shear Wave Distance <strong>Calibration</strong> (DC) Block


The DC AWS Block is a metal path distance and beam exit point calibration<br />

standard that conforms to the requirements of the American Welding Society<br />

(AWS) and the American Association of State Highway and Transportation<br />

Officials (AASHTO). Instructions on using the DC block can be found in the<br />

annex of American Society for <strong>Testing</strong> and Materials Standard E164,<br />

Standard Practice for Ultrasonic Contact Examination of Weldments.


AWS Resolution <strong>Calibration</strong> (RC) Block<br />

The RC Block is used to determine the resolution of angle beam transducers<br />

per the requirements of AWS and AASHTO. Engraved Index markers are<br />

provided for 45, 60, and 70 degree refracted angle beams.


The RC Block is used to determine the resolution of angle beam transducers<br />

per the requirements of AWS and AASHTO. Engraved Index markers are<br />

provided for 45, 60, and 70 degree refracted angle beams.


30 FBH Resolution Reference Block<br />

The 30 FBH resolution reference block is used to evaluate the near-surface<br />

resolution and flaw size/depth sensitivity of a normal-beam setup. The block<br />

contains number 3 (3/64"), 5 (5/64"), and 8 (8/64") ASTM flat bottom holes at<br />

ten metal-distances ranging from 0.050 inch (1.27 mm) to 1.250 inch (31.75<br />

mm).


Miniature Resolution Block<br />

The miniature resolution block is used to evaluate the near-surface resolution<br />

and sensitivity of a normal-beam setup It can be used to calibrate highresolution<br />

thickness gages over the range of 0.015 inches (0.381 mm) to<br />

0.125 inches (3.175 mm).


Step and Tapered <strong>Calibration</strong> Wedges<br />

Step and tapered calibration wedges come in a large variety of sizes and<br />

configurations. Step wedges are typically manufactured with four or five steps<br />

but custom wedge can be obtained with any number of steps. Tapered<br />

wedges have a constant taper over the desired thickness range.


Distance/Sensitivity (DS) Block<br />

The DS test block is a calibration standard used to check the horizontal<br />

linearity and the dB accuracy per requirements of AWS and AASHTO.


Area Amplitude Blocks provide standards for discontinuities of different size<br />

at the same depth<br />

Distance Amplitude Blocks provide standards for discontinuities of same size<br />

at the different depth


The ASTM basic set of Area/Distance Amplitude Blocks consists of ten, two<br />

inches diameter blocks


The ASTM basic set of Area/Distance Amplitude Blocks consisits of ten, two<br />

inches diameter blocks


Distance/Area-Amplitude Blocks<br />

Distance/area amplitude correction blocks typically are purchased as a tenblock<br />

set, as shown above. Aluminum sets are manufactured per the<br />

requirements of ASTM E127 and steel sets per ASTM E428. Sets can also be<br />

purchased in titanium. Each block contains a single flat-bottomed, plugged<br />

hole. The hole sizes and metal path distances are as follows:<br />

• 3/64" at 3"<br />

• 5/64" at 1/8", 1/4", 1/2", 3/4", 11/2", 3", and 6"<br />

• 8/64" at 3" and 6"<br />

Sets are commonly sold in 4340 Vacuum melt Steel, 7075-T6 Aluminum, and<br />

Type 304 Corrosion Resistant Steel. Aluminum blocks are fabricated per the<br />

requirements of ASTM E127, Standard Practice for Fabricating and Checking<br />

Aluminum Alloy Ultrasonic Standard Reference Blocks. Steel blocks are<br />

fabricated per the requirements of ASTM E428, Standard Practice for<br />

Fabrication and Control of Steel Reference Blocks Used in Ultrasonic<br />

Inspection.


ASTM E 127


Area-Amplitude Blocks<br />

Area-amplitude blocks are also usually purchased in an eight-block set and<br />

look very similar to Distance/Area-Amplitude Blocks. However, areaamplitude<br />

blocks have a constant 3-inch metal path distance and the hole<br />

sizes are varied from 1/64" to 8/64" in 1/64" steps. The blocks are used to<br />

determine the relationship between flaw size and signal amplitude by<br />

comparing signal responses for the different sized holes. Sets are commonly<br />

sold in 4340 Vacuum melt Steel, 7075-T6 Aluminum, and Type 304 Corrosion<br />

Resistant Steel. Aluminum blocks are fabricated per the requirements of<br />

ASTM E127, Standard Practice for Fabricating and Checking Aluminum Alloy<br />

Ultrasonic Standard Reference Blocks. Steel blocks are fabricated per the<br />

requirements of ASTM E428, Standard Practice for Fabrication and Control of<br />

Steel Reference Blocks Used in Ultrasonic Inspection.


Distance-Amplitude #3, #5, #8 FBH Blocks<br />

Distance-amplitude blocks also very similar to the distance/area-amplitude<br />

blocks pictured above. Nineteen block sets with flat-bottom holes of a single<br />

size and varying metal path distances are also commercially available. Sets<br />

have either a #3 (3/64") FBH, a #5 (5/64") FBH, or a #8 (8/64") FBH. The<br />

metal path distances are 1/16", 1/8", 1/4", 3/8", 1/2", 5/8", 3/4", 7/8", 1", 1-1/4",<br />

1-3/4", 2-1/4", 2-3/4", 3-14", 3-3/4", 4-1/4", 4-3/4", 5-1/4", and 5-3/4". The<br />

relationship between the metal path distance and the signal amplitude is<br />

determined by comparing signals from same size flaws at different depth.<br />

Sets are commonly sold in 4340 Vacuum melt Steel, 7075-T6 Aluminum, and<br />

Type 304 Corrosion Resistant Steel. Aluminum blocks are fabricated per the<br />

requirements of ASTM E127, Standard Practice for Fabricating and Checking<br />

Aluminum Alloy Ultrasonic Standard Reference Blocks. Steel blocks are<br />

fabricated per the requirements of ASTM E428, Standard Practice for<br />

Fabrication and Control of Steel Reference Blocks Used in Ultrasonic<br />

Inspection.


Key Words:<br />

Distance Amplitude Blocks<br />

DSC<br />

DC<br />

SC<br />

AWS RC<br />

Distance sensitivity calibration<br />

Distance calibration<br />

Sensitivity calibration<br />

AWS Resolution <strong>Calibration</strong>.


Q56: On the area-amplitude ultrasonic standard test blocks, the flat-bottomed<br />

holes in the blocks are:<br />

A. All of the same diameter<br />

B. Different in diameter, increasing by 1/64 inch increments from the<br />

No. 1 block to the No. 8 block<br />

C. Largest in the No. 1 block and smallest in the No. 8 block<br />

D. Drilled to different depths from the front surface of the test block


Q: A primary purpose of a reference standard is:<br />

A. To provide a guide for adjusting instrument controls to reveal<br />

discontinuities that are considered harmful to the end use of the<br />

product.<br />

B. To give the technician a tool for determining exact discontinuity size<br />

C. To provide assurance that all discontinuities smaller than a certain<br />

specified reference reflector are capable of being directed by the test.<br />

D. To provide a standard reflector which exactly simulates natural<br />

discontinuities of a critical size.


4.2: The <strong>Calibration</strong>s<br />

4.2.1: Distance Amplitude Correction (DAC)<br />

Distance Amplitude Correction (DAC): Acoustic signals from the same<br />

reflecting surface will have different amplitudes at different distances from the<br />

transducer. Distance amplitude correction (DAC) provides a means of<br />

establishing a graphic ‘reference level sensitivity’ as a function of sweep<br />

distance on the A-scan display. The use of DAC allows signals reflected from<br />

similar discontinuities to be evaluated where signal attenuation as a function<br />

of depth has been correlated. Most often DAC will allow for loss in amplitude<br />

over material depth (time), graphically on the A-scan display but can also be<br />

done electronically by certain instruments. Because near field length and<br />

beam spread vary according to transducer size and frequency, and materials<br />

vary in attenuation and velocity, a DAC curve must be established for each<br />

different situation. DAC may be employed in both longitudinal and shear<br />

modes of operation as well as either contact or immersion inspection<br />

techniques.


DAC Curve


http://www.huatecgroup.com/china-digital_portable_dac_avg_curves_ultrasonic_flaw_detector_ut_flaw_detector_fd350-632512.html


DAC- Distance Amplitude Correction


DAC- Distance Amplitude Correction<br />

DGS- Distance Gain Size


A distance amplitude correction curve is constructed from the peak amplitude<br />

responses from reflectors of equal area at different distances in the same<br />

material. A-scan echoes are displayed at their non-electronically<br />

compensated height and the peak amplitude of each signal is marked on the<br />

flaw detector screen or, preferably, on a transparent plastic sheet attached to<br />

the screen. Reference standards which incorporate side drilled holes (SDH),<br />

flat bottom holes (FBH), or notches whereby the reflectors are located at<br />

varying depths are commonly used. It is important to recognize that<br />

regardless of the type of reflector used, the size and shape of the reflector<br />

must be constant. Commercially available reference standards for<br />

constructing DAC include ASTM Distance/Area Amplitude and ASTM E1158<br />

Distance Amplitude blocks, NAVSHIPS Test block, and ASME Basic<br />

<strong>Calibration</strong> Blocks.


The following applet shows a test block with a side drilled hole. The<br />

transducer was chosen so that the signal in the shortest pulse-echo path is in<br />

the far-field. The transducer may be moved finding signals at depth ratios of 1,<br />

3, 5, and 7. Red points are "drawn" at the peaks of the signals and are used<br />

to form the distance amplitude correction curve drawn in blue. Start by<br />

pressing the green "Test now!" button. After determining the amplitudes for<br />

various path lengths (4), press "Draw DAC" and then press the green "Test<br />

now!" button.


DAC Java<br />

http://www.ndt-ed.org/EducationResources/CommunityCollege/Ultrasonics/<strong>Calibration</strong>Meth/applet2/applet2.htm


Developing a Distance Amplitude Correction (DAC) Curve<br />

Distance Amplitude Correction (DAC) provides a means of establishing a<br />

graphic ‘reference level sensitivity’ as a function of sweep distance on the A-<br />

scan display. The use of DAC allows signals reflected from similar<br />

discontinuities to be evaluated where signal attenuation as a function of depth<br />

may be correlated. In establishing the DAC curve, all A-scan echoes are<br />

displayed at their non-electronically compensated height.<br />

Construction of a DAC involves the use of reference standards which<br />

incorporate side drilled holes (SDH), flat bottom holes (FBH), or notches<br />

whereby the reflectors are located at varying depths. It is important to<br />

recognize regardless of the type of reflector that is used in constructing the<br />

DAC, the size and shape of the reflector must be constant over the sound<br />

path distance. Commercially available reference standards for constructing<br />

DAC include ASTM Distance/Area Amplitude and ASTM E1158 Distance<br />

Amplitude blocks, NAVSHIPS Test block, and ASME Basic <strong>Calibration</strong><br />

Blocks.


Sequence for constructing a DAC curve when performing a straight<br />

beam contact inspection on 1 ¾” thick material.<br />

1.) Using a suitable reference standard, calibrate the sweep for a distance<br />

appropriate for the material to be inspected, i.e.. using a 1” thick standard,<br />

calibrate the sweep for 2” of material travel.


Back Wall Echo


Back Wall Echo<br />

Sweep 2” / Distance 1”


2.) This example represents the use a 1 3/4” thick reference standard with<br />

1/8” side drilled holes located at 1/4 T and 3/4 T respectively. ‘T’ being equal<br />

to the block thickness.


3.) Position the transducer over the 1/4T hole and peak the signal to<br />

approximately 80% FSH (Full screen height), mark the peak of the echo on<br />

the display using a suitable marker, and record the gain setting.


4.) With no further adjustments to the gain control, position the transducer<br />

over the 3/4T hole and peak the signal, mark the peak of the echo on the<br />

display.


5.) To complete the DAC curve connect the dots with a smooth line. The<br />

completed curve represents the ‘reference level sensitivity’ for this application.


Plotting DAC Curve


DAC Curve


DAC Curve


Gain Control for FSH: It should be remember that the dB is a means of<br />

comparing signals. All <strong>UT</strong> sets are different and a FSH with a gain controls of<br />

36dB in one <strong>UT</strong> set and be at FSH at another <strong>UT</strong> set with a gain control<br />

reading of 26dB.<br />

The gain controls allow us to set sensitivity and form the basis of Ultrasonic<br />

Sizing Techniques.


Birring NDT Series, Ultrasonic Distance Amplitude Correction - DAC<br />

www.youtube.com/embed/qUqaF0PnLGA?list=UUZncq6JFram3pfQDlzGggwA


Alta Vista <strong>UT</strong> <strong>Calibration</strong> DAC Curve<br />

www.youtube.com/embed/VNgMKlp43I8


4.2.2: Finding the probe index


Exit Point<br />

A2 Block


Exit Point- A5 Block


Q16: Notches are frequently used as a reference reflector for:<br />

A. Distance amplitude calibration for shear wave<br />

B. Area amplitude calibration<br />

C. Thickness calibration for plate<br />

D. Determining the near-surface resolution


5.2.3: Checking the probe angle


Probe Angles- A2 Block


Probe Angles- A5 Block


4.2.4: <strong>Calibration</strong> of shear waves for range V1 Block


<strong>Calibration</strong> of shear waves for range V1 Block


1 st Echo from circular <strong>Section</strong>


Echo from 100mm circular <strong>Section</strong>


<strong>Calibration</strong> of shear waves for range V1 Block<br />

Test block 1 for calibrating the<br />

time base (depth scale) of a flaw<br />

detector for vertical probes<br />

(longitudinal waves) for angle<br />

probes (transverse waves), for<br />

determining the probe index and<br />

beam angle of angle probes, and<br />

for checking the short term<br />

consistency of the sensitivity of<br />

vertical probes


<strong>Calibration</strong> of shear waves for range V2 Block


25 mm radius from V2 Block


50 mm radius from V2 Block


100 mm radius from K2 Block


<strong>Calibration</strong> of shear waves for range V2 Block


Shear Wave Distance <strong>Calibration</strong> IIW Block & DSC Blocks<br />

www.youtube.com/embed/RmtHmtOozic


Exit Point /Range/Probe Angle calibration using IIW Block (Repeat-Code1)<br />

www.youtube.com/embed/Qr0dGNuq9yY


4.2.5: Dead Zone<br />

Determine the dead zone by finding the hole echo which is easily<br />

identifiable from the probe noise at the shortest range


Dead Zone<br />

Determine the dead zone by<br />

finding the hole echo which is<br />

easily identifiable from the probe<br />

noise at the shortest range


4.2.6: 20 dB Profile- A5 Block


20 dB Profile<br />

Probe Beam Line of Symmetry


20 dB Profile<br />

Probe Beam Sound Pressure


4.2.7: Transfer Correction<br />

<strong>Methods</strong> of compensating for transfer and attenuation loss differences for<br />

0attenuation 000compression probes and for shear wave compression<br />

probes. These are based on obtaining similar echo responses on both the<br />

calibration block and on the component.<br />

For 0degree probes backwall echoes are used to probes establish transfer<br />

and attenuation correction.<br />

For shear wave probes two identical probes are used in “pitch-catch” in<br />

order to obtain what are effectively backwall echoes.<br />

either method cannot be used if the either component does not have a<br />

convenient parallel section.


Example:<br />

0 degree Probe <strong>Calibration</strong><br />

40mm thick block – Gain to achieve FSH


Example:<br />

0 degree Probe <strong>Calibration</strong><br />

30mm thick block – Gain to achieve FSH


TRANSFER & ATTENUATION CORRECTION:<br />

0 degree Probes<br />

If the results are plotted on<br />

log -linear paper they will<br />

form straight parallel lines<br />

provided that there is no<br />

attenuation difference if an<br />

attenuation difference<br />

occurs then the resultant<br />

lines will no longer be<br />

parallel.


Transfer and Attenuation Correction: Shear Probe<br />

The principle for obtaining transfer correction for shear wave probes is the<br />

same as it was for compression probes except that backwall echoes are<br />

replaced by pitch --catch responses.


4.2.8: Linearity Checks (Time Base / Equipment Gain / Vertical Gain)


4.2.8.1: Linearity of time base<br />

General<br />

This check may be carried out using a standard calibration block eg A2,<br />

and a compressional wave probe. The linearity should be checked over a<br />

range at least equal to that which is to be used in subsequent testing.<br />

Method<br />

a) Place the probe on the 25mm thickness of the A2 block and adjust the<br />

controls to display ten BWEs.<br />

b) Adjust the controls so that the first and last BWEs coincide with the scale<br />

marks at 1 and 10.<br />

c) Increase the gain to bring successive backwall echoes to 80% FSH. The<br />

leading edge of each echo should line up with the appropriate reticules<br />

line.<br />

d) Record any deviations at approximately half screen height. Deviations<br />

should be expressed as a percentage of the range between the first and<br />

last echoes displayed (ie 225mm).


Tolerance<br />

Unless otherwise specified by the testing standard, a tolerance of ±2% is<br />

considered acceptable.<br />

Frequency of checking<br />

This check shall be carried out at least once per week.


Ultrasonic <strong>Testing</strong> - Horizontal Linearity (<strong>Calibration</strong>)<br />

www.youtube.com/embed/NuS6j0SmjKQ


4.2.8.2: Linearity of Equipment Gains<br />

General<br />

This is a check on both the linearity of the amplifier within the set and the<br />

calibrated gain control. It can be carried out on any calibration block<br />

containing a side-drilled hole and should be the probe to be used in<br />

subsequent testing. Reject/suppression controls shall be switched off.<br />

Method<br />

• Position the probe on a calibration block to obtain a reflected signal from a<br />

small reflector eg 1.5mm hole in the A2 block.<br />

• Adjust the gain to set this signal to 80% FSH and note the gain setting (dB).<br />

- Increase the gain by 2dB and record the amplitude of the signal.<br />

- Remove the 2dB and return the signal to 80% FSH.<br />

- Reduce the gain by 6dB and record signal amplitude.<br />

- Reduce the gain by a further 12dB (18 intotal) and record signal amplitude.<br />

- Reduce the gain by a further 6dB (24 in total) and record signal amplitude.


Tolerance<br />

Frequency of checking<br />

The check shall be carried out at least once per week.


5.2.8.3-1: Linearity of vertical display to EN12668-1<br />

Procedure: Test the ultrasonic instrument screen linearity by altering the<br />

amplitude of a reference input using an external calibrated attenuator and<br />

observing the change in the signal height on the ultrasonic instrument<br />

screen. Report the gain setting at the beginning of the test. Check the<br />

linearity at prescribed intervals from 0 dB to - 26 dB of full screen height.<br />

Repeat the test for centre frequencies for of each filter as measured in 9.5.2.<br />

Using the same set-up shown in Figure 6 set the external calibrated attenuator<br />

to 2 dB and adjust the input signal and the gain of the ultrasonic instrument<br />

so the signal is 80 % of full screen height. Without changing the gain of the<br />

ultrasonic instrument switch the external calibrated attenuator to the values<br />

given in the Table 4. For each setting measure the amplitude of the signal on<br />

the ultrasonic instrument screen.<br />

Extract from: BS EN 12668-1:2010 Non-destructive testing- Characterization and verification of ultrasonic examination equipment<br />

Part 1: Instruments


Figure 6 — General purpose set-up for equipment


4.2.8.3-2: Linearity of vertical display to ASTM E317-01<br />

Vertical Limit and Linearity:<br />

Significance—Vertical limit and linearity have significance when echo signal<br />

amplitudes are to be determined from the display screen or corresponding<br />

output signals, and are to be used for evaluation of discontinuities or<br />

acceptance criteria. A specified minimum trace deflection and linearity limit<br />

may<br />

be required to achieve the desired amplitude accuracy. For other situations they<br />

may not be important, for example, go/no-go examinations with flaw alarms<br />

or evaluation by comparison with a reference level using calibrated gain<br />

controls.<br />

This practice describes both the two-signal ratio technique (Method A) and the<br />

input/output attenuator technique (Method B).<br />

Extract from: ASTM E317-01 Standard Practice for Evaluating Performance Characteristics of Ultrasonic Pulse-Echo Examination<br />

Instruments and Systems without the Use of Electronic Measurement Instruments<br />

Note: Method A: two-signal ratio technique collecting 2 signal from the<br />

reflectors of same size at different depth.


Method A:<br />

6.3.2.1 Apparatus—A test block is required that produces two non interfering<br />

signals having an amplitude ratio of 2 to 1. These are compared over the<br />

usable screen height as the instrument gain is changed. The two amplitudes<br />

will be referred to as HA and HB (HA > HB). The two signals may occur in<br />

either screen order and do not have to be successive if part of a multipleecho<br />

pattern. Unless otherwise specified in the requesting document, any<br />

test block that will produce such signals at the nominal test settings specified<br />

can be used. For many commonly used search units and test conditions, the<br />

test block shown in Fig. 1 will usually be satisfactory when the beam is<br />

directed along the H dimension toward the two holes. The method is<br />

applicable to either contact or immersion tests; however, if a choice exists,<br />

the latter may be preferable for ease of set-up and coupling<br />

stability……(more…)


4.2.9: Time Correction Gain (TCG)<br />

Please read:<br />

http://aqualified.com/tcg-dac-ndt-ultrasound/


Q61: The vertical linear range of a test instrument may be determined by<br />

obtaining ultrasonic responses from:<br />

A. a set of distance amplitude blocks<br />

B. steel ball located at several different water path distances<br />

C. a set of area amplitude blocks<br />

D. all of the above


Q29: Test sensitivity correction for a metal distance and discontinuity area<br />

responses are accomplished by using:<br />

A. An area amplitude set of blocks<br />

B. An area amplitude and a distance amplitude set of blocks<br />

C. A distance amplitude set of blocks<br />

D. Steel balls of varying diameters.


4.3: Curvature Correction<br />

Curvature in the surface of a component will<br />

have an effect on the shape of the ultrasonic<br />

beam. The image to the right shows the beam<br />

from a focused immersion probe being<br />

projected on to the surface of a<br />

component. Lighter colors represent areas of<br />

greater beam intensity. It can be seen that<br />

concave surfaces work to focus the beam and<br />

convex surfaces work to defocus the<br />

beam. Similar effects are also seen with<br />

contact transducers. When using the<br />

amplitude of the ultrasonic signal to size flaws<br />

or for another purpose, it is necessary to<br />

correct for surface curvature when it is<br />

encountered. The "correction" value is the<br />

change in amplitude needed to bring signals<br />

from a curved surface measurement to the flat<br />

surface or DAC value.


Convex surfaces work to defocus the beam<br />

Diverge if the surface is convex.<br />

Concave surface contour-<br />

Focusing effects


Concave surfaces work to focus the beam<br />

Diverge if the surface is convex.<br />

Concave surface contour-<br />

Focusing effects


convex surfaces work to defocus the beam<br />

Convex surfaces work to defocus the beam


Convex surfaces work to defocus the beam- When sound travels from a<br />

liquid through a metal, it will converge if the surface is concave or diverge if<br />

the surface is convex.


Q: In an immersion method, the incident sound path enter the specimen<br />

interface with convex geometry, the sound path on entry into the specimen,<br />

the convex surface works to<br />

a) De-focus the sound<br />

b) Focus the sound<br />

c) Has no effect on the focusing or de-focusing the sound<br />

d) Reflected totally all the incident sound.


Q: In transmitting sound energy into a part shown below in a immersion<br />

testing, the sound beam will be:<br />

a) Diverge<br />

b) Converge<br />

c) Straight into<br />

d) Will not enter


A curvature correction curve can be generated experimentally in a manner<br />

similar to that used to generate a DAC curve, This simply requires a<br />

component with a representative reflector at various distances below the<br />

curved surface. Since any change in the radius will change the focus of the<br />

sound beam, it may be necessary to develop reference standards with a<br />

range of surface curvatures.<br />

However, computer modeling can also be used to generate a close<br />

approximation of the curvature correction value. Work by Ying and Baudry<br />

(ASME 62-WA175, 1962) and by Birchak and Serabian (Mat. Eval. 36(1),<br />

1978) derived methods for determining "correction factors" to account for<br />

change in signal amplitude as a function of the radius of curvature of convex,<br />

cylindrical components.<br />

An alternative model for contact and immersion probe inspection was more<br />

recently by researchers at the Center for NDE at Iowa State University. This<br />

mathematical model further predicts transducer radiation patterns using the<br />

Gauss-Hermite model, which has been used extensively for simulation of<br />

immersion mode inspections.


The resulting model allows computationally efficient prediction of the full<br />

ultrasonic fields in the component for<br />

1. any frequency, including broadband measurements.<br />

2. both circular and rectangular crystal shapes.<br />

3. general component surface curvature<br />

4. both normal and oblique incidence (e.g., angle beam wedges) transducers.<br />

When coupled with analytical models for defect scattering amplitudes, the<br />

model can be used to predict actual flaw waveforms. The image shown<br />

above was generated with this model.


The plot to the right shows an example curvature correction curve and DAC<br />

curve. This curvature correction curve was generated for the application of<br />

detecting a #4 flat bottom hole under a curved surface as shown in the<br />

sketch and photograph. An immersion techniques was used generate a<br />

shear wave since the reflective surface of the target flaw was not parallel with<br />

the surface. The DAC curve drops monotonically since the water path<br />

ensures that the near field of the sound beam is always outside the part. The<br />

correction factor starts out negative because of the focusing effect of the<br />

curved surface. At greater depths, the correction factor is positive due to the<br />

increased beam spread beyond the focal zone caused by the surface<br />

curvature.


Curvature Corrections


A table of correction values and the DAC and curvature correction curves for<br />

different size radiuses can be found at the following link.<br />

https://www.nde-ed.org/EducationResources/CommunityCollege/Ultrasonics/<strong>Calibration</strong>Meth/table/table.htm


Curvature Correction


Curvature Correction


4.4: <strong>Calibration</strong> References & Standards<br />

What are standards?<br />

Standards are documented agreements containing technical specifications or<br />

other precise criteria to be used consistently as rules, guidelines, or<br />

definitions of characteristics, in order to ensure that materials, products,<br />

processes, and services are fit for their purpose.<br />

For example, the format of the credit cards, phone cards, and "smart" cards<br />

that have become commonplace is derived from an ISO International<br />

Standard. Adhering to the standard, which defines such features as an<br />

optimal thickness (0.76 mm), means that the cards can be used worldwide.


An important source of practice codes, standards, and recommendations for<br />

NDT is given in the<br />

Annual Book of the American Society of <strong>Testing</strong> and Materials,<br />

ASTM. Volume 03.03, Nondestructive <strong>Testing</strong><br />

is revised annually, covering acoustic emission, eddy current, leak testing,<br />

liquid penetrant, magnetic particle, radiography, thermography, and<br />

ultrasonics.<br />

There are many efforts on the part of the National Institute of Standards and<br />

Technology (NIST) and other standards organizations, both national and<br />

international, to work through technical issues and harmonize national and<br />

international standards.


Reference Reflectors:<br />

are used as a basis for establishing system performance and sensitivity.


Spherical reflectors are often used in immersion techniques for assessing<br />

sound fields.<br />

1. Omni direction<br />

2. Sphere directivity patterns reduce reflectance as compare with plane<br />

reflector<br />

3. Sphere of any materials could be used, however steel balls are often<br />

preferred.


Reference Reflectors are used as a basis for establishing system<br />

performance and sensitivity.


4.5: Questions & Answers<br />

Exercises


Q80: The 50 mm diameter hole in an IIW block is used to:<br />

(a) Determine the beam index point<br />

(b) Check resolution<br />

(c) Calibrate angle beam distance<br />

(d) Check beam angle<br />

Q81: The 100 mm radius in an IIW block is used to:<br />

(a) Calibrate sensitivity level<br />

(b) Check resolution<br />

(c) Calibrate angle beam distance<br />

(d) Check beam angle


Q6: The Notches are frequently used for reference reflectors for:<br />

A. Distance amplitude calibration for shear wave<br />

B. Area amplitude calibration<br />

C. Thickness calibration of plate<br />

D. Determine of near surface resolution<br />

Q17: Notches provide good reference discontinuities when <strong>UT</strong> examination is<br />

conducted to primarily detect defects such as:<br />

A. Porosity in rolled plate<br />

B. Inadequate penetration at the root of weld<br />

C. Weld porosity<br />

D. Internal inclusion


4.6: Video Time<br />

http://v.pps.tv/play_315ARS.html


Birring NDT Series, <strong>UT</strong> of Welds Part 1 of 2 - CALIBRATION<br />

https://www.youtube.com/embed/SRJktrHUlM4


Birring NDT Series, Ultrasonic <strong>Testing</strong> # 4, Angle Beam Shear Wave <strong>UT</strong> as<br />

per AWS D1.1<br />

www.youtube.com/embed/vXcAI-Zci30

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!