Rinse wash time At the

Rinse wash time At the end of the thermal disinfection cycle the chamber and bedpans are sprayed with cold water to reduce the internal temperature. This parameter permits adjustment to the final rinse time. Default setting is 10 seconds. Default setting: 15 Seconds - Mini Default setting: 15 Seconds - Midi Default setting: 10 Seconds – Optima 2 Minimum setting: 10 seconds Maximum setting: 60 seconds Lid pre-open delay time Parameter for controlling the time in seconds before the lid will automatically open: Default setting: 00 Seconds Minimum setting: 00 Seconds Maximum setting: 20 Seconds Lid auto close time Parameter for controlling the time in seconds before the lid will automatically close: Default setting: 60 Seconds Minimum setting: 00 Seconds Maximum setting: 60 Seconds Copyright © 2017 DDC Dolphin Ltd. Page 14 of 41 Feb 18 SGPana – V27 1.6

Disinfection temperature setting (degrees centigrade) Thermal disinfection is defined by holding the chamber at a given minimum temperature for a given period of time. In the UK the minimum temperature is 80 Degrees Centigrade which has to be held for a period of 60 seconds. Regional thermal disinfection standards may vary around the world; adjustments made to this parameter define what temperature the chamber must achieve before the one minute cycle will start. Default setting is 82 Degrees Centigrade which provides a safety margin ensuring that the 80 Degrees centigrade is met. Should the chamber temperature drop below the number set in this parameter; a 60 second timer (Thermal disinfection time) will reset to zero starting again once the temperature has reached the minimum temperature set within this parameter. Default setting: 82 Degrees Centigrade Minimum setting: 75 Degrees Centigrade Maximum setting: 90 Degrees Centigrade Initial fill (seconds) The initial cold-wash time in seconds. Default setting 30 seconds. Extending this time may produce better results from the wash but should only be considered were the water pressure is sufficient to refill the water tank in line with the flow rate of the pump. If adjusting this parameter test and ensure that the water pump does not run dry. Default setting: 40 Seconds - Mini Default setting: 30 Seconds - Midi Default setting: 20 Seconds – Optima 2 Minimum setting: 5 Seconds Maximum setting: 45 Seconds Copyright © 2017 DDC Dolphin Ltd. Page 15 of 41 Feb 18 SGPana – V27 1.6

cNODE® Transponders Maxi, Midi and Mini - Kongsberg Maritime
Kategorie mini midi maxi maxi 2 Modell HP 2170p ... - u:book
Line 6 PODStudio & TonePort MIDI Setup Guide (Rev 2 ... - Thomann
YOTTA Mini Software installation guide(5914KB) - Axus
Yotta Mini Sport Software installation guide - Axus
Vernier LabQuest Mini User's Manual - Vernier Software & Technology
Vision MINI Smart Camera Quick Start Guide - Microscan
Samsung Galaxy Mini 2 S6500D Unlock Software - GalaxyUnlocker
Line 6 FBV MkII Advanced User Guide, English (Rev. 2)
Samsung Galaxy Mini 2 S6500L Unlock Software - GalaxyUnlocker
User Guide for Alliance Planner Programming Software - UniMac
WinTEC3 Software user's Guide - Electromotive Engine Controls
MONOPACK 2 Manual - Pro/Motion Mechanik + Elektronik + Software
Installer's handbook - 1/3 Types of installation - 2/3 Software guide ...