Patient Parameter

Patient Parameter Specifications Appendix 5.2.6.6 NIBP Start Pressure Settings and Ranges The Start Pressure is adjustable and is set to the following defaults: PATIENT SIZE PRESSURE INCREMENT (DEPENDING ON ALGORITHM) DEFAULT START PRESSURE Adult Mode 30 – 60 mmHg 178 ±5 mmHg Pediatric Mode 30 – 60 mmHg 133 ±5 mmHg 5.2.6.7 NIBP Measurement Cycle There is one mode of measurement operation: manual. The manual mode requires the operator to initiate the measurement cycle. In the manual mode, the unit adjusts the inflation pressure according to the previous systolic pressure. After the first successful measurement is made, the subsequent inflation pressure becomes +50 ±10 mmHg above the previous systolic pressure measurement. 5.2.6.8 SpO 2 Agency Requirements The SpO 2 function performs in accordance with the requirements of EN 865: 1997. 5.2.7 SpO 2 Performance Requirements The Rosebud monitor is capable of providing SpO2 functional saturation level measurements via the DPM pulse oximeter. SpO 2 Measurement Range: 0 – 100% Resolution: 1% Accuracy: ±2% from 70 to 100%; 0~69% unspecified Display Refresh Rate: Once per second. Pulse Rate Range: 25 – 254 bpm ±1 bpm Resolution: 1 bpm Accuracy: ±1 bpm NOTE: The sensor measurement wavelengths are nominally 660 nm for the red LED and 940 nm for the infrared LED. Maximum optical power output for LED is 4 mW. 5 - 6 0070-10-0603-60 Rosebud® Operating Instructions

Appendix Patient Parameter Specifications 5.2.8 Power Supply 5.2.8.1 Power Source The Rosebud monitor auto-selects its power source from those available. The monitor uses the following priority in choosing the power source: 1. AC Mains Power 2. Internal battery power The monitor operates from AC Mains power with or without the internal batteries installed. 5.2.9 AC Mains Power Source Input Voltage: 100 – 240 VAC (+/-10%) Line Frequency: 50/60 Hz (+/-3 Hz) Current: 0.7 – 0.4 Amps 5.2.10 Battery Power Time to Shutdown from Low Battery: 5.2.10.1 Lithium Ion Battery: P/N 0146-00-0079 The battery pack is 7.2 VDC, 6.6 Amp-hr. >10 minutes but < 20 minutes after indication, with 1 new, fully charged battery. The minimum Battery Run Time: 14.0 hours from one fully charged new battery at 25 °C with continuous SpO2 measurement and NIBP measurements taken at 7 minute intervals. The Battery Charge Time: 4.5 hours maximum. 17.5 hours with continuous SpO2 measurement and no NIBP measurements being taken. Rosebud® Operating Instructions 0070-10-0603-60 5 - 7

Guide to Operations - Abandonia
OPERATING INSTRUCTIONS - KTM
Operator Manual - Crown Equipment Corporation
NavOne 4500/4000 Operating Instruction Manual
Operating instructions
IN13 Operating Instruct - Euroheat
Operating Instructions - Skywalk
Ambient Light Powered Wireless Weather Station OPERATING ...
Operating Instructions - Vetter GmbH
Installation and Operating Instructions R.O.C.O.F ... - Enersom.com
OPERATING MANUAL English - Scubapro
OPERATING INSTRUCTIONS www.ozito.com.au
und Betriebsanleitung Assembly and operating instructions ...
Operating Instructions - Help - Weber
Operating instructions - mobilenavigation.mybecker.com - Harman ...
Console Operation - perfusion.ws
Operating Manual TEC PRIME
Automatic Gate Operators - Linear
JVL Operators Manual
operating instructions - Babystyle
Bedienungsanleitung Operating Manual Mode d'emploi ...