Table 7: Study No. 14-VIN-657– outcome of validation of the analytical methods<br />
Parameter Acceptance criterion Found value acceptable<br />
(yes/no)<br />
Selectivity<br />
6 individual sources,<br />
14 lots, no significant yes<br />
response is LLOQ – ULOQ<br />
≥5 samples at ≥4 concentration levels<br />
each,<br />
±20% for LLOQ<br />
±15% for >LLOQ – ULOQ<br />
≥5 samples at ≥4 concentration levels<br />
each,<br />
±20% for LLOQ<br />
±15% for >LLOQ – ULOQ<br />
Accuracy and precision should be within<br />
the set criteria, i.e. within ±15%<br />
normalized MF calculated from the 6 lots<br />
of matrix ≤15 % for low and high<br />
concentration level<br />
Low, medium, high<br />
QC within ±15%<br />
Low, medium, high<br />
QC within ±15%<br />
Low, medium, high<br />
QC within ±15%<br />
Low, medium, high<br />
QC within ±15%<br />
Dilution factor 2 & 4;<br />
complies<br />
6 lots, no significant<br />
matrix effect<br />
Carry over response is