Les données d’exposition<br />

Class of equipment Acoustic I ISPTA ISPPA pm<br />

Power<br />

mW mW/cm 2 mW/cm 2 W/cm 2 Atm<br />

Manual Compound scanners 0.5-20 0.4-10 1 0-170 0.5-280 1.7-50<br />

Auto-sector scanners 9-20 7-15 6-30 25-100 1.7-27<br />

Sequential linear arrays 0.1-33 0.06-10 0.01-12 0.2-70 1.1-17<br />

Ophthalmic A-mode, B-mode 0.6-1 0. 3-5 20-34<br />

Ophthalmic real time 0.2-0.06 0.07-0.2 0.2-0.6<br />

Pulsed Doppler-cardiac 8-24 6-32 50-290 3-14<br />

CW Doppler-obstetrical 8-37 3-19 9-57<br />

CW Doppler fetal monitoring 5-32 3-20 9-80<br />

CW Doppler peripheral 6-105 38-840 110-2500<br />

Pulsed Doppler peripheral 6-10 87-175 350-700<br />

Valeurs extrêmes <strong>des</strong> paramètres d’exposition pour différents appareils