(1) If reproducibility was carried out, intermediate precision does not need to be carried out.

(2) In the identification test, combining one or more analytical procedures may be needed to reach the required level

of discrimination.

(3) It may be required in some cases.

ANNEX II

(Republicado no DOU nº 156, de 15 de agosto de 2017)

Formula 1. Accuracy calculation.

Experimental mean concentration

Recovery = x 100

Theoretical concentration

Or

CA (added sample) – CA (sample)

Recovery = x 100

CTA

Where: CA is the experimental concentration of analyte and CTA is the theoretical

concentration of analyte added.

Formula 2. Limit of detection calculation.

LD= 3.3.σ

IC

Where: IC is the calibration curve inclination, σ is the standard deviation and may be

obtained in 3 ways:

I – from the standard deviation of the intercept with Y-axis of at least 3 calibration curves

constructed containing analyte concentration near the assumed limit of detection;