SC450 Single Parameter Colorimeter Chlorine - Orbeco-Hellige
SC450 Single Parameter Colorimeter Chlorine - Orbeco-Hellige
SC450 Single Parameter Colorimeter Chlorine - Orbeco-Hellige
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Calibration ModeFactory calibration resetResetting the user calibration to the original factory calibration willreset all methods and ranges.StorePrintCalRESULTDateTimeA user calibrated method is indicated by a "Cal" symbol while the testresult is displayed.ModeOnOffZeroTestTo reset the calibration press both the [MODE] and [ZERO/TEST] keyand hold.Turn the unit on using the [ON/OFF] key.Release the [MODE] and [ZERO/TEST] keys after approx. 1 second.The following messages will appear in turn on the display:SELCALThe factory setting is active.(SEL stands for Select)or:SELcALCalibration has been set by the user.(If the user calibration is to be retained, turn the unit off using the[ON/OFF] key).ModeSELCALOnOffCalibration is reset to the factory setting by pressing the [MODE] key.The following messages will appear in turn on the display:Turn the unit off using the [ON/OFF] key.<strong>SC450</strong>_V1_11.201017