3. Important Notes<br />

3.1. User information in display read-out<br />

Z E R O E R R O R<br />

+ + + E R R O R<br />

– – – E R R O R<br />

Light absorption for zero value is too great. Cause<br />

could be, for <strong>ins</strong>tance, a dirty optical component in<br />

the cuvette chamber.<br />

Measurement range has been exceeded or water<br />

sample is too opaque.<br />

Measurement range has been exceeded.<br />

L O W B A T<br />

The 9 V battery must be changed for a new battery<br />

(or the storage battery must be charged). Initiated<br />

measuring routines should not be continued.<br />

B A T T O F F<br />

T E M P D R I F T<br />

* * * * * * * * * * * * * * * *<br />

(16 blinking stars)<br />

Power supply to the device is inadequate. The device<br />

can only be turned off by pressing the ON/OFF key.<br />

The temperature difference between the photometer<br />

and ambient atmosphere is too great. Restarting the<br />

measurement routine after 15 minutes at the earliest<br />

is recommended.<br />

The device can no longer be shut off under control.<br />

There is an electronic fault which can, for example,<br />

be caused by the entry of water.<br />

It is urgently recommended that you call servicing!<br />

The unit must be switched off by disconnecting the<br />

battery or storage battery.<br />

26