3. Important Notes<br />
3.1. User information in display read-out<br />
Z E R O E R R O R<br />
+ + + E R R O R<br />
– – – E R R O R<br />
Light absorption for zero value is too great. Cause<br />
could be, for <strong>ins</strong>tance, a dirty optical component in<br />
the cuvette chamber.<br />
Measurement range has been exceeded or water<br />
sample is too opaque.<br />
Measurement range has been exceeded.<br />
L O W B A T<br />
The 9 V battery must be changed for a new battery<br />
(or the storage battery must be charged). Initiated<br />
measuring routines should not be continued.<br />
B A T T O F F<br />
T E M P D R I F T<br />
* * * * * * * * * * * * * * * *<br />
(16 blinking stars)<br />
Power supply to the device is inadequate. The device<br />
can only be turned off by pressing the ON/OFF key.<br />
The temperature difference between the photometer<br />
and ambient atmosphere is too great. Restarting the<br />
measurement routine after 15 minutes at the earliest<br />
is recommended.<br />
The device can no longer be shut off under control.<br />
There is an electronic fault which can, for example,<br />
be caused by the entry of water.<br />
It is urgently recommended that you call servicing!<br />
The unit must be switched off by disconnecting the<br />
battery or storage battery.<br />