25.12.2020 Views

Comparison between class A And meter PQM

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

Voltage Sags and Total Harmonic Distortion

Monitoring in Power Systems. A case study

Niculai STANCIU 1 , Dorel STĂNESCU 2 , Petru POSTOLACHE 3 , Willibald SZABO 4

1 S.C. FDEE Electrica Distribuţie Transilvania Sud S.A., Brasov, ROMANIA, niculai_stanciu@yahoo.com

2 S.C. FDEE Electrica Distribuţie Transilvania Sud S.A., Brasov, ROMANIA, dorel.stanescu@electricats.ro

3 University „POLITEHNICA” of Bucharest, ROMANIA, petrupostolache@yahoo.com

4 University “TRANSILVANIA” Brasov, ROMANIA, w_l_szabo@yahoo.co.uk

Abstract: The purpose of Power Quality (PQ) measurements

can be either continuous “surveillance” as a task performed

by the Distribution System Operator (DSO) or specific parameter

“evaluation” for limit violation. Individual strategies

can be applied for each goal but a combined approach

is preferred.

This paper presents a comparison between results obtained

using an electricity meter with PQ capabilities (surveillance

task) and a class A PQ monitoring equipment (used to evaluate

limit violation determination).

Measurement uncertainty was calculated for each device

and measurand and the result was compared with IEC

61000-4-30 requirements.

KEYWORDS: power quality, voltage sags, Total Harmonic

Distortion, measurement uncertainty

I. INTRODUCTION

Monitoring Power Quality parameter is one of the

DSO’s constant concerns. Improving the service quality

when the number of power electronics applications is increasing

must be accompanied with a continuous preoccupation

for finding new means to identify and evaluate the

PQ problems. Preventions measures may then be taken.

PQ parameter evaluation is used to determine limit violations.

Measurements must have a proper accuracy in

order to obtain efficient correction measures.

PQ parameters surveillance and evaluation are actually

two complementary stages of the same process of providing

the customer with high quality services [1].

II. MEASURING AND MONITORING EQUIPMENT

Nowadays, measurement technology is capable of very

fast acquisition and processing and high accuracy.

Systems for acquisition, management and data storage

can record signals corresponding to phenomena that affect

electric networks over relevant periods of time.

The equipment for measuring/monitoring PQ parameters

is classified in compliance with IEC 61000-4-30, [2],

standard that is correlated with measuring methods/procedures

and accuracy level:

• Class A instruments;

• Class S instruments;

• Class B instruments.

The metrological and technical conditions for each

class, as well as the use of the results, are regulated by the

same standard and connected standards IEC 61000-4-7,

[3], 61000-4-15, [4], and IEC 61557-12, [5].

The „Necessary Condition” is fulfilled when an instrument

included in one of the classes above is available.

In the measurement process the “Sufficient Condition”

must be proven. This means that the methods, procedures

and recommendations must be correctly applied. We will

try to demonstrate this below.

Proper measuring and monitoring instruments will be

selected depending on the actual purpose.

Choosing metering / monitoring equipment takes into

account the technical and economical analyses. First of

all, the technical and metrological requirements must be

fulfilled and then the initial investment and maintenance

costs must be considered.

The content of „Technical Specification” is very important

during this kind of analysis because the lack of

relevant information can influence the correct decision,

[6].

III. TECHNICAL SPECIFICATIONS FOR MEASURING

INSTRUMENTS

Technical specification for measuring/monitoring instruments

of PQ parameters includes information for static

and dynamic performance characteristics. It is important

to remember that those characteristics refer to:

• Performance of the measuring instrument for

each characteristic;

• Time intervals recommended for instrument recalibration

to confirm metrological status;

• Probability of measurement performance.

Important: If the tolerated limits and probability distribution

are provided in the Technical Specification for

some errors, then the measurement uncertainty can be

estimated, [7]. While measuring uncertainty evaluation, it

is very recommended that the Technical Specification

should make references to basic probability distribution

used to establish tolerance limits for each performance

parameter.

Evaluation of static and dynamic characteristics of

measuring instruments is required at certain time intervals

(no longer than the time interval specified by manufacturer).

This process of recalibration ensures that those

characteristics stay within the specified tolerance limits.

On these occasions the systematic errors and associate

uncertainty are determined. Then, the risk analysis is performed

to check whether the instrument is outside imposed

tolerance, [8].

In this way, the measurement results characterize phenomena

with high objectivity.


IV. MEASUREMENT UNCERTAINTY

Every measurement is accompanied by errors associated

with the measuring equipment, environmental conditions

during measurements as well as the methods /

procedures used. Because systems that get involved in

measuring process cannot ensure perfect reliability the

errors can also vary. In most cases phenomena can be

controlled only to a certain extent, with a certain probability.

For this reason the deviation of measuring error as

magnitude and sign is named „measurement uncertainty”.

A. Measurement uncertainty estimation based on

technical specification, before actual measurements

In this case, it is important the user knows that some

technical specifications are established by testing a selected

sample from mass production of that model. Because

the results are applied to the whole population, limits

are stated to ensure that the majority would comply.

Thus, we deal with a certain confidence level and degrees

of freedom depending on the sample size. Tolerance limits

for performance characteristic, „x”, will be written as:

± L = ± tα / 2,

ν ⋅ s x

(1)

where:

t = t – statistic coefficient;

α / 2,ν

α =relevance level = 1 – C/100;

C = confidence level, [%];

ν = degrees of freedom = n-1;

n = sample size;

s

x

= sample standard deviation.

Uncertainty, u, related to a characteristic with

normal distribution of error variation, estimated to the

tolerated error limits, ± L, confidence level, p, normal

−1

inverse cumulative distribution function, Φ (.)

, is:

L

u =

(2)

−1⎛ 1+

p ⎞

Φ ⎜ ⎟

⎝ 2 ⎠

Uncertainty estimation using technical specifications

facilitates the adequate choice of measuring instrument.

B. Measurement uncertainty estimation after actual

measurements

Uncertainty evaluation methodology used after actual

measurement ensures proper statement of measurement

result as close as possible to the real value.

Following the recommendations of standard ISO/IEC

98-3, i.e. „Evaluation of measurement data — Guide to

the expression of uncertainty in measurement”, [9], the

measurement uncertainty can be determined for each

measured parameter taking into account the mathematical

model for variation law and nature of errors in measurement

process.

In this way, the measurement results characterize phenomena

with high objectivity, with probability of 95%.

V. EXPERIMENTAL RESULTS

Voltage sags and Total Harmonics Distortion for voltage

and current are most common PQ parameters and, as

a consequence, measurements in “surveillance” and

“evaluation” took them into account.

For this purpose we used:

• Energy static meter A1RLQ+, SN: 2748989 for

surveillance;

• PQ Analyzer Fluke 435, SN: N10140 for evaluation.

Electrical signals were generated using C300 Calmet

Three phase Calibrator, SN: CT/260/2011.

Metrologic trace is ensured by making use of the following

documents: Calibration Certificate for C300 Calemt

Calibrator, [10], and Fluke 435, [11], and Metrological

Certificate, [12], for Energy static meter A1RLQ+.

A. Voltage sags measurement/ monitoring

Predefined voltage residual values for voltage sags are

presented in Table 1.

Table 1

Inferior

limit of

U res As in IEC 61000-4-30

supply

voltage [V] % din U din Class

0,9U din 207,00 0 A S B

206,77 -0,10%

206,54 -0,20%

204,70 -1%

202,40 -2%

Voltage levels generated with C300 Calmet Calibrator,

SN: CT / 260 / 2011 are indicated in Table 2. Test performed

on 03.12.2012.

Table 2

Time U 1 U 2 U 3

[hh:mm:ss] [V] [V] [V]

17:39:13 230,00 230,00 230,00

17:39:13 207,00 207,00 207,00

17:39:13 230,00 230,00 230,00

17:39:13 206,77 206,77 206,77

17:39:13 230,00 230,00 230,00

17:39:13 206,54 206,54 206,54

17:39:13 230,00 230,00 230,00

17:39:13 206,54 206,54 206,54

17:39:13 230,00 230,00 230,00

17:39:13 204,70 204,70 204,70

17:39:13 230,00 230,00 230,00

17:39:13 202,40 202,40 202,40

17:39:13 230,00 230,00 230,00

Table 3 shows the recordings of voltage sags measuring

/ monitoring using PQ Analyzer Fluke 435, SN:

N10140. Test performed on 03.12.2102.

Table 3


Time

Event

Residual

voltage

[hh:mm:ss] Type Duration [V]

17:39:56 650ms Sag 0m.1s.426ms. 206,95

17:39:59 915ms Sag 0m.1s.576ms. 206,71

17:40:03 854ms Sag 0m.1s.637ms. 206,48

17:40:07 854ms Sag 0m.1s.637ms. 206,49

17:40:11 713ms Sag 0m.1s.787ms. 204,65

17:40:15 653ms Sag 0m.1s.856ms. 202,35

Table 4 presents the recordings of voltage sags measuring

/ monitoring using Energy static meter A1RLQ+, SN:

2748989. Test performed on 03.12.2012.

Table 4

Time

Event

[hh:mm:ss]

17:38:29

Started sag [7202]

1

17:38:32 End sag [7203]

17:38:33

Started sag [7202]

2

17:38:36 End sag [7203]

17:38:37

Started sag [7202]

3

17:38:40 End sag [7203]

17:38:41

Started sag [7202]

4

17:38:44 End sag [7203]

17:38:45

Started sag [7202]

5

17:38:48 End sag [7203]

17:38:49

Started sag [7202]

6

17:38:53 End sag [7203]

B. Measuring / monitoring of voltage and current

distorted waveforms

Using the C300 Calmet Calibrator several three phase

distorted signals were generated:

• three phase distorted voltage waveforms with

THD value of 10%;

• three phase distorted current waveforms with

THD value of 50%;

Figures 1, a)... d) present screen captures and values

measured with Fluke 435 PQ analyzer for voltage waveforms.

Fig. 1, b) Three phase voltage waveforms

Fig. 1, c) Harmonics spectrum

Fig. 1, d) Voltage THD and individual harmonics values

Test performed on 20.11.2011 during the time interval

18:23 - 18:50.

Voltage THD values per phase recorded with Energy

static meter A1RLQ+, with SN: 2748989, are presented

in Table 5. Test performed on 20.11.2012.

Fig. 1, a) Voltage RMS values and phasors for voltage


Table 5

THD THD THD

Time

phase L 1 phase L 2 phase L 3

18:47 9,8 10,0 9,9

Figures 2, a) ... d) present screen captures and values

measured with PQ Analyzer FLUKE 435 for current

waveforms.

Fig. 2, d) Current THD and individual harmonics values

Fig. 2, a) Current RMS values and phasors for voltage and current

Fig. 2, b) Three phase current waveforms

Fig. 2, c) Harmonics spectrum

Test performed on 20.11.2011 during the time interval

18:23 to 18:50.

Current THD values per phase recorded with Energy

static meter A1RLQ+, with SN: 2748989, are presented

in Table 6.

Table 6

THD THD THD

Ora

phase L 1 phase L 2 phase L 3

18:47 49,8 48,8 48,9

Residual voltage measurement uncertainty for voltage

sags detected by Energy static meter A1RLQ+ was estimated

using information from technical specification,

[13], and formula (2):

2,3

U = = 1, 02 V

2,248

where:

• 2,3 V represents the limit value L = ± 1% from

voltage reference value U ref = 230 V, as in IEC

62053-22:2003, [14];

• 2,248 is the value of normal inverse cumulative

−1

distribution function Φ (.)

, for 95% probability,

assigned for limit L.

We can state that voltage sags detected by Energy static

meter A1RLQ+ have a measurement uncertainty of 1V

with a 95% probability.

The residual voltage measurement uncertainty for voltage

sags detected measured with PQ Analyzer Fluke 435

was estimated using recommendations included in Guide

ISO/IEC 98-3, [9], with formula:

U = k ⋅u

n

2

c

= k ⋅ ∑u i

i=

1

(3)

where:

• k multiplier of the combined uncertainty; k = 2

when the real value falls within ± U interval with

95% probability;

• u c composed standard uncertainty;

• u i components of standard uncertainty allocated to

different error types.


Measurements were performed in reference conditions

with ambient temperature inside (21,7 ... 23,2) ºC interval

and humidity inside (47,6 ... 58,9)% interval. In this case

only following components have significant contributions:

• Instrument uncertainty stated by Calibration Certificate

no. 1287, [11]:

U

PMD

0,05 V

PMD

= = = 0, 025V

k 2

• Hysteresis uncertainty:

Vd

−Vc

0,01V

H

= = = 0, 00577V

3 3

• Resolution uncertainty:

0,5d

0,5 ⋅ 0,01V

rez

= = = 0, 00288V

3 3

Note: 10 measurements were performed under

repeatability conditions for each Ures value. The

experimental values measured with the network analyzer

FLUKE 435 were identical, as they were generated with

Calibrator Calmet C300, which has a high level of

accuracy. In this case the standard deviation is zero. The

standard uncertainty of type A is implicitely null.

So, composed uncertainty calculated according “uncertainty

propagation law” [9], is:

2 2 2

uc = uδ

PMD

+ uδH

+ uδrez

= 0, 0258V

Extended uncertainty according formula (3) is:

U = 2 ⋅ 0,0258V

= 0,0516 V ≅ 0, 05V

In case of THD measurements, technical specification,

[15], or calibration certificate, [10], do not include tolerance

limits or uncertainty value.

For this reason, uncertainty cannot be calculated.

Instead, we compared the measured values against prescribed

values:

• maximal measuring error for Energy static meter

A1RLQ+ :

o 2% for distorted voltage

o 2,4% for distorted current

• maximal measuring error for PQ Analyser Fluke

435 :

o 2% for distorted voltage

o 1,8% for distorted current

As far as measurement accuracy for THD is concerned,

both instruments fall within limit of 5% for Class I

equipments stated by IEC 61000-4-7:2003, Electromagnetic

compatibility guide, [3].

VI. CONCLUSIONS

The goal of experimental measurements was to identify

and propose an adequate strategy for choice of instruments

used for measuring/monitoring PQ parameters.

Perturbations simulated using Calmet C300, SN: CT /

260 / 2011 calibrator were measured / monitored correctly

with both Energy static meter A1RLQ+ and PQ

Analyzer Fluke 435:

a) voltage sags;

b) distorted waveforms for voltage and current with

predefined harmonics spectrum and THD.

This means that:

• sag threshold settings were correct;

• correct settings were possible because the uncertainty

was known;

• the necessary condition is fulfilled in order to develop

a sag monitoring strategy based on modeling

software for choosing optimal location of PQ

enabled electronic energy meters and/or PQ network

analyzers.

Both types of PMD can be used for scheduling permanent

and/or pre-determined measurements/monitoring of

distorted waveforms of voltage and current. Thus, based

on these measurements/monitoring, one can highlight

distortions between “starting time” (initiation of distortion

event) and “inhibition time” (end of the event), [17].

We can then choose to use the Energy static meter with

PQ capabilities for parameter surveillance, or in case of

further investigation, PQ analyzers must be employed.

REFERENCES

[1] Golovanov, I.C., Measurement of electrical quantities in

power system, Romanian Academy and AGIR Ed.

Bucharest, 2009

[2] ***Electromagnetic compatibility (EMC). Part 4-30:

Testing and measurement techniques. Power quality measurement

methods, IEC 61000-4-30

[3] ***Electromagnetic compatibility (EMC) - Part 4-7: Testing

and measurement techniques - General guide on harmonics

and interharmonics measurements and instrumentation, for

power supply systems and equipment connected thereto, IEC

61000-4-7

[4] Electromagnetic compatibility (EMC) - Part 4-15: Testing

and measurement techniques - Flickermeter - Functional

and design specifications, IEC 61000-4-15

[5] ***Electrical safety in low voltage distribution systems up to

1 000 V a.c. and 1 500 V d.c. – Equipment for testing.

Measuring or monitoring of protective measures – Part 12:

Performance measuring and monitoring devices (PMD),

IEC 61557-12

[6] *** Measuring and Test Equipment Specifications, NASA

Measurement Quality Assurance Handbook - Annex 2,

NASA-HDBK-8739.19-2, Approved: 2010-07-13

[7] *** Measurement Uncertainty Analysis, Principles and

Methods, NASA Measurement Quality Assurance Handbook

- Annex 3, NASA-HDBK-8739.19-3, Approved: 2010-07-13

[8] ***Estimation and Evaluation of Measurement Decision

Risk, NASA Measurement Quality Assurance Handbook -

Annex 4, NASA-HDBK-8739.19-4, Approved: 2010-07-13

[9] ***Uncertainty of measurement. Part 3: Guide to the

expression of uncertainty in measurement (GUM:1995)

ISO/CEI GUIDE 98-3

[10] ***Certificate of Calibration CT/260/2011, for C300 Calmet

Calibrator, Serial number: 21042

[11] ***Certificate of Calibration 1287 of 09/26/2011, for PQ

Analyzer Fluke 435, Serial number N10140


[12] ***Metrological Certificate nr. TM2196551 of 03/24/2011

for Energy static meter A1RLQ+, SN: 2748989

[13] ***User manual – Energy meter Alpha plus

[14] ***Electricity metering equipment (a.c.). Particlar requirements

Part 22: Static meters for active energy (classes 0,2S

and 0,5S), IEC 62053-22:2003

[15] ***User manual – Calmet C300 Calibrator

[16] Olguin, G., An Optimal Trade-off between Monitoring and

Simulation for Voltage Dip Characterization of

Transmission Systems, IEEE/PES Transmission and

Distribution, Conference & Exhibition: Asia and Pacific

Dalian, China, 2005

[17] Bollen, M.H.J, Gu, Y.H.I, Signal processing of power quality

disturbance, IEEE Press, A John Wiley & Sons, INC.,

PUBLICATION, 2006

[18] N. Golovanov. P. Postolache. C. Toader. "Efficiency and

power quality" AGIR Ed. 2009

[19] Szekely, I., Szabo, W., Gerican, C., Systems of data acquisition

and processing, Transilvania University of Brasov,

1997

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!