06.03.2013 Views

〈85〉 bacterial endotoxins test - US Pharmacopeial Convention

〈85〉 bacterial endotoxins test - US Pharmacopeial Convention

〈85〉 bacterial endotoxins test - US Pharmacopeial Convention

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Interim Revision Announcement<br />

4 〈85〉 Bacterial Endotoxins Test Official April 1, 2011<br />

Table 4. Preparation of Solutions for the Inhibition/Enhancement Test for Photometric Techniques<br />

Solution to Which<br />

Solution Endotoxin Concentration Endotoxin Is Added Number of Replicates<br />

A a None Sample Solution Not less than 2<br />

B b Middle concentration of the standard curve Sample Solution Not less than 2<br />

C c At least 3 concentrations (lowest concentration is Water for BET Each not less than 2<br />

designated λ)<br />

D d None Water for BET Not less than 2<br />

a Solution A: The Sample Solution may be diluted not to exceed MVD.<br />

b Solution B: The preparation under <strong>test</strong> at the same dilution as Solution A, containing added endotoxin at a concentration equal to or near the middle of the<br />

standard curve.<br />

c Solution C: The standard endotoxin at the concentrations used in the validation of the method described for Assurance of Criteria for the Standard Curve<br />

under Preparatory Testing (positive controls).<br />

d Solution D: Water for BET (negative control).<br />

PHOTOMETRIC QUANTITATIVE TECHNIQUES<br />

Turbidimetric Technique<br />

concentrations within the range indicated by the lysate<br />

manufacturer to generate the standard curve. Perform the<br />

assay using at least three replicates of each standard endotoxin<br />

concentration according to the manufacturer’s instructions<br />

for the lysate (volume ratios, incubation time, temperature,<br />

pH, etc.). If the desired range is greater than two logs<br />

This technique is a photometric assay measuring increases in the kinetic methods, additional standards should be inin<br />

reactant turbidity. On the basis of the particular assay cluded to bracket each log increase in the range of the stanprinciple<br />

employed, this technique may be classified as ei- dard curve. The absolute value of the correlation coeffither<br />

an endpoint-turbidimetric assay or a kinetic-turbidimet- cient, r, must be greater than or equal to 0.980, for the<br />

ric assay. The endpoint-turbidimetric assay is based on the range of endotoxin concentrations set up.<br />

quantitative relationship between the concentration of <strong>endotoxins</strong><br />

and the turbidity (absorbance or transmission) of<br />

the reaction mixture at the end of an incubation period.<br />

The kinetic-turbidimetric assay is a method to measure either<br />

the time (onset time) needed to reach a predetermined<br />

absorbance or transmission of the reaction mixture, or the<br />

rate of turbidity development. The <strong>test</strong> is carried out at the<br />

incubation temperature recommended by the lysate manufacturer<br />

(which is usually 37 ± 1°).<br />

Test for Interfering Factors—Select an endotoxin con-<br />

centration at or near the middle of the endotoxin standard<br />

curve. Prepare Solutions A, B, C, and D as shown in Table 4.<br />

Perform the <strong>test</strong> on Solutions A, B, C, and D at least in dupli-<br />

cate according to the instructions for the lysate employed,<br />

for example, concerning volume of Sample Solution and Ly-<br />

sate TS, volume ratio of Sample Solution to Lysate TS, incu-<br />

bation time, etc.<br />

The <strong>test</strong> is considered valid when the following conditions<br />

are met.<br />

Chromogenic Technique<br />

1. The absolute value of the correlation coefficient of the<br />

standard curve generated using Solution C is greater than or<br />

This technique is an assay to measure the chromophore<br />

released from a suitable chromogenic peptide by the reaction<br />

of <strong>endotoxins</strong> with lysate. On the basis of the particular<br />

assay principle employed, this technique may be classified as<br />

either an endpoint-chromogenic assay or a kinetic-chromogenic<br />

assay. The endpoint-chromogenic assay is based on<br />

the quantitative relationship between the concentration of<br />

<strong>endotoxins</strong> and the release of chromophore at the end of<br />

an incubation period. The kinetic-chromogenic assay is a<br />

method to measure either the time (onset time) needed to<br />

reach a predetermined absorbance of the reaction mixture,<br />

or the rate of color development. The <strong>test</strong> is carried out at<br />

the incubation temperature recommended by the lysate<br />

manufacturer (which is usually 37 ± 1°).<br />

equal to 0.980.<br />

2. The result with Solution D does not exceed the limit of<br />

the blank value required in the description of the lysate rea-<br />

gent employed, or it is less than the endotoxin detection<br />

limit of the lysate reagent employed.<br />

Calculate the mean recovery of the added endotoxin by<br />

subtracting the mean endotoxin concentration in the solu-<br />

tion, if any (Solution A, Table 4), from that containing the<br />

added endotoxin (Solution B, Table 4). In order to be consid-<br />

ered free of factors that interfere with the assay under the<br />

conditions of the <strong>test</strong>, the measured concentration of the<br />

endotoxin added to the Sample Solution must be within<br />

50% to 200% of the known added endotoxin concentration<br />

after subtraction of any endotoxin detected in the solution<br />

without added endotoxin.<br />

When the endotoxin recovery is out of the specified<br />

Preparatory Testing<br />

range, the Sample Solution under <strong>test</strong> is considered to contain<br />

interfering factors. Then, repeat the <strong>test</strong> using a greater<br />

To assure the precision or validity of the turbidimetric and<br />

chromogenic techniques, preparatory <strong>test</strong>s are conducted to<br />

verify that the criteria for the standard curve are valid and<br />

that the sample solution does not interfere with the <strong>test</strong>.<br />

Validation for the <strong>test</strong> method is required when conditions<br />

that are likely to influence the <strong>test</strong> result change.<br />

Assurance of Criteria for the Standard Curve—The <strong>test</strong><br />

must be carried out for each lot of lysate reagent. Using the<br />

Standard Endotoxin Solution, prepare at least three endotoxin<br />

dilution, not exceeding the MVD. Furthermore, interference<br />

of the Sample Solution or diluted Sample Solution not to ex-<br />

ceed the MVD may be eliminated by suitable validated<br />

treatment, such as filtration, neutralization, dialysis, or heat<br />

treatment.To establish that the treatment chosen effectively<br />

eliminates interference without loss of <strong>endotoxins</strong>, perform<br />

the assay described above using the preparation to be examined<br />

to which Standard Endotoxin has been added and<br />

which has then been submitted to the chosen treatment.<br />

© 2011 The United States <strong>Pharmacopeial</strong> <strong>Convention</strong> All Rights Reserved.

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!