14<br />
Year(s) Country Economic group Stage of therapy Cause/contributing<br />
factors of error<br />
1999–2003 Japan High income Planning Output factor input error in renewal<br />
of treatment planning system<br />
1999–2004 Treatment Insufficient dose delivery caused<br />
delivery by an incorrect operation<br />
of dosimeter<br />
2000–2001 Panama Upper middle Planning Error shielding block related<br />
income data entry into TPS resulted<br />
in prolonged treatment time<br />
2001 Poland Upper middle Treatment Failure of safety system on a<br />
income delivery Linac after power failure<br />
2003 Japan High income Planning & Input error of combination of<br />
Information transfer total dose and<br />
transfer fraction number<br />
2003–2004 Planning & Misapplication of tray factor<br />
Information to treatment delivery without tray<br />
transfer<br />
2004–2005 France High income Planning Wrong setting of the linear<br />
accelerator after introduction of<br />
new treatment planning system<br />
(TPS) (static wedges changes to<br />
dynamic wedges but dose intensity<br />
modification not done)<br />
Information Miscommunication of field size<br />
transfer & estimation, error in patient<br />
Treatment identification, incorrect<br />
delivery implantation of source during<br />
brachytherapy<br />
2004–2007 Canada High income Planning Incorrect output determinations<br />
for field sizes other than the<br />
calibration field size for superficial<br />
skin treatments.<br />
2005–2006 UK High income Planning Change in operational procedures<br />
while upgrading the data<br />
management system resulting in<br />
incorrect treatment dose