How flow is estimated<br />
• Flow is estimated from known:<br />
• pump speed<br />
• pump power<br />
• blood viscosity (hematocrit = HCT)<br />
• Accurate flow estimate depends on correct HCT range being<br />
entered<br />
• Extremely high flows with no apparent cause are suspicious of pump<br />
problems<br />
Hb (g/L) X 0.3 ~ HCT eg. Hb 100 X 0.3 ~ 30<br />
Hb (g/dL) X 3 ~ HCT eg. Hb 10 X 3 ~ 30<br />
The contents of this document are confidential and are not to be reproduced or made available to third parties without prior consent from Ventracor.<br />
Caution: Investigational device - this device is limited by law to investigational use only in the US, Australia and NZ (under Federal Law in the US)<br />
Effect of treatments on flow<br />
Pump is preload-dependent and afterload-sensitive<br />
If preload remains the same but ! afterload, flows "<br />
If ! preload and afterload remains the same, flows !<br />
Any treatment which ! cardiac output should !<br />
pump output<br />
eg Inotropes, intravascular filling, pacing<br />
The contents of this document are confidential and are not to be reproduced or made available to third parties without prior consent from Ventracor.<br />
Caution: Investigational device - this device is limited by law to investigational use only in the US, Australia and NZ (under Federal Law in the US)<br />
Assistance provided by the pump<br />
Flow depends on:<br />
• volume (coming into the pump)<br />
• pressure (preload and afterload)<br />
• speed (of the pump)<br />
The contents of this document are confidential and are not to be reproduced or made available to third parties without prior consent from Ventracor.<br />
Caution: Investigational device - this device is limited by law to investigational use only in the US, Australia and NZ (under Federal Law in the US)<br />
Right Heart Support<br />
• Minimize blood products<br />
• Blood products =<br />
"RV distension<br />
"Release of proinflammatory cytokines = PH<br />
"RV has to work harder, " difficulty moving blood to LV<br />
"liver congestion – cycle<br />
"Possibility lead to hypercoaguability<br />
• Support right heart with inotropes<br />
• Lower right heart afterload<br />
The contents of this document are confidential and are not to be reproduced or made available to third parties without prior consent from Ventracor.<br />
Caution: Investigational device - this device is limited by law to investigational use only in the US, Australia and NZ (under Federal Law in the US)