<strong>laser</strong> assembly (TLA). The TLA includes a microprocessor and other control electronics<br />

to support <strong>laser</strong> commands (Fig.7). These commands may be accessed using the<br />

evaluation software or directly through the TLA interface.<br />

III. EXPERIMENTS<br />

Fig.7. Evaluation board with mounted SG-DBR <strong>laser</strong>.<br />

You will be using Anritsu Optical Spectrum Analyzer (OSA), model MS9710B.<br />

Switch-on Procedure of the Optical Spectrum Analyzer.<br />

• Switch-on the line power of OSA.<br />

• Set the spectrum analyzer center wavelength to 1520 nm and the span to 200 nm.<br />

• Set the wavelength resolution to the maximum of 0.07 nm.<br />

• Set the vertical scale to be logarithmic.<br />

A. Spectral Characteristics of LD<br />

In this experiment you will be measuring the optical spectrum of a LD, operating in the<br />

near-infrared (NIR) region with a center wavelength around 1550 nm.<br />

- Use the fiber patch-cord to couple the light from the LD to the OSA.<br />

Note that the fiber connector has notch, and correctly fits into the receptacle in only<br />

one orientation.<br />

- Set ILD current at 5mA and observe the optical spectrum of the emitted light.<br />

- Measure the center wavelength and bandwidth of the output spectrum on -3db (FWHM)<br />

level.<br />

- Set ILD current at 50mA and repeat measurements<br />

6