ECE 183 2012 Lab #3 Tunable laser diode (Agility)
ECE 183 2012 Lab #3 Tunable laser diode (Agility)
ECE 183 2012 Lab #3 Tunable laser diode (Agility)
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>laser</strong> assembly (TLA). The TLA includes a microprocessor and other control electronics<br />
to support <strong>laser</strong> commands (Fig.7). These commands may be accessed using the<br />
evaluation software or directly through the TLA interface.<br />
III. EXPERIMENTS<br />
Fig.7. Evaluation board with mounted SG-DBR <strong>laser</strong>.<br />
You will be using Anritsu Optical Spectrum Analyzer (OSA), model MS9710B.<br />
Switch-on Procedure of the Optical Spectrum Analyzer.<br />
• Switch-on the line power of OSA.<br />
• Set the spectrum analyzer center wavelength to 1520 nm and the span to 200 nm.<br />
• Set the wavelength resolution to the maximum of 0.07 nm.<br />
• Set the vertical scale to be logarithmic.<br />
A. Spectral Characteristics of LD<br />
In this experiment you will be measuring the optical spectrum of a LD, operating in the<br />
near-infrared (NIR) region with a center wavelength around 1550 nm.<br />
- Use the fiber patch-cord to couple the light from the LD to the OSA.<br />
Note that the fiber connector has notch, and correctly fits into the receptacle in only<br />
one orientation.<br />
- Set ILD current at 5mA and observe the optical spectrum of the emitted light.<br />
- Measure the center wavelength and bandwidth of the output spectrum on -3db (FWHM)<br />
level.<br />
- Set ILD current at 50mA and repeat measurements<br />
6