7 Battery Care Information<br />

Optimizing Battery Service Life<br />

Battery Preventive Maintenance Program<br />

Charging the Batteries<br />

-<br />

Ensuring the following procedures are a part of routine device use can<br />

optimize battery service life:<br />

. Connect pumps to AC power at all times except in the event of<br />

AC power loss or short-term portable operation.<br />

. Store pumps plugged in to AC power to maintain the battery<br />

charge whenever possible. See the Storage Temperature table in<br />

dtis chapter for information on the effects of temperature on the<br />

b"attery storage time.<br />

. Notify Central Supply or other appropriate departments as soon as<br />

a pump is removed from patient use so that it can be cleaned and<br />

the batteries can be recharged.<br />

. Recharge a pump that is in the Battery Low alert condition for a<br />

minimum of 12 uninterrupted lK>urs.<br />

The most reliable way to check the condition (capacity) of the pump's<br />

batteries at any time during their life is to measure discharge time. The<br />

batteries should provide an operating ti~ of no less dIaD 25 hours<br />

(Colleague pump) or 2 hours (Colleague 3 pump) from fully charged to<br />

Battery Depleted alarm wIY:n all channels infuse at 100 mI.Jbr on<br />

battery power. This test is described in the Colleague Pump Global<br />

Service Manual. Battery packs that do not pass the test are at the end of<br />

their service life and must be replaced. Implementing a program to<br />

replace the batteries prior to the projected end of their service life can<br />

reduce battery problems.<br />

For OOditional infomJation about Iead-ocid battery use in <strong>Baxter</strong> infusion<br />

devices, contact <strong>Baxter</strong> <strong>Medication</strong> <strong>Delivery</strong> Services at<br />

1-800- THE-PUMP.<br />

The batteries are charging whenever the pump is plugged into a 100-120<br />

VAc. 5