Technical Data<br />

THERMOVAC Transmitter<br />

Display range<br />

Measurement uncertainty<br />

Principle of measurement<br />

Supply voltage<br />

Power consumption<br />

mbar (Torr)<br />

VA<br />

TTR 211 PB/D / TTR 216 PB/D<br />

5 x 10 -4 to 1000 (3.75 x 10 -4 to 750)<br />

20% in the range 1 x 10 -3 to 1 x 10 -2 mbar (0.75 x 10 -3 to 0.75 x 10 -2 Torr)<br />

15% in the range 1 x 10 -2 to 30 mbar (0.75 x 10 -2 to 22.5 Torr)<br />

Thermal conductivity acc. to Pirani<br />

14.5 to 36 V DC, typ. 24 V DC<br />

Ripple voltage < 2 V pp<br />

< 2<br />

Storage / nominal temperature range °C<br />

-20 to +70 / +10 to +50<br />

Max. rel. humidity (climatic class F)<br />

Protection class<br />

Dimensions (H x W x D)<br />

Weight, approx.<br />

Inflammability<br />

Sensor<br />

Filament<br />

<strong>Vacuum</strong> connection<br />

% n.c.<br />

mm<br />

kg (lbs)<br />

DN<br />

95<br />

IP 40<br />

106 x 73 x 56<br />

0.29 (0.64)<br />

UL 94 - V 2<br />

Exchangeable sensing cell<br />

Tungsten (TTR 211)/platinum (TTR 216)<br />

16 KF<br />

Bake out temperature, max °C<br />

Dead volume, max. cm 3<br />

Materials in contact with the medium<br />

Over-pressure rating, abs.<br />

Signal output (R a<br />

≥ 10 kΩ)<br />

Measurement signal<br />

Status signal<br />

80 at the flange<br />

11<br />

TTR 211:<br />

Aluminum, nickel-plated steel, Vacon, tungsten, CrNi8020, glass, epoxy cement;<br />

TTR 216 S:<br />

Stainless steel, CrNi, Al 2 O 3 ceramics, NiFe, Mo, Ni, platinum<br />

TTR 211: 3 bar, TTR 216: 10 bar<br />

0 to 10.6 V<br />

1.9 to 10 V, corresp. 5 x 10 -4 to 1 x 10 3 mbar<br />

logarithmic divisions 1.286 V/decade<br />

broken filament 10.5 V<br />

Trigger<br />

Adjustment range<br />

Hysteresis<br />

Reaction time<br />

Rating<br />

Error status<br />

Status indicators<br />

Monitor output (R a<br />

≥ 100 kΩ)<br />

Electrical connection<br />

Cable length, max.<br />

Interface<br />

TTR 211 D / 216 D<br />

TTR 211 PB / 216 PB<br />

mbar (Torr)<br />

ms<br />

m<br />

n.o./changeover relay contact<br />

1 x 10 -3 to 500 (0.75 to 375)<br />

about 30% of the adjusted pressure<br />

< 50<br />

60 V, 0.5 A DC<br />

contact open in case of broken filament or supply off<br />

Operation (Power): Orange LED<br />

Trigger (active): Green LED<br />

Jack socket (3.5 mm) at which the measurement signal and the trigger setting is available<br />

FCC-68 socket, 8 way with shield<br />

100<br />

DeviceNet<br />

Profibus DP<br />

C16 28<br />