EXTRACT PREPARATION<br />

The extraction vehicles and conditions <strong>of</strong> extraction used shall<br />

be appropriate to the nature and use <strong>of</strong> the final product and to<br />

the purpose <strong>of</strong> the test<br />

a) (37 ± 1) °C for (72 ± 2) h;<br />

b) (50 ± 2) °C for (72 ± 2) h;<br />

c) (70 ± 2) °C for (24 ± 2) h;<br />

d) (121 ± 2) °C for (1 ± 0,1) h.<br />

The increased temperature may cause cross-linking and/or<br />

polymerization <strong>of</strong> the polymer and, therefore, decrease the amount <strong>of</strong><br />

free monomer that is available to migrate from the polymer<br />

<br />

The increased temperature could cause degradation products to form<br />

that are not typically found in the finished device under conditions <strong>of</strong><br />

use<br />

Harlan Laboratories 17