EXTRACT PREPARATION<br />
The extraction vehicles and conditions <strong>of</strong> extraction used shall<br />
be appropriate to the nature and use <strong>of</strong> the final product and to<br />
the purpose <strong>of</strong> the test<br />
a) (37 ± 1) °C for (72 ± 2) h;<br />
b) (50 ± 2) °C for (72 ± 2) h;<br />
c) (70 ± 2) °C for (24 ± 2) h;<br />
d) (121 ± 2) °C for (1 ± 0,1) h.<br />
The increased temperature may cause cross-linking and/or<br />
polymerization <strong>of</strong> the polymer and, therefore, decrease the amount <strong>of</strong><br />
free monomer that is available to migrate from the polymer<br />
<br />
The increased temperature could cause degradation products to form<br />
that are not typically found in the finished device under conditions <strong>of</strong><br />
use<br />