How do you measure how good a sphere is?<br />

• Sphere Multiplier “M” – a measure of the efficiency of<br />

the sphere.<br />

• Calculated from the reflectivity and the port fraction.<br />

• Bigger is better<br />

• M isn’t everything<br />

• Think of the light being diluted across the surface of the sphere<br />

• Light is more concentrated in a smaller sphere<br />

Sphere diameter<br />

Sphere Surface<br />

Area (mm 2 )<br />

Port Fraction<br />

f<br />

Sphere Multiplier<br />

M<br />

% of incident light falling on<br />

64 mm 2 detector port<br />

60 mm 11310 11.3 8.1 4.6<br />

70 mm 15394 9.3 9.7 4.0<br />

75 mm 17671 3.6 21.7 7.9<br />

150 mm 70686 2.5 28.5 2.6<br />

Erich Heiden UV/Vis Salesmanager<br />