Data from the dedicated positive <strong>and</strong> negative experiment were<br />

compared to data from the polarity switching experiment. Peak<br />

areas <strong>and</strong> signal-to-noise (S/N) were compared, as well as %CV<br />

of the peak areas.<br />

Intensity, cps<br />

Intensity, cps<br />

Intensity, cps<br />

Intensity, cps<br />

XIC of +MRM (78 pairs): 195.5/13... Max. 2.3e4 cps.<br />

2.0e4<br />

1.0e4<br />

(+) Caffeine<br />

6.9<br />

1.9e4<br />

(+/-)Caffeine<br />

S/N = 24.7 1.0e4 S/N = 20.0<br />

0.0<br />

5.0 5.5 6.0 6.5 7.0 7.5<br />

Time, min<br />

8.0 8.5<br />

XIC of +MRM (78 pairs): 235.2/86... Max. 1.1e6 cps.<br />

1.00e6<br />

5.00e5<br />

(+) Lidocaine<br />

6.5<br />

1.0e6<br />

(+/-) Lidocaine<br />

S/N = 705<br />

5.0e5<br />

S/N = 922<br />

0.00<br />

4.5 5.0 5.5 6.0 6.5 7.0<br />

Time, min<br />

7.5 8.0 8.5<br />

XIC of +MRM (78 pairs): 213.1/72... Max. 9.1e4 cps.<br />

9.1e4<br />

5.0e4<br />

0.0<br />

9.0 9.5 10.0 10.5 11.0<br />

Time, min<br />

11.5 12.0 12.5<br />

XIC of -MRM (33 pairs): 312.9/15... Max. 2.8e5 cps.<br />

2.8e5<br />

2.0e5<br />

1.0e5<br />

0.0<br />

Variations between the sets of data were minimal, demonstrating<br />

that no significant loss of data quality was observed when a<br />

positive/negative switching method was used versus a dedicated<br />

method (see Figure 1-4).<br />

Dedicated polarity Positive negative switching<br />

XIC of +MRM (78 pairs): Exp 1, 1... Max. 1.9e4 cps.<br />

0.0<br />

5.0 5.5 6.0 6.5 7.0 7.5<br />

Time, min<br />

8.0 8.5<br />

XIC of +MRM (78 pairs): Exp 1, 2... Max. 1.0e6 cps.<br />

Figure 3. Comparison of S/N <strong>for</strong> representative analytes: Data from the dedicated positive or negative experiment is shown on the left <strong>and</strong> data acquired<br />

using a positive/negative switching experiment is shown on the right. No significant degradation of data quality was observed when using the polarity<br />

switching experiment<br />

Intensity, cps<br />

Intensity, cps<br />

6.9<br />

6.5<br />

0.0<br />

4.5 5.0 5.5 6.0 6.5 7.0<br />

Time, min<br />

7.5 8.0 8.5<br />

XIC of +MRM (78 pairs): Exp 1, 2... Max. 9.1e4 cps.<br />

10.7<br />

(+) Chlorotoluron<br />

9.1e4<br />

10.7<br />

(+/-) Chlorotoluron<br />

S/N = 52.6 5.0e4 S/N = 48.8<br />

11.5 12.0 12.5 13.0 13.5 14.0 14.5 15.0<br />

Time, min<br />

Intensity, cps<br />

0.0<br />

9.0 9.5 10.0 10.5 11.0<br />

Time, min<br />

11.5 12.0 12.5<br />

XIC of -MRM (33 pairs): Exp 2, 3... Max. 1.9e5 cps.<br />

(-) Triclocarban<br />

13.3<br />

1.9e5<br />

13.3<br />

(+/-) Triclocarban<br />

S/N = 869<br />

1.0e5<br />

S/N = 592<br />

Intensity, cps<br />

0.0<br />

11.5 12.0 12.5 13.0 13.5 14.0 14.5 15.0<br />

Time, min<br />

p 3