WSRC-TR-2000-00182<br />

MetLab Treatability Study: June 2000 Update<br />

1312 photo-acoustic infrared gas analyzer (new model <strong>of</strong> the Bruel and Kjaer Model<br />

1302). The instrument was calibrated to measure TCE, PCE, Freon 113 and CO 2 . The<br />

Freon-113 interference problems on the instrument were corrected in January 2000<br />

and a reliable measurement is now possible with the infrared instrument. The correction<br />

was done by installing an additional band pass optical filter in the instrument to<br />

measure Freon-113 and to compensate for interference <strong>of</strong> the Freon-113 on the optical<br />

filters used to measure TCE and PCE. Samples were also obtained in Tedlar sample<br />

bags and analyzed on an HP 5890 series gas chromatograph (GC) to verify the<br />

reliability <strong>of</strong> the infrared instrument.<br />