Probes<br />

Standard probes HC2-S / HC2-S3<br />

The HC2-S / HC2-S3 is the most versatile probe from ROTRONIC and forms the basis of the<br />

product portfolio. It measures humidy and temperature and calculates the dew/frost point.<br />

HC2-S<br />

HC2-S3<br />

Applications<br />

HVAC, food industry, building services equipment, paper, textile and pharmaceutical industries<br />

Features<br />

• Accuracy: ±0.8 %RH, ±0.1 K, at 23 °C ±5 K<br />

• Range of application: -50…100 °C / 0…100 %RH<br />

• Digital interface (UART) and scalable analog outputs, 0…1 V<br />

• Standard output scaling: 0…1 V = -40…60 °C / 0…100 %RH<br />

• Adjusted at 23 °C and 10, 35, 80 %RH<br />

83<br />

15<br />

Order code HC2-S HC2-S3<br />

Probe type Standard probe, black Meteorology probe, white<br />

Dimensions<br />

Range of application<br />

Accuracy<br />

Power supply<br />

Current consumption<br />

Long-term stability<br />

Sensor type<br />

Filter type<br />

Response time<br />

Maximum wind<br />

velocity<br />

Housing material<br />

Weight<br />

Ø 15 x 83 mm<br />

-50…100 °C, 0…100 %RH<br />

±0.8 %RH, ±0.1 K, at 23 °C ±5 K<br />

3.2…5 VDC, calibrated at 3.3 VDC<br />

~4.5 mA<br />