02.12.2014 Views

PIDs As HazMat Response Tools

PIDs As HazMat Response Tools

PIDs As HazMat Response Tools

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Addressing PID Sensor Moisture Sensitivity<br />

Technical Note TN-163<br />

rev 5 ab.05-07<br />

The RAE Systems’ photoionization detectors (<strong>PIDs</strong>)<br />

were designed to have little or no response to<br />

moisture. However, when the sensor is dirty, bent<br />

out of shape, or corroded, RAE Systems’ monitors<br />

may show a strong response to humidity at or above<br />

about 85% RH (relative humidity).<br />

Dirty Sensor Causes Current Leakage<br />

A dirty sensor is the number-one reason for humidity<br />

response. The sensor has two electrodes: a bias<br />

electrode and a sensing electrode. One has a greater<br />

positive charge than the other. <strong>As</strong> the UV light<br />

breaks down chemicals into positive ions and<br />

negative electrons, they migrate to the oppositely<br />

charged electrodes. With clean, dry air and sensor<br />

components, no current can leak across the air space<br />

between the two electrodes.<br />

Clean Sensor<br />

Bias Electrode<br />

Sensing Electrode<br />

No dirt build-up to<br />

foster a decrease in<br />

airspace resistance<br />

However, even microscopic dirt accumulations<br />

on the<br />

electrodes and Teflon parts can promote leakage,<br />

especially when the relative humidity is greater than<br />

85%. This is because water adheres more readily<br />

onto dust particles than onto a clean Teflon or glass<br />

surface. A sensor may appear to be visually clean,<br />

but actually can be dirty enough to cause current<br />

(electron) leakage. This current leakage is<br />

interpreted by the PID as a drifting reading of<br />

typically more than 10, and as much as hundreds of<br />

ppm, with no VOC present (over 1 ppm for a<br />

ppbRAE).<br />

Dirty Sensor<br />

Bias Electrode<br />

Sensing Electrode<br />

Dirt build-up absorbs water and<br />

breaks down airspace resistance,<br />

leading to sensor “ leakage” or<br />

moisture response<br />

At very high levels of leakage due to a dirty sensor,<br />

the MiniRAE 2000 displays a “Clear Sensor, Quit?”<br />

message on the screen. Pressing the “Y” key tells the<br />

MiniRAE 2000 to continue. The “Clear Sensor,<br />

Quit?” message can also be displayed if the<br />

MiniRAE 2000 pump has been shut off with a high<br />

level of sample left in the instrument.<br />

Prevention is the Best Cure<br />

Preventing dirt and dust from entering the PID sensor<br />

is the best cure for moisture response. Make sure that<br />

all of the filters in the instrument are clean and<br />

effective. In dirty environments, it may be necessary<br />

to add an additional filter. However, under normal<br />

circumstances there should be no response to<br />

humidity with just the standard sintered metal and<br />

external filter installed. A Humidity Filtering Type II<br />

Tube can be used to dry the sample stream and<br />

measure VOCs at high humidity for about a half hour<br />

(see Technical Note TN-178). Many compounds can<br />

be measured this way, especially those commonly<br />

encountered in soil remediation such as gasoline and<br />

trichloroethylene. Caution needs to be taken because<br />

there are some<br />

very heavy compounds that can be<br />

lost on the tube<br />

Quick Humidity <strong>Response</strong> Tests<br />

Try exhaling gently into a MiniRAE 2000, ppbRAE,<br />

or ppbRAE plus for 10 to 15 seconds or cupping your<br />

hand over the inlet probe; moisture from your hand<br />

provides a fairly continuous high humidity stream.<br />

Be sure not to block the air flow. This test can also<br />

be preformed on the UltraRAE, MultiRAE and<br />

EntryRAE in a similar manner.<br />

1<br />

RAE Systems Inc.<br />

3775 N. First St., San Jose, CA 95134-1708 USA<br />

Phone: +1.888.723.8823<br />

Email: raesales@raesystems.com<br />

Web Site: www.raesystems.com


Technical Note TN-163<br />

rev 5 ab.05-07<br />

inside the probe, including a green dust filter and a<br />

white<br />

“C” filter. Both of these should be removed.<br />

The instrument should show little or no response<br />

from these tests. If the MiniRAE 2000 shows more<br />

than 5 ppm isobutylene units or the ppbRAE shows<br />

more than 500 ppb increase in reading, then<br />

something is wrong. The design of the instruments<br />

makes it quick and easy to remove and clean the<br />

probe, lamp, and sensor to address this problem.<br />

Focus on the following areas:<br />

1. Replace filters<br />

2. Clean the probe, PID lamp, and sensor<br />

3. Inspect the sensor for damage or corrosio n.<br />

4. Clean the sensor housing.<br />

5. Clean the sensor using an ultrasonic bath.<br />

MiniRAE 2000, ppbRAE Plus, and UltraRAE<br />

When testing the UltraRAE monitor, make sure that the<br />

tube of interest is installed before proceeding with the<br />

test. If you plan to test for Benzene, install the Benzene<br />

tube and use the same moisture test discussed for the<br />

MiniRAE 2000. If the readings increase more than 2<br />

ppm Benzene units, then there is a problem with dirt in<br />

the monitor.<br />

Replace Filter<br />

The MiniRAE 2000, ppbRAE, ppbRAE plus, and<br />

UltraRAE all use two filters, an external filter and a<br />

sintered metal filter. The external filter (PN 002-<br />

3022-005 for a pack of 5, 002-3022-010 for a pack of<br />

10, and PN 002-3022-100 for a pack of 100) is<br />

attached to the end of the probe and is white. If it<br />

appears to have any discoloration, then it is time to<br />

replace the filter. If the instrument goes into pump<br />

alarm with the filter attached and is able to run<br />

without any problems when not attached, it is time to<br />

replace the filter. When the instrument is responding<br />

to moisture, it is time to replace the filter. The metal<br />

sintered filter (PN 490-0047-005 for a pack of 5)<br />

should be replaced when it is dirty and when the<br />

instrument responds to moisture. The MiniRAE<br />

2000 and ppbRAE instruments have other filters<br />

Clean the Probe<br />

Unscrew the lamp housing cap; inspect the sintered<br />

metal filter at the bottom of the probe. If it appears to<br />

be dirty, it needs to be replaced. A dirty internal filter<br />

indicates that the probe has dirt inside and requires<br />

cleaning. Cleaning the probe depends on the kind of<br />

exposure or dirt that has contaminated the probe. If it is<br />

a chemical that might affect the reading of the PID, use<br />

methanol to clean the probe. If it is just dirt or dust that<br />

has been sucked into the instrument, blow clean, dry air<br />

through the probe to clear the internal tubing.<br />

Clean the PID Lamp<br />

Access to both the lamp and the sensor of a MiniRAE<br />

2000 and ppbRAE plus requires no tools. To remove<br />

the PID sensor, grasp the top lip of the sensor and pull<br />

straight out, to avoid bending the electrical pins on the<br />

sensor. If the sensor does not pull out immediately, a<br />

gentle rocking motion should help loosen the sensor.<br />

Also try rotating your grip as you pull the sensor. Once<br />

the PID sensor is removed, set it aside to clean or to<br />

inspect for damage or corrosion.<br />

The PID lamp has a white O-ring that keeps it from<br />

sitting too low in the lamp housing. Before reaching for<br />

the lamp, get out the lamp cleaning kit (PN 081-0002-<br />

000), which includes methanol, finger cots, cotton<br />

swabs, and lens tissue. Use finger cots on the fingers<br />

that will handle the lamp.<br />

CAUTION: Never touch the window surface with<br />

anything that might scratch, including your fingers, or<br />

any liquid that might leave a film.<br />

Carefully remove the lamp by grabbing onto the white<br />

O-ring, and set the instrument aside. Dip a cotton swab<br />

into the methanol and wipe the window surface of the<br />

lamp. The sides of the lamp can be cleaned if it is<br />

obviously dirty, but the flat-surface window of the lamp<br />

is the only part of the PID lamp that requires cleaning.<br />

Once the lamp is cleaned, use a piece of the lens tissue<br />

to dry and polish the lamp before dropping it back into<br />

the instrument. If there is visible dirt, clean the lamp<br />

housing prior to placing the lamp back into the<br />

instrument. Do not force the lamp all the way down to<br />

the bottom of the lamp housing. Allow the PID sensor<br />

to push the lamp the rest of the way into place.<br />

2<br />

RAE Systems Inc.<br />

3775 N. First St., San Jose, CA 95134-1708 USA<br />

Phone: +1.888.723.8823<br />

Email: raesales@raesystems.com<br />

Web Site: www.raesystems.com


Technical Note TN-163<br />

rev 5 ab.05-07<br />

To access the UltraRAE lamp for cleaning, you will<br />

need to first remove any RAE-SEP tube from the probe.<br />

Then unscrew the four screws holding the bar code<br />

reader cap to the instrument base. Lift the Bar Code<br />

Reader cap straight off of the instrument base until the<br />

PCB (printed circuit board) clears. Set it aside and<br />

make sure that the screws do not fall out and disappear.<br />

Take the silver sensor adapter off of the PID senso r.<br />

Underneath is a rubber tube connector and a silver<br />

sintered filter. Remove the rubber tube connector with<br />

filter. If it appears to be dirty, it needs to be replaced. A<br />

dirty internal filter indicates that the probe has dirt<br />

inside and requires cleaning. Cleaning the probe<br />

depends on the kind of exposure or dirt that has<br />

contaminated the probe. If it is a chemical that might<br />

affect the reading of the PID, use methanol to clean the<br />

probe. If it is just dirt or dust that has been sucked into<br />

the instrument, blow clean, dry air through the probe to<br />

clear the internal tubing.<br />

bent or the Teflon mask is warped, the sensor needs<br />

replacement. Note: A warped Teflon mask can be a<br />

sign that the sensor has been exposed to too much heat.<br />

Straight Electrodes<br />

with mask removed<br />

for clarity<br />

Good Sensor<br />

(partly disassembled)<br />

Warped<br />

Teflon<br />

Mask<br />

Damaged Sensor<br />

Bent<br />

Electrode<br />

If the sensing electrode is exposed to UV light from the<br />

lamp, it gives an undesired positive response to the<br />

light. If the fingers look aligned but are not in the same<br />

plane (some fingers are bent towards or away from the<br />

lamp), then the sensor should be replaced. The fingers<br />

should be close to, but not touching, the Teflon mask.<br />

In some cases it is possible to carefully bend the<br />

electrodes back into the correct position if the damage is<br />

very minor.<br />

Clean the Sensor Housing<br />

If the PID sensor is not damaged, thoroughly wipe clean<br />

the sensor housing area using a cotton swab dipped in<br />

lamp-cleaning methanol. Let the sensor housing area<br />

dry completely. Reassemble the PID, and perform a<br />

humidity response test. If it shows no response,<br />

calibrate the instrument and return it to service.<br />

Follow the directions for the MiniRAE 2000 and<br />

ppbRAE plus for PID sensor and lamp removal. The<br />

UltraRAE PID sensor and lamp are the same as those in<br />

the MiniRAE 2000.<br />

Sensor<br />

Housing<br />

Inspect the Sensor for Damage or Corrosion<br />

If the sensor appears corroded, it should be replaced. If<br />

the sensor is damaged or even slightly bent, it can show<br />

increased response to moisture. The bias and sensing<br />

electrodes are two sets of “fingers” in the MiniRAE<br />

2000 sensor. These fingers should be straight and<br />

parallel. The bias electrode is easy to see when the<br />

bottom of the sensor is inspected. However, the sensing<br />

electrode should be blocked from sight (and the UV<br />

light from the PID lamp) by a Teflon mask. If the<br />

sensing electrode can be seen because the fingers are<br />

Cleaning the PID Sensor<br />

Cleaning the lamp and sensor surfaces using the lamp<br />

cleaning kit with cotton swabs is not good enough to<br />

remove humidity response. In fact, cotton swabs can<br />

leave behind fibers that continue to cause moisture<br />

issues after cleaning. An ultrasonic cleaning bath<br />

MUST be used. Ultrasonic cleaners are available<br />

from laboratory suppliers. Alternatively, a jewelry<br />

3<br />

RAE Systems Inc.<br />

3775 N. First St., San Jose, CA 95134-1708 USA<br />

Phone: +1.888.723.8823<br />

Email: raesales@raesystems.com<br />

Web Site: www.raesystems.com


Technical Note TN-163<br />

rev 5 ab.05-07<br />

cleaner from a department store can be used. To<br />

clean the sensor, remove it from the instrument<br />

and<br />

immerse it in lamp-cleaning solution (anhydrous<br />

methanol) in a small beaker or baby food jar. Fill the<br />

ultrasonic cleaning bath with water. Place the beaker<br />

containing the sensor into the water bath of the<br />

ultrasonic cleaner. The ultrasonic waves pass through<br />

the water and container and assist in cleaning the<br />

sensor. The lamp-cleaning solution in the jar can be<br />

reused a few times until it becomes dirty. Remove the<br />

through the tube easily. Some heavy vapors such as<br />

diesel fuel or polar compounds such as acetone and<br />

alcohols require extended sampling times. See<br />

Technical Note TN-178 for more details and cautions<br />

on use. (Note that Type I Humidity Filtering tubes<br />

cannot be used during measurements.)<br />

Flex-I-Probe<br />

Tube Tip Breaker<br />

(p/n 021-2400-100)<br />

Tube Adapter<br />

(p/n 025-3002-000)<br />

Humidity Filtering II Tube<br />

(10-pack, p/n 025-2002-010)<br />

Water Bath<br />

Methanol<br />

in Jar<br />

Ultrasonic Cleaner<br />

sensor and allow it to dry for at least a few hours, and<br />

preferably overnight (warm but not hot air will speed<br />

the drying process).<br />

Caution! Do not overheat the sensor, and never<br />

disassemble the sensor!<br />

It is very important to blow or shake out any residual<br />

lamp-cleaning solution from the sensor before letting it<br />

air dry. Otherwise, some of the dirt that was just<br />

extracted into the cleaning solvent could be deposited<br />

back onto the sensor components as they dry.<br />

Reinstall the clean, dry sensor into the MiniRAE 2000.<br />

Exhale into the instrument and check the response. If it<br />

shows no response to the moisture in your breath,<br />

calibrate the instrument and return it to service.<br />

Using Humidity Filtering II tubes<br />

Humidity Filtering Tubes (Type II) can be used to dry<br />

the sample gas and measure many VOCs for about 15<br />

minutes to 1 hour. Their use prevents moisture effects,<br />

even on a dirty sensor. The tubes (p/n 025-2002-010)<br />

are inserted into an adapter (PN 025-3002-000) attached<br />

to a Flex-I-Probe (PN 021-2400-000). Many<br />

compounds such as trichloroethylene and gasoline pass<br />

MultiRAE Plus and EntryRAE<br />

While the MultiRAE and EntryRAE do not normally<br />

have as many issues with dirt in PID sensor,<br />

contamination is possible. The moisture test is done<br />

by cupping your hand over the end or breathing<br />

gently into the end, being careful not to block the air<br />

flow. If the VOC reading increases over 5 ppm<br />

isobutylene, there is a problem with dirt.<br />

Filters<br />

The MultiRAE uses a smaller external filter (PN 008-<br />

3022-001 for a pack of 3, PN 008-3022-010 for a<br />

pack of 10, PN 008-3022-100 for a pack of 100, and<br />

PN 008-3014-000 for a single external filter with<br />

adapter for the silver MultiRAE). This white filter<br />

should be replaced when it is dirty and when the<br />

instrument responds to moisture. If the instrument<br />

goes into pump alarm with the filter attached and is<br />

able to run without any problems when it is not<br />

attached, it is time to replace the filter.<br />

The EntryRAE uses a small external filter (PN 490-<br />

0122-000), which should be replaced when it appears<br />

dirty or when the instrument goes into pump alarm<br />

with the filter attached and clears the pump alarm<br />

when the filter is removed.<br />

MultiRAE PLUS<br />

There are two versions of the MultiRAE Plus, silver<br />

housing and black housing. There are small differences<br />

when cleaning them. If there is a response to moisture,<br />

dirt has gotten to the lamp. Due to the gas flow within<br />

the instrument, if dirt has gotten to the lamp, it has<br />

probably gotten into the filter, pump, tubing (silver<br />

MultiRAE only) and the gas plate.<br />

4<br />

RAE Systems Inc.<br />

3775 N. First St., San Jose, CA 95134-1708 USA<br />

Phone: +1.888.723.8823<br />

Email: raesales@raesystems.com<br />

Web Site: www.raesystems.com


Technical Note TN-163<br />

rev 5 ab.05-07<br />

Pump<br />

Start by replacing the filter, and then work on cleaning<br />

the internal parts of the instrument. Loosen the two<br />

screws on the back of the instrument and remove the<br />

top housing cover. Unplug the battery. If you have a<br />

black-box MultiRAE, you have access to the gas plate,<br />

pump and sensors now. If you have a silver-box<br />

MultiRAE, you still need to unscrew the four screws<br />

holding the gas plate to the instrument. After loosening<br />

the four screws, pull the gas plate off of the silver stand-<br />

plugged into the<br />

offs. Unplug the tubing from the pump and the wiring<br />

PCB.<br />

To clean the tubing, remove both pieces of tubing<br />

from<br />

the instrument and the gas plate and immerse it in<br />

methanol. Agitate it by hand or in an ultrasonic bath.<br />

Remove it and let it dry overnight. The black<br />

MultiRAE does not have tubing, but there are O-rings<br />

in the adapter attached to the tubing. Make sure there is<br />

no dirt in the adapter or O-rings by removing the pump<br />

from the top housing, and unplug the pump from the<br />

adapter. Remove all eight O-rings and inspect the<br />

adapter and O-rings. Clean the adapter and O-rings<br />

with methanol. Dry the O-rings with a piece of the lens<br />

tissue that comes with lamp-cleaning kit. Clean the<br />

pump by removing the four screws that hold the head to<br />

the diaphragm. Use a cotton swab dipped in methanol<br />

to clean all surfaces, including the diaphragm.<br />

Reassemble the pump and set it aside to dry for 5<br />

minutes. Now that everything is removed from the gas<br />

plate, clean the gas plate by pushing air or methanol<br />

through the airflow path. Allow it to dry overnight.<br />

The PID sensor<br />

The PID sensor is covered by a shielding cover that<br />

needs to be removed. This silver cover can be dipped in<br />

methanol to be cleaned. To clean the sensor, remove it<br />

from the instrument and immerse it in lamp-cleaning<br />

solution (anhydrous methanol) in a small beaker or<br />

baby food jar.<br />

Agitate it by hand or in an ultrasonic cleaner. Fill the<br />

ultrasonic cleaning bath with water. Place the beaker<br />

containing the sensor into the water bath of the<br />

ultrasonic cleaner. The ultrasonic waves pass through<br />

the water and container and assist in cleaning the<br />

sensor. The lamp-cleaning solution in the jar can be<br />

reused a few times until it becomes dirty. Remove<br />

the PID sensor and let it dry for at least a few hours,<br />

and preferably overnight (warm but not hot air will<br />

speed the drying process).<br />

Caution! Do not overheat the sensor!<br />

It is very important to blow or shake out any residual<br />

lamp-cleaning solution from the sensor before letting it<br />

air dry. Otherwise, some of the dirt that was just<br />

extracted into the cleaning solvent could be deposited<br />

back onto the sensor components as they dry. Reinstall<br />

the clean, dry sensor into the MultiRAE. Exhale into<br />

the instrument and check the response. If it shows no<br />

response to the moisture in your breath, calibrate the<br />

instrument and return it to service.<br />

EntryRAE<br />

If moisture problems still persist, contact the RAE<br />

Systems Service Department at 888-723-4800.<br />

5<br />

RAE Systems Inc.<br />

3775 N. First St., San Jose, CA 95134-1708 USA<br />

Phone: +1.888.723.8823<br />

Email: raesales@raesystems.com<br />

Web Site: www.raesystems.com

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!