CANNON
CANNON
CANNON
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
BS/IP/SL SUSPENDED<br />
LEVEL VISCOMETERS<br />
For measurement of kinematic<br />
viscosity of transparent Newtonian<br />
liquids according to ASTM D 445 and<br />
ISO 3104. Specifications conform to<br />
ASTM D 446 and ISO 3105.<br />
Provided with certificate of<br />
calibration and instruction sheet.<br />
See page 2 for expanded<br />
uncertainty.<br />
Minimum sample volume 11 mL.<br />
Require liquid bath depth of<br />
250 mm (10 inches).<br />
Holder is not supplied. For holder,<br />
see 9726-M99 on page 15.<br />
Size<br />
Specifications for BS/IP/SL<br />
Suspended Level Viscometers<br />
Approx.<br />
Constant<br />
mm 2 /s 2 Viscosity Range,<br />
(cSt/s) mm 2 /sec (cSt)*<br />
Calibrated<br />
BS/IP/SL Series<br />
1 0.01 3.5 to 10 9724-F10<br />
1A 0.03 6 to 30 9724-F13<br />
2 0.1 20 to 100 9724-F16<br />
2A 0.3 60 to 300 9724-F19<br />
3 1.0 200 to 1000 9724-F22<br />
3A 3.0 600 to 3000 9724-F25<br />
4 10 2000 to 10 000 9724-F28<br />
4A 30 6000 to 30 000 9724-F31<br />
5 100 20 000 to 100 000 9724-F34<br />
* Ranges in centistokes represent flow times of 350 to 1000 seconds<br />
for Size 1 and 200 to 1000 seconds for all other sizes.<br />
9724-F10 Series<br />
Size<br />
BS/IP/SL(S)<br />
VISCOMETERS<br />
For measurement of kinematic<br />
viscosity of transparent Newtonian<br />
liquids according to ASTM D 445 and<br />
ISO 3104. Specifications conform to<br />
ASTM D 446 and ISO 3105.<br />
Provided with certificate of<br />
calibration and instruction sheet.<br />
See page 2 for expanded<br />
uncertainty.<br />
Minimum sample volume 10 mL.<br />
Require liquid bath depth of 230 mm (9 inches).<br />
Holder is not supplied. For holder, see 9726-M99 on page 15.<br />
Specifications for Shortened Form BS/IP/SL(S)<br />
Suspended Level Viscometers<br />
Approx.<br />
Constant,<br />
mm 2 /s 2 Viscosity Range, Calibrated<br />
(cSt/s) mm 2 /sec (cSt) BS/IP/SL(S) Series<br />
1 0.0008 1.05 minimum* 9724-F50<br />
2 0.003 2.1† to 3 9724-F53<br />
3 0.01 3.8‡ to 10 9724-F56<br />
4 0.03 6¤ to 30 9724-F59<br />
5 0.1 20 to 100 9724-F62<br />
6 0.3 60 to 300 9724-F65<br />
7 1.0 200 to 1000 9724-F68<br />
8 3.0 600 to 3000 9724-F71<br />
9 10 2000 to 10 000 9724-F74<br />
* 1 320 seconds minimum flow time.<br />
† 600 seconds minimum flow time to 1000 seconds.<br />
‡ 380 seconds minimum flow time to 1000 seconds.<br />
¤ Ranges in centistokes represent flow times of 200 to 1000 seconds<br />
for all other sizes.<br />
9724-F50 Series<br />
12<br />
<strong>CANNON</strong> INSTRUMENT COMPANY • 2139 HIGH TECH ROAD • STATE COLLEGE • PA • 16803 • USA<br />
800-676-6232 • 814-353-8000 • Fax 814-353-8007 • cannon@cannoninstrument.com • www.cannoninstrument.com