Can't Touch This! The Three W's of Barrier Isolator Technology
Can't Touch This! The Three W's of Barrier Isolator Technology
Can't Touch This! The Three W's of Barrier Isolator Technology
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
A<br />
Different <strong>Isolator</strong> Applications<br />
„Open <strong>Isolator</strong>s“ for continuous aseptic Operations<br />
• Classifications:<br />
Filling:<br />
Capping:<br />
Support Area:<br />
ISO 5 / Grade A<br />
ISO 7 / Grade C<br />
ISO 8 / Grade D<br />
D<br />
4 Pa<br />
E<br />
A<br />
40 Pa<br />
ISO 5<br />
• VHP sterilization for ISO 5 / Grade A<br />
- stopper bowls included in VHP sterilization<br />
- CIP / SIP for filling equipment<br />
4-6Pa.<br />
30Pa.<br />
• Caps and capping equipment:<br />
- no sterilisation<br />
- controlled disinfection<br />
B<br />
B<br />
30 Pa<br />
ISO 5<br />
20Pa.<br />
• Environmental Monitoring:<br />
- particles 0.5µm, 5.0µm<br />
- viable air monitoring<br />
-surfaces (RODAC)<br />
- Temp, diff. pressure, relative humidity<br />
-10 Pa<br />
0<br />
a-<br />
e<br />
-10Pa.<br />
D<br />
C<br />
H<br />
F<br />
10Pa.<br />
G<br />
E<br />
F<br />
G<br />
D<br />
10 Pa<br />
ISO 7<br />
C<br />
Andockstation<br />
f╧r Stopfen<br />
2270<br />
Hubs╠ule<br />
Stopfen<br />
2295<br />
G<br />
12