ECAT FOUNDATION EXERCISE <strong>2010</strong>-4<br />

INTERNATIONAL EXTERNAL QUALITY ASSESSMENT Page 3 of 11<br />

PROGRAMME IN HAEMOSTASIS AND THROMBOSIS Print date: 28 February 2011<br />

Copyright © <strong>2010</strong> ECAT Foundation<br />

Labcode: 900<br />

Overview ratio ECAT plasma / normal plasma<br />

RATIO SCREENING n mean CV Range<br />

APTT 253 1.43 16.8% 0.69 – 2.12<br />

APTT SP (IL) 32 1.24 8.1% 1.06 – 1.40<br />

HemosIL SynthASil (IL) 14 1.16 6.4% 1.06 – 1.33<br />

Pathromtin SL (Siemens) 24 1.35 8.1% 1.13 – 1.56<br />

Actin FSL (Siemens) 42 1.39 10.1% 0.99 – 1.74<br />

PTT Automate / STA APTT<br />

(Stago)<br />

24 1.71 9.9% 1.24 – 2.12<br />

PTT <strong>LA</strong> (Stago) 71 1.66 9.0% 1.21 – 1.95<br />

DRVVT 289 1.02 10.8% 0.79 – 2.10<br />

DVVtest (AD) 18 0.96 9.6% 0.82 – 1.17<br />

<strong>LA</strong>C screen (I.L.) 66 1.07 10.3% 0.91 – 1.59<br />

<strong>LA</strong> screen (Life Diagnostics) 26 1.03 8.5% 0.88 – 1.19<br />

<strong>LA</strong> check (Precision Biologic) 15 0.96 7.1% 0.79 – 1.08<br />

<strong>LA</strong>1 screen (Siemens) 105 1.00 13.0% 0.79 – 2.10<br />

DRVVT Screen (Stago) 38 0.99 6.1% 0.89 – 1.21<br />

KCT 22 1.05 26.7% 0.69 – 1.73<br />

dAPTT 16 1.40 22.1% 0.95 – 2.08<br />

dPT 24 1.04 10.6% 0.83 – 1.28<br />

Innovin (Siemens) 11 1.03 9.7% 0.83 – 1.15<br />

SCT 17 1.11 14.4% 0.80 – 1.48<br />

Other 2 1.26 - 1.24 – 1.28<br />

Remarks:<br />

The high CV of the dAPTT-group and KCT-group in comparison to the other assay groups is<br />

remarkable. The relatively high CV in the aPTT group in comparison to previous surveys is<br />

caused by the differences in sensitivity to CRP between the different reagents. Within a particular<br />

aPTT reagent group the CVs relatively low.<br />

RATIO MIXING n mean CV range<br />

APTT 204 1.21 14.0% 0.82 – 2.02<br />

APTT SP (IL) 22 1.12 7.9% 0.82 – 1.25<br />

Pathromtin SL (Siemens) 16 1.11 6.2% 0.99 – 1.24<br />

Actin FSL (Siemens) 36 1.20 6.9% 0.99 – 1.38<br />

PTT Automate / STA APTT<br />

(Stago)<br />

24 1.34 10.4% 1.15 – 1.75<br />

PTT <strong>LA</strong> (Stago) 69 1.32 13.6% 0.92 – 2.02<br />

DRVVT 100 1.00 8.5% 0.77 – 1.59<br />

<strong>LA</strong>C screen (I.L.) 17 1.04 14.4% 0.92 – 1.59<br />

<strong>LA</strong> screen (Life Diagnostics) 10 0.98 4.7% 0.90 – 1.05<br />

<strong>LA</strong>1 screen (Siemens) 45 0.99 6.3% 0.77 – 1.10<br />

DRVVT Screen (Stago) 12 1.03 6.5% 0.94 – 1.21<br />

KCT 19 0.96 12.5% 0.67 – 1.20<br />

dAPTT 14 1.20 13.3% 0.99 – 1.59<br />

dPT 12 1.03 7.1% 0.96 – 1.23<br />

SCT 3 1.02 - 0.99 – 1.06<br />

Other 1 0.97 - -<br />

Remarks:<br />

The relatively high CV in the aPTT group in comparison to other assay groups is caused by the<br />

differences in sensitivity to CRP between the different reagents. Within a particular aPTT reagent<br />

group the CV is in most cases relatively low.