<strong>Variable</strong> <strong>Anamorphic</strong> <strong>Prism</strong> <strong>Pair</strong> <strong>Manual</strong> 5. Alignment<br />
2. Choose an appropriate angular orientation of prism 1.<br />
For an estimate Figure 2 shows the theoretical dependence of overall anamorphic magnification<br />
and orientation of prism 1 for a wavelength range of 390 - 1500 nm.<br />
NOTE ! The values given in Figure 2 are calculated for ideal prisms and a wavelength range of<br />
390 - 1500 nm. Different wavelengths lead to slightly different dependences and production<br />
tolerances can lead to deviations.<br />
Figure 2 Angle and Magnification values of ideal prism for wavelength range between 390 nm and<br />
1500 nm.<br />
2.1 Make sure the beam enters the entrance aperture centered.<br />
2.2 Loosen Screw 1 (see Figure 3) and carefully loosen Screws 2 and 3 until <strong>Prism</strong> 1 can be<br />
rotated with a small force.<br />
2.3 Choose the desired magnification and determine the orientation of <strong>Prism</strong> 1 e.g. by the help<br />
of Figure 2.<br />
2.3 Fasten Screw 1.<br />
3. Rotate <strong>Prism</strong> 2 until the exiting beam propagates parallel to the incoming beam.<br />
As an estimate Figure 2 shows calculated angular orientations of <strong>Prism</strong> 2 as a function of the orientation<br />
of <strong>Prism</strong> 1 for a wavelength range of 390 - 1500 nm.<br />
3.1 Loosen Screw 4 until <strong>Prism</strong> 2 can be shifted and rotated.<br />
3.2 Shift <strong>Prism</strong> 2 into the middle of the range of possible positions.<br />
3.3 Rotate <strong>Prism</strong> 2 until the exiting beam hits mark B (cf. step 1.3).<br />
Page 6<br />
Status: 25.5.10