Every patient responds to electrical stimulation differentlyand their needs may vary from the conventional settingssuggested here. If the initial results are not positive, speakto your physician about alternative stimulation settingsand/or electrode placements. Once an acceptable locationhas been achieved, mark down the electrode sites and thedevice settings, so the patient can easily continue treatment.4.5 Turn the device onIn order to turn on the device, press [ON/OFF] button,mode LED indicator and timer LED indicator will light up.4.6 Select the therapy mode(anterior neck) or transcerebrally (through the head).4.7 Adjust the pulse width and pulse rateBy turning the [P. W.] knob clockwise to select pulse width,the pulse width is adjustable from 30 us to 260 us.By turning the [P. R.] knob clockwise to select pulse rate,the pulse rate is adjustable from 2 Hz to 150 Hz.4.8 Select therapy timeThere are 4 therapy time (15, 30, 60, C) for you to select.Press the [TIMER] button, the appropriate time LEDindicator will be light. Press [MODE] button cycle to selectthe therapy time.15