MESSAGE PUMP RESPONSE POSSIBLE CAUSE OPERATOR ACTIONCHECK TIMINGActive in OPERATOR Mode only.Message.Pumping switched from AUTOPILOT Mode to OPERATOR ModeReminder to assess timing. Adjustas needed.A 2 second audible alert mayaccompany message if enabled.Assist interval set to short to fullyinflate balloon.Press ASSIST FREQ to 1:2 to evaluatetiming. Readjust timing as necessary.ECG DETECTEDMessage and beep tone every5 seconds.Cardiac activity (ECG) detected whilepumping in INTERNAL trigger.Switch trigger mode according topatient’s rhythm. Readjust timing.FILL FAILUREAudible tone, message, deflateballoon; stop pumpingGAS ALARMS OFF Mode: NoneFailure to fill to 6-10mmHg duringfill state.Press ASSIST key again. Check foradequate supply of helium. If alarmpersists, operate in GAS ALARMSOFF override mode until anotherconsole is available.FILL PRESSUREAudible tone, message, deflateballoon; stop pumping.Changing balloon volume whilepumping.If IAB volume was changed whilepumping, depress ASSIST and beginpumping again.High helium pressure in balloon line.Notify service if alarm conditioncontinues.GAS ALARMS OFF Message only Pump is set to GAS ALARMS OFFMode.Press and hold GAS ALARMS OFFkey for 3 seconds to turn this modeoff (LED off) and reactivate theautomatic surveillance systems.