MODEL G&D METER METER DETAILS<br />
E YEPIECE<br />
The focus of the eyepiece (ocular) is adjustable. To move the<br />
eyepiece up or down, first loosen the knurled ring below the top of<br />
the eyepiece. Turning counterclockwise loosens. While looking<br />
through the eyepiece, move it up or down until a good focus is<br />
found. Then, retighten the knurled ring to hold the eyepiece in<br />
position.<br />
A RRESTMENT K NOB<br />
The arrestment knob is knurled and made of aluminum. It is held<br />
on the shaft by two very small set screws. It is positioned very<br />
close to the arrestment stalk to minimize the chance of dirt entering<br />
the top of the stalk.<br />
If something is dropped on the top of the knob, the knob may be<br />
driven down and drag on the stalk. To repair, use a 3 ⁄32-inch Allen<br />
tool to loosen the knob set screws, raise the knob about a tenth of<br />
a millimeter (a few thousands of an inch) and retighten the set<br />
screws.<br />
If the meter is subjected to salt spray or is stored damp, enough<br />
corrosion may develop between the knob and the stalk to cause<br />
dragging. Repair in the same way as above.<br />
If the knob is very hard to turn and is not dragging on the stalk, the<br />
problem is more serious. The bearings or seals in the stalk may<br />
require servicing at L and R .<br />
INSTRUCTION MANUAL 3-2001 4 - 3