MODEL G&D METER METER DETAILS<br />

E YEPIECE<br />

The focus of the eyepiece (ocular) is adjustable. To move the<br />

eyepiece up or down, first loosen the knurled ring below the top of<br />

the eyepiece. Turning counterclockwise loosens. While looking<br />

through the eyepiece, move it up or down until a good focus is<br />

found. Then, retighten the knurled ring to hold the eyepiece in<br />

position.<br />

A RRESTMENT K NOB<br />

The arrestment knob is knurled and made of aluminum. It is held<br />

on the shaft by two very small set screws. It is positioned very<br />

close to the arrestment stalk to minimize the chance of dirt entering<br />

the top of the stalk.<br />

If something is dropped on the top of the knob, the knob may be<br />

driven down and drag on the stalk. To repair, use a 3 ⁄32-inch Allen<br />

tool to loosen the knob set screws, raise the knob about a tenth of<br />

a millimeter (a few thousands of an inch) and retighten the set<br />

screws.<br />

If the meter is subjected to salt spray or is stored damp, enough<br />

corrosion may develop between the knob and the stalk to cause<br />

dragging. Repair in the same way as above.<br />

If the knob is very hard to turn and is not dragging on the stalk, the<br />

problem is more serious. The bearings or seals in the stalk may<br />

require servicing at L and R .<br />

INSTRUCTION MANUAL 3-2001 4 - 3