ARMONIZACIÓN EN DOSIMETRÍA INTERNA OCUPACIONAL
ARMONIZACIÃN EN DOSIMETRÃA INTERNA OCUPACIONAL
ARMONIZACIÃN EN DOSIMETRÃA INTERNA OCUPACIONAL
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
Norma ISO 28218 “Performance criteria for radiobioassay”<br />
Límite de Detección (LD) y Umbral de Decisión (DT) – CRC - CIEMAT<br />
Counting System Tc(s) Radionuclide LD (Bq) DT (Bq)<br />
NaI(Tl) detector - Total Body 1200 Mn-54 (834 keV) 56,7 28<br />
Co-57 (122 keV) 70 34,7<br />
Co-60 (1173.2 keV) 55,7 27,4<br />
Y-88 (898 keV) 60,6 30<br />
Tc-99m (140.5 keV) 64,5 32<br />
Sn-113 (391.7) 82,9 41<br />
Cs-134 (798,9 keV) 60 29,6<br />
Cs-137 (661.7 keV) 66,9 33<br />
K-40 (1460 keV) 1390 692<br />
FASTSCAN Counter - Total body 300 Mn-54 (834 keV) 94 46,4<br />
Co-57 (122 keV) 201,2 100,1<br />
Co-60 (1173.2 keV) 94,5 46,6<br />
Y-88 (898 keV) 102,3 50,5<br />
Tc-99m (140.5 keV) 176,7 87,9<br />
Sn-113 (391.7) 156,1 77,4<br />
Cs-134 (798,9 keV) 96,1 47,5<br />
Cs-137 (661.7 keV) 109 53,9<br />
K-40 (1460 keV) 2144 1065<br />
NaI(Tl) det +collimator - 131 I in thyroid 1200 I-131 (364,5 keV) 18,4 9,1<br />
FASTSCAN - 131 I in thyroid 300 I-131 (364,5 keV) 51,7 25,6<br />
LE Ge detector - 131 I in thyroid 1200 I-131 (364,5 keV) 6,4 3,45<br />
LE Ge system - 241 Am in lungs (CWT= 3cm) 2700 Am-241 (59.54 keV) 7,7 3,7