<strong>Micro</strong><br />

<strong>Dancer</strong><br />

The High Speed<br />

Evaporation<br />

System<br />

from Hettich AG<br />

L A B O R A P P A R A T E


High efficient<br />

solvent<br />

evaporation<br />

<strong>Micro</strong><br />

<strong>Dancer</strong><br />

The new <strong>Micro</strong><strong>Dancer</strong> infrared vortex vacuum evaporating<br />

system impresses by its small extend and fits<br />

therefore in every laboratory. Due to its easy operability<br />

<strong>Micro</strong><strong>Dancer</strong> is fast and reliably ready for operation –<br />

especially designed for laboratories with high sample<br />

throughput. Here the <strong>Micro</strong><strong>Dancer</strong> shows its productive<br />

efficency – day after day – for an exceptional attractive<br />

value for money!<br />

Operation mode<br />

<strong>Micro</strong><strong>Dancer</strong> represents the new generation of vacuum<br />

evaporator. Vortexing, heating and vacuum are the keywords<br />

to a high efficient evaporation of solvents. Orbital<br />

shaking of the sample rack enlarges the surface of the<br />

sample. The patented contactless drive of the shaker<br />

plate eases loading and removing of the sample racks.<br />

It is not necessary to balance the samples. A major advantage<br />

for samples, obtained by HPLC gradient separation.<br />

A temperature sensor, placed in the vacuum chamber measures the<br />

drop in temperature of the sample and controls the infrared rod placed in the<br />

lid of the vacuum chamber.<br />

Vortexing<br />

Through orbital shaking (vortexing) of the samples the<br />

evaporating surface is substantially increased and they<br />

are kept in constant movement. The advantages over<br />

centrifugation are obvious. There is no imbalance problem<br />

if samples evaporate at different speeds. The quality<br />

and long-life cycle of the <strong>Micro</strong><strong>Dancer</strong> is not influenced. You can use completely<br />

different vials at the same time with the universal rack, as shown in<br />

the picture attached.<br />

Vacuum chamber<br />

insight with temperature<br />

sensor.<br />

Heating<br />

The temperature sensor in the vacuum chamber detects<br />

the temperature gradient during the evaporation<br />

process, which is caused by the latent heat under vacuum.<br />

It controls the infrared rod, which generates a<br />

constant heat distribution due to the parabolic reflectors.<br />

This avoids icing of the samples. Additionally the<br />

compartment heater can be activated to prevent condensation<br />

of the solvent at the wall of the vacuum<br />

chamber.<br />

Vacuum<br />

The diaphragm pump Vacuubrand PC3003 Vario generates optimum vacuum<br />

to operate the <strong>Micro</strong><strong>Dancer</strong>. Intelligent control based on a sensitive vapour<br />

pressure sensor avoids boiling delays by adapting the pump speed to the<br />

evaporation process.


The shaker plate is equipped with<br />

three magnetic rotary feet and can<br />

be easily placed in the circular holes<br />

at the bottom of the vacuum chamber<br />

– without any further adjustment.<br />

Individually racks can be mounted at<br />

Sample racks<br />

The patented magnetic linkage couples the racks to the<br />

bottom of the vacuum chamber. Vortexing of samples<br />

operates without any contact to the actuator.<br />

the shaker plate. There are two alternatives available:<br />

Rack with 50 ml Falcon<br />

tubes.<br />

Magnetic linkage at<br />

the bottom of the vacuum<br />

chamber.<br />

Rack with scintillation<br />

vials.<br />

Version II:<br />

To use many different vials we suggest<br />

to use sample tube blocks,<br />

which can be easily exchanged. The<br />

shaker plate is equipped with a special<br />

holder to carry two different<br />

blocks simultaneously.<br />

Shaker plate –<br />

top side.<br />

Shaker plate –<br />

bottom side.<br />

Version I:<br />

To evaporate multiple samples with<br />

the same sample tube we recommend<br />

to use specially designed<br />

racks.<br />

Shaker plate with<br />

special hold for sample<br />

tube blocks.<br />

Applications with the <strong>Micro</strong><strong>Dancer</strong> have no rival:<br />

For recovery of the evaporated solvents we recommend<br />

the coldtrap PoleStar. It is equipped with a fully<br />

inert 3.3 litre container.<br />

The superfast defrost mode saves time during the daily<br />

routine work. PoleStar is the ideal completion of your<br />

evaporation system.<br />

Variety of different<br />

sample tube blocks.<br />

Solvent Sample Tube type Number Temperature Evaporation<br />

volume (ml) of tubes (°C) time (min)<br />

Hexane 20 50 ml cone tubes 36 40 20<br />

Water 0,5 1.5 ml microlitre tube 192 40 56<br />

Ether derivatives 3.5–4 10 ml tube 96 30 40<br />

Methanol<br />

Water/<br />

8 20 ml scintillation vial 45 40 40<br />

Ethanol 50/50 4 16x100 mm tube 96 40 120<br />

New Impulse<br />

for the<br />

Laboratory<br />

<strong>Micro</strong><br />

<strong>Dancer</strong>


Device<br />

specifications<br />

<strong>Micro</strong><br />

<strong>Dancer</strong><br />

Dimensions W x H x D (lid closed) 448 x 400 x 530 mm<br />

Dimensions W x H x D (lid opened) 448 x 700 x 620 mm<br />

Weight approx. 50 kg net<br />

(without base plate)<br />

Power source 230 VAC ±5%<br />

Power consumption total 940 VAC<br />

Power consumption of single components:<br />

– Heater (IR rod) 500 VA<br />

– Compartment heater (vacuum chamber) 400 VA<br />

– Shaker motor 24 VA<br />

– Fan 5 VA<br />

– Release valve 6 VA<br />

– Temperature controller 5 VA<br />

– LED chamber light 3 VA<br />

Fuses 2x T5A L<br />

glass tube fuse 5x20 mm<br />

Vortex rate 0–1000 rpm ±5%<br />

Adjustable temperature range (controller) 5°C–60°C<br />

Temperature compartment heater<br />

(factory-provided)