General<br />
Supply<br />
for 0 - 1V 10 - 35V DC or 9 - 29V AC<br />
for 0 - 5V 12 - 35V DC or 15 - 29V AC<br />
for 0 - 10V 15 - 35V DC or 15 - 29V AC<br />
for 4 - 20mA 10V +R L x 0,02 < U V < 35V DC; R L < 500 Ohm<br />
Current consumption for DC supply: typ. 5mA for AC supply: typ. 15mA eff<br />
Electrical connection screw terminals max. 1.5 mm 2 (AWG 16)<br />
Cable gland M16x1.5 or connection plug (only snap-in models N + M)<br />
cable Ø 4.5 - 10 mm (0.18 - 0.39”)<br />
Sensor protection membrane filter, sintered stainless steel filter, metal grid filter, PTFE filter<br />
Electromagnetic compatibility EN61326-1 EN61326-2-3 ICES-003 ClassB<br />
Industrial Environment FCC Part15 ClassB<br />
Temperature ranges working temperature probe: -40...60°C (-40...140°F)<br />
working temperature electronics: -40...60°C (-40...140°F)<br />
storage temperature: -25...60°C (-13...140°F)<br />
1) Please refer to working range of HC1000!<br />
Dimensions (mm)<br />
Type A<br />
Connection Diagram<br />
<strong>EE21</strong>-FT1/2/3xxx / <strong>EE21</strong>-F1/2/3xxx<br />
power supply<br />
for 0 - 1V: 10 - 35V DC<br />
9 - 29V AC<br />
for 0 - 5V: 12 - 35V DC<br />
15 - 29V AC<br />
for 0 -10V: 15 - 35V DC<br />
15 - 29V AC<br />
<strong>EE21</strong>-FT6xxx / <strong>EE21</strong>-F6xxx<br />
power supply<br />
20...35V DC; R L < 500 Ohm<br />
11...35V DC; R L < 50 Ohm<br />
Ordering Guide<br />
MODEL OUTPUT T-SENSOR HOUSING TYPE PROBE LENGTH FILTER<br />
(only passive) (see dimensions “A”)<br />
humidity + temperature (FT) 0 - 1 V (1) Pt 100 DIN A (A) wall mounting (A) 50 mm (1.9”) (2) membrane filter (1)<br />
humidity (F) 0 - 5 V (2) Pt 1000 DIN A (C) duct mounting (B) 200 mm (7.9”) (5) sintered stainless steel filter (3)<br />
humidity+temp. passive (FP) 0 - 10 V (3) snap in - wall mounting 1)<br />
(M) metal grid filter (6)<br />
4 - 20 mA (6) snap in - duct mounting 1)<br />
(N)<br />
<strong>EE21</strong>-<br />
Type B<br />
COATING T-UNIT SCALING OF T-OUTPUT<br />
no (no code) °C (no code) -40...60 (T02)<br />
yes (HC01) °F (E01) -20...80 (T24)<br />
-30...70 (T08)<br />
other (Txx)<br />
1) Combination snap - in mounting and model FP is not possible<br />
housing: PC<br />
protection class: IP65, Nema 4<br />
Accessories<br />
- radiation shield (HA010501) - filter caps (HA0101xx)<br />
37<br />
<strong>EE21</strong>-FP3xxx<br />
power supply<br />
for 0 - 1V: 10 - 35V DC<br />
9 - 29V AC<br />
for 0 - 5V: 12 - 35V DC<br />
15 - 29V AC<br />
for 0 -10V: 15 - 35V DC<br />
15 - 29V AC<br />
<strong>EE21</strong>-FP6xxx<br />
power supply<br />
20...35V DC; R L < 500 Ohm<br />
11...35V DC; R L < 50 Ohm<br />
1 mm = 0.03937” / 1”= 25.4 mm<br />
snap in - mounting kit<br />
Type N<br />
passive<br />
passive<br />
passive<br />
passive<br />
Type M<br />
Order Example<br />
<strong>EE21</strong>-FT3A26/T24<br />
model: RH/T transmitter<br />
output: 0 - 10V<br />
housing type: wall mounting<br />
probe length: 50 mm (7.9”)<br />
filter: metal grid filter<br />
sensor coating: no<br />
calibration: standard<br />
T-unit: °C<br />
Scaling of T-output: -20...80°C<br />
<strong>EE21</strong>