General<br />

Supply<br />

for 0 - 1V 10 - 35V DC or 9 - 29V AC<br />

for 0 - 5V 12 - 35V DC or 15 - 29V AC<br />

for 0 - 10V 15 - 35V DC or 15 - 29V AC<br />

for 4 - 20mA 10V +R L x 0,02 < U V < 35V DC; R L < 500 Ohm<br />

Current consumption for DC supply: typ. 5mA for AC supply: typ. 15mA eff<br />

Electrical connection screw terminals max. 1.5 mm 2 (AWG 16)<br />

Cable gland M16x1.5 or connection plug (only snap-in models N + M)<br />

cable Ø 4.5 - 10 mm (0.18 - 0.39”)<br />

Sensor protection membrane filter, sintered stainless steel filter, metal grid filter, PTFE filter<br />

Electromagnetic compatibility EN61326-1 EN61326-2-3 ICES-003 ClassB<br />

Industrial Environment FCC Part15 ClassB<br />

Temperature ranges working temperature probe: -40...60°C (-40...140°F)<br />

working temperature electronics: -40...60°C (-40...140°F)<br />

storage temperature: -25...60°C (-13...140°F)<br />

1) Please refer to working range of HC1000!<br />

Dimensions (mm)<br />

Type A<br />

Connection Diagram<br />

<strong>EE21</strong>-FT1/2/3xxx / <strong>EE21</strong>-F1/2/3xxx<br />

power supply<br />

for 0 - 1V: 10 - 35V DC<br />

9 - 29V AC<br />

for 0 - 5V: 12 - 35V DC<br />

15 - 29V AC<br />

for 0 -10V: 15 - 35V DC<br />

15 - 29V AC<br />

<strong>EE21</strong>-FT6xxx / <strong>EE21</strong>-F6xxx<br />

power supply<br />

20...35V DC; R L < 500 Ohm<br />

11...35V DC; R L < 50 Ohm<br />

Ordering Guide<br />

MODEL OUTPUT T-SENSOR HOUSING TYPE PROBE LENGTH FILTER<br />

(only passive) (see dimensions “A”)<br />

humidity + temperature (FT) 0 - 1 V (1) Pt 100 DIN A (A) wall mounting (A) 50 mm (1.9”) (2) membrane filter (1)<br />

humidity (F) 0 - 5 V (2) Pt 1000 DIN A (C) duct mounting (B) 200 mm (7.9”) (5) sintered stainless steel filter (3)<br />

humidity+temp. passive (FP) 0 - 10 V (3) snap in - wall mounting 1)<br />

(M) metal grid filter (6)<br />

4 - 20 mA (6) snap in - duct mounting 1)<br />

(N)<br />

<strong>EE21</strong>-<br />

Type B<br />

COATING T-UNIT SCALING OF T-OUTPUT<br />

no (no code) °C (no code) -40...60 (T02)<br />

yes (HC01) °F (E01) -20...80 (T24)<br />

-30...70 (T08)<br />

other (Txx)<br />

1) Combination snap - in mounting and model FP is not possible<br />

housing: PC<br />

protection class: IP65, Nema 4<br />

Accessories<br />

- radiation shield (HA010501) - filter caps (HA0101xx)<br />

37<br />

<strong>EE21</strong>-FP3xxx<br />

power supply<br />

for 0 - 1V: 10 - 35V DC<br />

9 - 29V AC<br />

for 0 - 5V: 12 - 35V DC<br />

15 - 29V AC<br />

for 0 -10V: 15 - 35V DC<br />

15 - 29V AC<br />

<strong>EE21</strong>-FP6xxx<br />

power supply<br />

20...35V DC; R L < 500 Ohm<br />

11...35V DC; R L < 50 Ohm<br />

1 mm = 0.03937” / 1”= 25.4 mm<br />

snap in - mounting kit<br />

Type N<br />

passive<br />

passive<br />

passive<br />

passive<br />

Type M<br />

Order Example<br />

<strong>EE21</strong>-FT3A26/T24<br />

model: RH/T transmitter<br />

output: 0 - 10V<br />

housing type: wall mounting<br />

probe length: 50 mm (7.9”)<br />

filter: metal grid filter<br />

sensor coating: no<br />

calibration: standard<br />

T-unit: °C<br />

Scaling of T-output: -20...80°C<br />