DICOM Conformance Statement - Medical Systems

fujimed.com

DICOM Conformance Statement - Medical Systems

Synapse DICOM Conformance Statement

Document Version 08

Saved Date: 0/0/0000 0:00:00 AM

SOP Class Name Class UID Service Group

1.2.840.10008.5.1.4.1.1.7.3 (SC-MF-Word)

1.2.840.10008.5.1.4.1.1.7.2 (SC-MF-Color)

1.2.840.10008.5.1.4.1.1.88.11 (Basic Txt SR)

1.2.840.10008.5.1.4.1.1.88.50 (MG CAD SR)

6.3 Supported DICOM Attributes during C-FIND and C-MOVE

C-FIND

C-MOVE

Query Level Attribute Name DICOM Tag

Match/

Return

All

SOP Class UID

Query/Retrieve Level

0002,0002

0008,0052

-

-

Patient Name 0010,0010 M, R

Patient ID 0010,0020 M, R

Study Date 0008,0020 M, R

Study

Study Time

Accession Number

0008,0030

0008,0050

R

M, R

Study ID 0020,0010 R

Study Instance UID 0020,000D R

Modality in Study 0008,0061 R

Modality 0008,0060 R

Series Series Number 0020,0011 R

Series Instance UID 0020,000E M, R

Instance Number 0020,0013 R

SOP Instance UID 0008,0018 M, R

Instance

Rows

Columns

0028,0010

0028,0011

R

R

Number of Frames 0028,0008 R

SOP Class UID 0008,0016 R

SOP Class UID 0002,0002 -

Query/Retrieve Level 0008,0052 -

Destination AE Title 0000,0600 -

Instance Patient ID 0010,0020 M

Study Instance UID 0020,000D M

Series Instance UID 0020,000E M

SOP Instance UID 0008,0018 M

FUJIFILM Medical Systems USA Page 18 of 22

Proprietary Information ENG-0012397-A

Effective Date: 5/14/2008