3.4. Flexel ® <strong>3D</strong> <strong>Bag</strong> Emptying<br />

a) Warning: Before starting bag emptying,<br />

remove the stainless steel hanger from its<br />

limit stopper (See Fig 18).<br />

b) Take the bottom port tube and position it<br />

vertically. Open the connector protective<br />

outer-wrap using the reversible cable tie<br />

on the dispensing line and carry out the<br />

connection according to the standard<br />

procedures that apply on the site<br />

(See Fig. C).<br />

c) Unclamp the drain line and start the<br />

draining operation (See Fig. A) - Gravity<br />

can be used, but for faster draining <strong>Sartorius</strong><br />

Stedim Biotech recommends the<br />

use of a peristaltic pump. (See Fig. 19)<br />

d) The Flexel ® <strong>3D</strong> <strong>Bag</strong> closes down on itself<br />

and collapses automatically. To obtain<br />

100% fluid recovery, lift the collapsed<br />

bag.<br />

e) When dispensing is complete, clamp the<br />

line (See Fig. B) and disconnect the main<br />

drain line from the drain tube of the<br />

Flexel ® <strong>3D</strong> <strong>Bag</strong>.<br />

f) Remove each stainless steel hook of the<br />

cable from holes located on the Flexel ®<br />

<strong>3D</strong> <strong>Bag</strong> top flaps.<br />

g) Open the three bottom gates to free the<br />

bottom ports by removing the retaining<br />

clip from the retaining clip fastening<br />

point; then grasp the retaining clip fastening<br />

point and swing the bottom gate<br />

open.<br />

h) Remove the collapsed bag from the<br />

Palletank ® (See Fig. 20).<br />

4. Trouble Shooting Guide 5. Precautions of Use<br />

The bag will not fill<br />

> Check that the clamp is open and that<br />

the tube is not pinched.<br />

When filling through the bottom, the<br />

bag comes out of the bottom gate.<br />

> The edge of the jaw is not closed inside<br />

the groove of the bottom port. Position<br />

it correctly in the groove.<br />

The bag does not open out correctly<br />

during filling:<br />

> The port system is incorrectly aligned.<br />

The bag is incorrectly positioned in<br />

the Palletank ®. Lift the bag out of the<br />

Palletank ®, turn it round and go through<br />

Flexel ® <strong>3D</strong> <strong>Bag</strong> installation procedure<br />

once again.<br />

> The bottom of the bag is incorrectly<br />

positioned.<br />

Place the corners of the bag in the corners<br />

of the Palletank ® (make sure that<br />

only 1-2% of nominal volume is filled<br />

before adjusting, otherwise bag failure<br />

may occur)<br />

Many folds appear in the bag during<br />

filling<br />

> The bag is too big, use the bag suited to<br />

the size of the Palletank ®<br />

The bag does not come up against the<br />

walls<br />

> The bag is too small, use the bag suited<br />

to the size of the Palletank ®<br />

The bag will not drain<br />

> Check that the clamp is open and that the<br />

tube is not pinched.<br />

a) The <strong>Sartorius</strong> Stedim Biotech Flexel ® <strong>3D</strong><br />

<strong>Bag</strong> must imperatively be used with the<br />

<strong>Sartorius</strong> Stedim Biotech Palletank ®.<br />

b) The Palletank ® must be clean and dry<br />

before inserting the Flexel ® <strong>3D</strong> <strong>Bag</strong>.<br />

Traces of cleaning agents left on the<br />

Palletank ® can potentially damage the<br />

Flexel ® <strong>3D</strong> <strong>Bag</strong>.<br />

c) It is under the user’s responsibility to<br />

verify the compatibility of the Flexel ® <strong>3D</strong><br />

<strong>Bag</strong> materials with the solutions used in<br />

your process. <strong>Sartorius</strong> Stedim Biotech<br />

can provide guidelines for chemical compatibility,<br />

bio-compatibility and extractables.<br />

d) Avoid leaving sharp objects (scissors, cutters,<br />

scalpels, etc.) in the vicinity of the<br />

Flexel ® <strong>3D</strong> <strong>Bag</strong>.<br />

e) The Flexel ® <strong>3D</strong> <strong>Bag</strong> is not to be handled<br />

by its connections (i.e. tubing).<br />

f) Avoid folding the Flexel ® <strong>3D</strong> <strong>Bag</strong> other<br />

than the original folds. Do not fold the<br />

bag in the direction of the “gussets”.<br />

(Gussets Heat-sealed seams)<br />

g) When positioning the Flexel ® <strong>3D</strong> <strong>Bag</strong> in<br />

the Palletank ® and when adjusting it after<br />

the initial fill, handle the bag gently and<br />

do not force fit or pull hard.<br />

h) The Flexel ® <strong>3D</strong> <strong>Bag</strong> is for single-use and<br />

should not be re-used.<br />