‣ Clean the product immediately after use.<br />
‣ Further detailed advice on hygienically safe and<br />
material-/value-preserving reprocessing can be<br />
found at www.a-k-i.org, link to Publications, Red<br />
Brochure – Proper maintenance of instruments.<br />
Preparations prior to cleaning<br />
‣ Have the product dry in a disposal container and<br />
ready for immediate cleaning and disinfecting<br />
within 30 min after use.<br />
Cleaning/Disinfecting<br />
WARNING<br />
Cut injuries due to touching of the<br />
diamond blade!<br />
‣ Do not touch the diamond<br />
blade.<br />
‣ Activate the sleeve when the<br />
instrument is not used.<br />
‣ Apply proper caution when<br />
handling or cleaning the<br />
instrument.<br />
CAUTION<br />
Damage to the product due to<br />
inappropriate cleaning/disinfecting<br />
agents and/or excessive<br />
temperatures!<br />
‣ Use cleaning and disinfecting<br />
agents approved for titanium<br />
and high-grade steel according<br />
to manufacturer’s instructions.<br />
‣ Observe specifications regarding<br />
concentration, temperature and<br />
exposure time.<br />
‣ Do not exceed the maximum<br />
allowable cleaning temperature<br />
of 55 °C.<br />
‣ If the microsurgical products can be securely fixed<br />
in machines or in storage devices in such a way<br />
that they will be thoroughly cleaned, clean and<br />
disinfect them mechanically.<br />
CAUTION<br />
Damage to the diamond blade!<br />
‣ Avoid contact of the delicate<br />
blades at tips with hard surfaces<br />
or other instruments.<br />
‣ Do not apply ultrasound<br />
treatment.<br />
‣ Do not carry out immersion<br />
disinfection.<br />
‣ Do not clean the diamond blade<br />
with cleaning brushes.<br />
‣ Process and store the product in<br />
an appropriate storage unit.<br />
Ensure that the product is<br />
securely fixed in the storage<br />
unit.<br />