Technical Data und Instructions Country-specific design<br />
<strong>IVV</strong> <strong>1054</strong> und <strong>IME</strong> <strong>1500</strong> (on request)<br />
electric-sockets<br />
Classification conventional ballast<br />
IIb connection diameter<br />
Nominal voltage/Nominal frequency<br />
outlet points for med. gases<br />
Electrical Data sizing of pipe diameters<br />
Nominal voltage: 230 V (400V) AC operating pressure<br />
Nominal frequency: 50 Hz signal current technology/nurse call<br />
Operating mode: continuous operation<br />
Protection class: Protection class 1<br />
Type of protection: closed design Installation<br />
Heavy current connection:<br />
For the connecting of the cables (on site) there are<br />
screw less connectors for cables of up to 2.5 mm².<br />
The following connection possibilities for potential<br />
compensation with a protective connector block are at<br />
your disposal:<br />
The medical supply system, types <strong>IVV</strong> <strong>1054</strong><br />
<strong>and</strong> <strong>IME</strong> <strong>1500</strong>, must be mounted on solid<br />
walls <strong>and</strong> on light-weight partitions using<br />
appropriate fastening materials. With light-<br />
weight partitions, a suitable substructure<br />
must be provided.<br />
1 x max. 16 mm²<br />
n x max. 4 mm²<br />
Night light TC 7 Watt<br />
Nominal voltage: 230 V AC<br />
Nominal frequency: 50 HZ<br />
Connection type: Connectors with spring hold<br />
Connection diameter: single wire/fine str<strong>and</strong> 1,5 mm²<br />
Cleaning<br />
The <strong>IVV</strong> <strong>1054</strong> <strong>and</strong> <strong>IME</strong> <strong>1500</strong>-systems can be<br />
cleaned with all st<strong>and</strong>ard cleaning agents <strong>and</strong><br />
can be disinfected with all alcohol-free<br />
disinfection liquids.<br />
The surface is coated with a high quality<br />
electro static powder coating. Smooth<br />
surfaces are made of electrically-oxidized<br />
Aluminium <strong>and</strong> chrome-nickel steel. The<br />
plastic parts are cleaning <strong>and</strong> disinfection<br />
Operating pressures for med. gases<br />
liquid proof.<br />
(outlet points in accordance with<br />
DIN EN 737-1, -3, -4)<br />
Oxygen:<br />
Compressed air:<br />
Nitrous oxide:<br />
Vacuum:<br />
AGSS:<br />
5 bar<br />
5 bar<br />
5 bar<br />
- 0,8 bar<br />
5 bar<br />
Maintenance <strong>and</strong> Repair<br />
System maintenance should be carried out<br />
after 5 years <strong>and</strong> every 2 years thereafter. If<br />
desired, a customer service agreement can<br />
be drawn up at a later date.