13.04.2013 Views

Infant Flow® LP system headgear - CareFusion

Infant Flow® LP system headgear - CareFusion

Infant Flow® LP system headgear - CareFusion

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>Infant</strong> Flow ®<br />

<strong>LP</strong> <strong>system</strong> <strong>headgear</strong>


Prepare <strong>Infant</strong> Flow ®<br />

<strong>LP</strong> SiPAP driver ...............1<br />

<strong>Infant</strong> Flow <strong>LP</strong> components ............................2<br />

Generator assembly components ...................3<br />

Low pressure generator .................................4<br />

Nasal prongs ..................................................5<br />

Nasal masks ...................................................6<br />

Nasal prongs and mask sizing ........................7<br />

Headgear sizing .............................................8<br />

Generator assembly application .....................9<br />

Headgear application ............................9-11<br />

Check for proper <strong>headgear</strong> placement .....12<br />

Incorrect <strong>headgear</strong> placement .............13-14<br />

Attach interface to generator ...................15<br />

Remove the support cradle ......................16<br />

Apply support cradle to <strong>headgear</strong> ............17<br />

Align generator assembly .........................18<br />

Table of contents<br />

i<br />

Secure the locking strap ...........................19<br />

Attach side straps to generator ................20<br />

Final check of generator assembly ................21<br />

Check for correct <strong>headgear</strong> placement .....21<br />

Check placement of exhalation tube ........22<br />

Check for proper nasal prong position .....23<br />

Incorrect nasal prong position ..................24<br />

Verify proper mask placement ..................25<br />

Incorrect mask placement ........................26<br />

Incorrect mask size—too small .................27<br />

Incorrect mask size—too large .................28<br />

Warnings and cautions ................................29<br />

Application tips ............................................30<br />

Inspect the <strong>system</strong> .......................................31<br />

Notes ......................................................32-33


Prepare <strong>Infant</strong> Flow <strong>LP</strong> SiPAP driver<br />

1. Attach the infant circuit to humidifier<br />

and nCPAP driver.<br />

2. Attach generator to the circuit.<br />

3. Perform oxygen sensor calibration<br />

and user verification.<br />

4. Enter desired settings.<br />

1<br />

NCPAP/<br />

Pres Low<br />

L/min<br />

14<br />

50<br />

60<br />

70<br />

40<br />

80<br />

12<br />

30<br />

90<br />

10<br />

21<br />

8<br />

100<br />

6<br />

4<br />

2<br />

<strong>Infant</strong> Flow<br />

SiPAP<br />

XDCR<br />

P PROX<br />

Pres High<br />

L/min<br />

5<br />

4<br />

3<br />

2<br />

1<br />

3 4<br />

1<br />

2


<strong>Infant</strong> Flow <strong>LP</strong> components<br />

A. Prongs and masks<br />

B. Generator<br />

C. Bonnet sizing guide<br />

D. Mask/prong sizing guide<br />

E. Headgear<br />

F. Bonnet<br />

D<br />

A<br />

B<br />

2<br />

E<br />

F<br />

C


Generator assembly components<br />

A. Exhalation tube<br />

B. Securing strap<br />

C. Generator receiver<br />

D. Fixation tab<br />

E. Mask/prong<br />

attachment base<br />

F. Support cradle<br />

G. Pressure and drive lines<br />

C<br />

D<br />

E<br />

3<br />

B<br />

F<br />

A<br />

G


Low pressure generator<br />

• Lower drive pressure<br />

– Prevents pressure backup into the<br />

water auto-feed <strong>system</strong><br />

• Pressure relief valve<br />

4<br />

Pressure relief valve


Nasal prongs<br />

Anatomically designed<br />

A. Flared tip<br />

B. Flexible bellows<br />

C. Base<br />

D. Size indicator<br />

E. Key<br />

5<br />

A<br />

B<br />

C<br />

D<br />

E<br />

A<br />

B<br />

C<br />

D


Nasal mask<br />

Anatomically designed<br />

A. Nasal bridge cushion<br />

B. Contour shape<br />

C. Key<br />

D. Flexible bellow<br />

E. Assessment window<br />

F. Size indicator<br />

6<br />

A<br />

B<br />

C<br />

D<br />

E<br />

E<br />

D<br />

F


Nasal prongs and mask sizing<br />

• Use sizing guide to determine appropriate<br />

mask or prong size<br />

• Nasal prongs: Choose dots that fill the<br />

nare space<br />

• If between sizes, choose the larger size<br />

• Nasal masks: Choose the triangle that fits<br />

over the nose<br />

7


Headgear sizing<br />

Measure the circumference of the infant’s head; from crown at back of head to brow line<br />

(in centimeters).<br />

8


Generator assembly application<br />

Headgear application<br />

1. Align <strong>headgear</strong> with the midline of face.<br />

2. Align bottom of the <strong>headgear</strong> with the<br />

nape of neck.<br />

9<br />

1.<br />

2


Headgear application (continued)<br />

3. Place #1 strap just above brow line.<br />

Note: Straps are labeled 1, 2, 3.<br />

4. Place #2 strap midline over strap #1.<br />

10<br />

3.<br />

4.


Headgear application (continued)<br />

5. Place #3 strap on top of #1 and<br />

#2 straps.<br />

Note: Strap should be snug—do<br />

not overtighten.<br />

6. Fold #2 strap up and over #3<br />

strap; secure to top strap.<br />

11<br />

5.<br />

6.


Check for proper <strong>headgear</strong> placement<br />

• Back of <strong>headgear</strong> rests on nape of neck<br />

• Forehead straps placed on brow line<br />

12


Incorrect <strong>headgear</strong> placement<br />

• Horizontal strap rests close to eye<br />

• Forehead strap is too far above the<br />

brow line<br />

13


Incorrect <strong>headgear</strong> placement (continued)<br />

• Above the nape of the neck, creates a<br />

fold/bulging of <strong>headgear</strong><br />

• Top strap too tight, forehead strap #3<br />

pulled back<br />

14


Attach interface to generator<br />

Align the notch on the prongs/mask to straddle the pressure lines.<br />

1.<br />

3. 4 .<br />

15<br />

2.


Remove the support cradle<br />

1. Expand the collapsible corrugated tubing.<br />

2. Remove the support cradle from the<br />

generator assembly.<br />

16<br />

1.<br />

2.


Apply support cradle to <strong>headgear</strong><br />

1. Apply the support cradle properly<br />

centered on the straps.<br />

2. Secure center tab.<br />

17<br />

1.<br />

2.


Align generator assembly<br />

1. Center the generator and interface over<br />

patient nose and support cradle.<br />

2. Use a gentle rocking motion to insert the<br />

nasal prongs.<br />

3. Place the drive line and pressure line into<br />

the support cradle with the exhalation tube<br />

resting on top.<br />

18<br />

1. 2. 3.


Secure the locking strap<br />

1. Wrap the locking strap over the<br />

exhalation tube and attach to the side<br />

of the support cradle.<br />

2. Position the small slit over the ridge of<br />

the exhalation tube.<br />

19<br />

1.<br />

2.


Attach side straps to generator<br />

1. Wrap the side straps around the<br />

generator tabs.<br />

2. Use the Velcro tip to secure the<br />

strap—loose is better.<br />

3. Use the side straps to adjust tension<br />

on interface—do not over-tighten.<br />

20<br />

1. 2.<br />

3.


Final check of generator assembly<br />

Check for correct <strong>headgear</strong> placement<br />

A. Generator aligned and<br />

centered with nose<br />

B. Side straps properly placed<br />

C. Support cradle tabs secured<br />

to <strong>headgear</strong><br />

Note: Straps should be snug,<br />

but not tight. B<br />

A<br />

21<br />

C


Check placement of exhalation tube<br />

Incorrect Correct<br />

22


Check for proper nasal prong position<br />

• Bellows are visible<br />

• Septum is visible<br />

• Generator assembly has<br />

no contact with skin<br />

23


Incorrect nasal prong position<br />

Too tight, bellows compressed<br />

Prong not in the nare<br />

24


Verify proper mask placement<br />

• Mask covers nasal profile<br />

• Mask rests below the eyes<br />

• Bellows not compressed<br />

• Side strap low on cheek<br />

Note: Do not over-tighten.<br />

25


Incorrect mask placement<br />

• Too much tension applied to one side<br />

• Mask is collapsed<br />

• Nares are not visible<br />

26


Incorrect mask size—too small<br />

• Mask blocks the nares<br />

• Mask unable to cover nose<br />

27


Incorrect mask size—too large<br />

• High on the bridge of the nose<br />

• Up against the eyes<br />

• Rest on the upper lip<br />

28


Warnings and cautions<br />

WARNING:<br />

• Use this product only as directed in the product literature to reduce the risk of nasal irritation,<br />

septal distortion, skin irritation and pressure necrosis<br />

• To be used by a trained practitioner, under the direct supervision of a qualified physician<br />

• Only use the <strong>Infant</strong> Flow <strong>LP</strong> generator with variable flow nCPAP drivers<br />

• Do not over-tighten the fixation straps<br />

29


Application tips<br />

• Select the appropriate size nasal mask to minimize leaks and dead space<br />

• Select the appropriate size nasal prongs; if between sizes, select the larger size<br />

• Application of an incorrectly sized prong, mask, bonnet or <strong>headgear</strong> will affect stability of<br />

the generator<br />

• Adjust the straps to stabilize the generator and maintain a seal at the nose using the least<br />

tension possible<br />

• Consider alternating the use of prong and mask interfaces at set intervals to change pressure<br />

points on the infant’s face<br />

• Continuously monitor patient’s respiratory status<br />

• Cover both ears evenly; ensure the ears are not folded<br />

• Recommend humidification be used with nCPAP <strong>system</strong>s<br />

30


Inspect the <strong>system</strong> at least every three to four hours to:<br />

• Ensure the patient’s eyes are clearly visible and the nares are not blocked, when using masks<br />

• Inspect the fixation device and straps for proper tension and adjust as needed to maintain<br />

a proper fit<br />

• Monitor the patient for gastric insufflation and abdominal distension<br />

• Monitor for excessive condensation in circuit and generator<br />

• Alternate the use of prong and mask interfaces to change pressure points on the<br />

infant’s face<br />

31


Notes<br />

32


Notes<br />

33


WARNING—U.S. Federal Law restricts this device to sale by or on the order of a physician.<br />

<strong>CareFusion</strong><br />

22745 Savi Ranch Parkway<br />

Yorba Linda, CA 92887<br />

800.231.2466 toll-free<br />

714.283.2228 tel<br />

714.283.8493 fax<br />

<strong>CareFusion</strong> Germany 234 GmbH<br />

Leibnizstrasse 7<br />

97204 Hoechberg<br />

Germany<br />

+49 931 4972-0 tel<br />

+49 931 4972-423 fax<br />

carefusion.com<br />

© 2011 <strong>CareFusion</strong> Corporation or one of its subsidiaries. All rights reserved.<br />

<strong>Infant</strong> Flow is a trademark or registered trademark of <strong>CareFusion</strong> Corporation or<br />

one of its subsidiaries. RC2661-01 (1211/5000) L3414 Rev. B

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!