18<br />
Metha ® . Implant and R<strong>as</strong>p Position.<br />
Valgus / Varus variability<br />
+15°<br />
+10°<br />
+5°<br />
The Metha Short Hip Stem can be implanted<br />
at various relative valgus or varus positions<br />
<strong>to</strong> fit the respective bone shape and<br />
implant size. The neutral position is defined<br />
<strong>as</strong> parallel <strong>to</strong> a 50° femoral osteo<strong>to</strong>my.<br />
Other implant positions are up <strong>to</strong> 15°<br />
relative valgus or 5° relative varus.<br />
Ple<strong>as</strong>e be aware that the varus / valgus<br />
variation of the stem influences respectively<br />
the offset situation of the femur.<br />
When preparing the medullary cavity, a<br />
position change of the r<strong>as</strong>p can be detected<br />
by intraoperative comparison with the<br />
osteo<strong>to</strong>my plane.<br />