Modes<br />

Specification Sheet<br />

Available modes IPPV Preset airway temp 40°C, chamber temp 37°C<br />

NPPV For NIV use, airway temp 34°C, chamber temp 31°C<br />

Manual User adjustable temperature settings:<br />

Set Temperature Range<br />

Patient Side<br />

Chamber Side<br />

Patient side: 30°C to 40°C (increments 1°C)<br />

Chamber side: 26°C to 43°C (increments 1°C)<br />

30°C to 40°C (increments 1°C)<br />

26°C to 43°C (increments 1°C)<br />

Temperature setting of patient side vs. chamber side temperature is in relationship within the range of -4°C to +3°C.<br />

Temperature Display<br />

Three digit 14mm 7 segment LED (Light Emitting Diode) display.<br />

Patient Side Temperature Display Range 10°C to 45°C increments 0.1°C<br />

Chamber Side Temperature Display Range 10°C to 55°C increments 0.1°C<br />

Alarms<br />

<strong>Intersurgical</strong><br />

<strong>7000</strong> <strong>Series</strong> <strong>Respiratory</strong> <strong>Humidifier</strong><br />

(PMH<strong>7000</strong>)<br />

Patient Temperature High. Exceeds set temperature by 2°C or >43°C<br />

Low; 4°C below set temperature<br />

Chamber Temperature High. Measured temperature exceeds set temperature by 4°C<br />

for more than 20 mins, or the measured temperature exceeds set<br />

temperature by 10°C<br />

Temperature Probes Faulty, Disconnected<br />

Heated Wire Faulty, Disconnected<br />

Low. Measured temperature is below set temperature by 4°C<br />

for more than 20 mins, or the measured temperature is below set<br />

temperature by 10°C<br />

Heater Plate Measured chamber temperature does not rise by more than 2°C<br />

in 20 minutes from start up<br />

Physical Dimensions<br />

Size (Without Humidification Chamber Installed) Height - 145mm / Width - 140mm / Depth – 180mm<br />

Weight 1.7 kg. (approx.)


<strong>Intersurgical</strong><br />

<strong>7000</strong> <strong>Series</strong> <strong>Respiratory</strong> <strong>Humidifier</strong><br />

(PMH<strong>7000</strong>)<br />

Power Requirements<br />

REF Operating Voltage Line Frequency<br />

<strong>7000</strong>000 230V ~ +/- 10% 50Hz<br />

<strong>7000</strong>001 100V ~ +/- 10% 50 / 60Hz<br />

<strong>7000</strong>002 230V ~ +/- 10% 50Hz<br />

Power Consumption 250 VA Maximum<br />

Current Rating 230V - 1.1A 100V - 2. 5A<br />

Heater Plate Power 150 Watts<br />

Heater Wire Power Inspiratory 44 Watts (Max. DC24V, 1.8A), regulated.<br />

Expiratory 44 Watts (Max. DC24V, 1.8A), regulated.<br />

Heater Plate Temperature 100°C Thermostat regulated (normal operation).<br />

130°C Thermostat limited (AC power disabled).<br />

Heater Wire Resistance 11 to 50 Ω (ohms)<br />

MDD Classification IIb<br />

Electric Shock Classification: Class I, type B Applied Part<br />

Protection against Water Classification: IPX1<br />

Mode of Operation Classification: Continuous Operation<br />

Environmental<br />

Operating Temperature 10°C to 40°C (50°F to 104°F)<br />

Recommended Ambient Temperature 20°C to 26°C (68°F to 79°F)<br />

Storage Temperature -20°C to 60°C (-4°F to 140°F)<br />

Operating Relative Humidity 30 to 75% Non-condensing<br />

Storage Relative Humidity 10 to 95% Non-condensing<br />

Atmospheric Pressure 700 to 1060 hPa<br />

Standards Applied<br />

JIS Medical Electrical Equipment, JIS Electrical Devices, ISO8185, UL60601-1 (CAN/CSA-C22.2 No. 601.1-M90),<br />

IEC60601-1, IEC60601-2.<br />

REF Product Description<br />

<strong>7000</strong>000 <strong>7000</strong> <strong>Series</strong> <strong>Respiratory</strong> <strong>Humidifier</strong> (UK)<br />

<strong>7000</strong>001 <strong>7000</strong> <strong>Series</strong> <strong>Respiratory</strong> <strong>Humidifier</strong> (JPN)<br />

<strong>7000</strong>002 <strong>7000</strong> <strong>Series</strong> <strong>Respiratory</strong> <strong>Humidifier</strong> (EU1): [ fr, es, it, de, nl ]<br />

5601000 Electrical adaptor lead (Dual heated wire system)<br />

5600000 Electrical adaptor lead (Single heated wire system)<br />

7<strong>7000</strong>00 Dual temperature probe assembly<br />

Manufactured by:<br />

Pacific Medico Co. Ltd, 2-6-4 Hongo, Bunkyo-ku, Tokyo 1133-0033 Japan<br />

Distributed by:<br />

Issue 3: TS 02.13<br />

United Kingdom (Head Office): Crane House, Molly Millars Lane, Wokingham, Berkshire, RG41 2RZ<br />

T: +44 (0)118 9656 300 F: +44 (0)118 965 6356 info@intersurgical.com | www.intersurgical.com<br />

France Deutschland España Pocc South Africa Japan<br />

info@intersurgical.fr info@intersurgical.de info@intersurgical-es.com info@intersurgical.ru info@intersurgical.co.za info@intersurgical.co.za<br />

Portugal Nederland Lietuva Česká Republika Philippines<br />

info@intersurgical.pt info@intersurgical.nl info@intersurgical.lt info@intersurgical.cz info@intersurgical.ph

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!